Electrothermal atomic absorption spectrometric determination of cadmium in Bangladeshi vegetables with a metal tube atomizer and slurry sampling technique
被引:6
作者:
Amin, MN
论文数: 0引用数: 0
h-index: 0
机构:
Mie Univ, Fac Engn, Dept Chem Mat, Tsu, Mie 5148507, JapanMie Univ, Fac Engn, Dept Chem Mat, Tsu, Mie 5148507, Japan
Amin, MN
[1
]
Kaneco, S
论文数: 0引用数: 0
h-index: 0
机构:Mie Univ, Fac Engn, Dept Chem Mat, Tsu, Mie 5148507, Japan
Kaneco, S
Suzuki, T
论文数: 0引用数: 0
h-index: 0
机构:Mie Univ, Fac Engn, Dept Chem Mat, Tsu, Mie 5148507, Japan
Suzuki, T
Ohta, K
论文数: 0引用数: 0
h-index: 0
机构:Mie Univ, Fac Engn, Dept Chem Mat, Tsu, Mie 5148507, Japan
Ohta, K
机构:
[1] Mie Univ, Fac Engn, Dept Chem Mat, Tsu, Mie 5148507, Japan
[2] Mie Univ, Environm Preservat Ctr, Tsu, Mie 5148507, Japan
Ultrasonic slurry sampling electrothermal atomic absorption spectrometry with a molybdenum tube atomizer has been applied for the determination of cadmium in vegetable samples in Bangladesh. The suspension-stabilizing medium was 10% glycerol solution. The optimum pyrolysis temperature was 300degreesC. The detection limit was 13 fg (3 S/N ). Matrix element interference was studied and it was found that thiourea as a chemical modifier eliminated the interference. The results for the determination of cadmium in vegetable samples by the proposed method were in good agreement with those measured in dissolved acid-digested samples. The method enables rapid calibration, and simple and rapid analysis of cadmium in vegetable samples at low cost.