Short-term in vivo precision of BMD and parameters of trabecular architecture at the distal forearm and tibia

被引:74
作者
Engelke, K. [1 ,3 ]
Stampa, B. [1 ]
Timm, W. [1 ]
Dardzinski, B. [4 ]
de Papp, A. E. [4 ]
Genant, H. K. [5 ]
Fuerst, T. [2 ]
机构
[1] Synarc Inc, D-22087 Hamburg, Germany
[2] Synarc Inc, San Francisco, CA USA
[3] Univ Erlangen Nurnberg, Inst Med Phys, D-91054 Erlangen, Germany
[4] Merck Sharp & Dohme Corp, Whitehouse Stn, NJ USA
[5] Univ Calif San Francisco, Dept Radiol, San Francisco, CA USA
关键词
Bone mineral density; High-resolution peripheral QCT; Multicenter study; Short-term in vivo precision; Trabecular structure; QUANTITATIVE COMPUTED-TOMOGRAPHY; HR-PQCT; BONE MICROARCHITECTURE; REPRODUCIBILITY; RADIUS; QUALITY;
D O I
10.1007/s00198-011-1829-1
中图分类号
R5 [内科学];
学科分类号
100201 [内科学];
摘要
In vivo hr-pQCT precision was determined in 42 postmenopausal women using double baseline measurements from a multicenter trial of odanacatib. Errors, e.g., at the radius below 1.3% for BMD and below 6.3% for trabecular structure, were comparable to single-center results. Motion artifacts remain a challenge, particularly at the forearm. The short-term in vivo precision of BMD, trabecular bone structure, cortical thickness and porosity of the forearm and tibia was measured by hr-pQCT. Also the effect of image quality on precision was evaluated. In 42 postmenopausal women (age 64.4 +/- 6.8 years) out of 214 subjects enrolled in a multi center advanced imaging phase III study of odanacatib (DXA spine or hip T-scores between -1.5 and -3.5), double baseline hr-pQCT (XtremeCT) measurements with repositioning were performed. The standard ultradistal location and a second, more proximally located VOI were measured at the radius and tibia to better assess cortical thickness and porosity. Image analysis and quality grading (grades: perfect, slight artifacts, pronounced artifacts, unacceptable) were performed centrally. At the radius RMS%CV values varied from 0.7% to 1.3% for BMD and BV/TV and from 5.6% to 6.3% for Tb.Sp, Tb.Th, Tb.N, and cortical porosity. Numerically at the tibia, precision errors were approx. 0.5% lower for BMD and 1% to 2% lower for structural parameters although most differences were insignificant. In the radius but not in the tibia, precision errors for cortical thickness were smaller at the distal compared to the ultradistal location (1% versus 2%). BMD precision errors were lower than those for trabecular architecture and cortical porosity. Motion artifacts remain a challenge, particularly at the forearm. Quality grading remains subjective, and more objective evaluation methods are needed. Precision in the context of a multicenter clinical trial, with centralized training and scan analysis, was comparable to single-center results previously reported.
引用
收藏
页码:2151 / 2158
页数:8
相关论文
共 17 条
[1]
[Anonymous], 2005, XTREMECT US GUID REV
[2]
In vivo assessment of trabecular bone microarchitecture by high-resolution peripheral quantitative computed tomography [J].
Boutroy, S ;
Bouxsein, ML ;
Munoz, F ;
Delmas, PD .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 2005, 90 (12) :6508-6515
[3]
Automatic segmentation of cortical and trabecular compartments based on a dual threshold technique for in vivo micro-CT bone analysis [J].
Buie, Helen R. ;
Campbell, Graeme M. ;
Klinck, R. Joshua ;
MacNeil, Joshua A. ;
Boyd, Steven K. .
BONE, 2007, 41 (04) :505-515
[4]
A local adaptive threshold strategy for high resolution peripheral quantitative computed tomography of trabecular bone [J].
Burghardt, Andrew J. ;
Kazakia, Galateia J. ;
Majumdar, Sharmila .
ANNALS OF BIOMEDICAL ENGINEERING, 2007, 35 (10) :1678-1686
[5]
Reproducibility of direct quantitative measures of cortical bone microarchitecture of the distal radius and tibia by HR-pQCT [J].
Burghardt, Andrew J. ;
Buie, Helen R. ;
Laib, Andres ;
Majumdar, Sharmila ;
Boyd, Steven K. .
BONE, 2010, 47 (03) :519-528
[6]
Comparison of anatomic coordinate systems with rigid multi-resolution 3D registration for the reproducible positioning of analysis volumes of interest in QCT [J].
Eisa, Fabian ;
Museyko, Oleg ;
Hess, Andreas ;
Kalender, Willi A. ;
Engelke, Klaus .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (05) :1429-1439
[7]
Fuerst T, 2010, 32 ANN M AM SOC BON
[8]
GRAMPP S, 1995, J BONE MINER RES, V10, P1566
[9]
In vivo determination of bone structure in postmenopausal women: a comparison of HR-pQCT and high-field MR imaging [J].
Kazakia, Galateia J. ;
Hyun, Benedict ;
Burghardt, Andrew J. ;
Krug, Roland ;
Newitt, David C. ;
de Papp, Anne E. ;
Link, Thomas M. ;
Majumdar, Sharmila .
JOURNAL OF BONE AND MINERAL RESEARCH, 2008, 23 (04) :463-474
[10]
Precision of the XCT 3000 and Comparison of Densitometric Measurements in Distal Radius Scans Between XCT 3000 and XCT 2000 Peripheral Quantitative Computed Tomography Scanners [J].
Koenig, Cory ;
Wey, Howard ;
Binkley, Teresa .
JOURNAL OF CLINICAL DENSITOMETRY, 2008, 11 (04) :575-580