Spiral CT and radiation dose

被引:44
作者
Imhof, H
Schibany, N
Ba-Ssalamah, A
Czerny, C
Hojreh, A
Kainberger, F
Krestan, C
Kudler, H
Nöbauer, I
Nowotny, R
机构
[1] Univ Vienna, Klin Radiodiagnost, Dept Osteol, A-1090 Vienna, Austria
[2] AKH Wien, Inst Biomed Techn & Phys, Vienna, Austria
关键词
spiral CT; radiation dose; computed tomography dose index-value; low-dose CT;
D O I
10.1016/S0720-048X(02)00232-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Recent Studies in the USA and Europe state that computed tomography (CT) scans compromise only 3-5'Y,, of all radiological exams. but they contribute 35-45% of total radiation dose to the patient population. These studies lead to concern by several public authorities. Basis of CT-dose measurements is the computed tomography dose index (CTDI), which was established 1981. Nowadays there are several modifications of the CTDI values, which may lead to confusion. It is suggested to use the standardized CTDI-100 w. value together with the dose length product in all CT-examinations. These values should be printed on all CT-images and allows an evaluation of the individualized patient dose. Nowadays, radiologist's aim must be to work at the lowest maximal diagnostic acceptable signal to noise ratio. To decrease radiation dose radiologist should use low kV and mA, but high pitches. Newly developed CT-dose-reduction soft-wares and filters should be installed in all CT-machines. We should critically compare the average dose used for a specific examination with the reference dose used in this country and/or Europe. Greater differences should caution the radiologist. Finally, we as radiologists must check very carefully all indications and recommend alternative imaging methods. But we have also to teach our customers-patients and medical doctors who are non-radiologists-that a 'good' image is not that which show all possible information, but that which visualize 'only' the diagnostic necessary information. (C) 2002 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:29 / 37
页数:9
相关论文
共 15 条
  • [1] [Anonymous], 1999, 16262 EUR
  • [2] DONELLY LF, 2001, CT RAD CHILDREN
  • [3] Galanski M, 2000, Rofo, V172, pM164
  • [4] Kachelriess M, 1999, RADIOLOGY, V213P, P283
  • [5] Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadavers
    Kalender, WA
    Wolf, H
    Suess, C
    Gies, M
    Greess, H
    Bautz, WA
    [J]. EUROPEAN RADIOLOGY, 1999, 9 (02) : 323 - 328
  • [6] Kalender Willi A., 2000, COMPUTERTOMOGRAPHIE
  • [7] Effective doses to members of the public from the diagnostic application of ionizing radiation in Germany
    Kaul, A
    Bauer, B
    Bernhardt, J
    Nosske, D
    Veit, R
    [J]. EUROPEAN RADIOLOGY, 1997, 7 (07) : 1127 - 1132
  • [8] ANATOMICALLY ADAPTED VARIATION OF THE TUBE CURRENT IN CT - STUDIES ON REDUCTION OF RADIATION-DOSE AND IMAGE QUALITY
    KOPKA, L
    FUNKE, M
    BREITER, N
    HERMANN, KP
    VOSSHENRICH, R
    GRABBE, E
    [J]. FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1995, 163 (05): : 383 - 387
  • [9] Kudler H., 2002, STRAHLENDOSISMESSUNG
  • [10] Rothenberg L.N., 2001, RC, V823, P97