Mean cardiac output by thermodilution with a single controlled injection

被引:12
作者
Jansen, JRC
Schreuder, JJ
Punt, KD
van den Berg, PCM
Alfieri, O
机构
[1] Leiden Univ, Med Ctr, Dept Intens Care, NL-2300 RC Leiden, Netherlands
[2] Hosp San Raffaele, Dept Cardiac Surg, I-20132 Milan, Italy
[3] Atrium Med Ctr, Dept Anesthesiol & Intens Care, Heerlen, Netherlands
关键词
cardiac output; mechanical ventilation; thermodilution; controlled injection; flow modulation;
D O I
10.1097/00003246-200110000-00003
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective. A new method to estimate mean cardiac output by thermodilution with a single duration-controlled injection was evaluated in patients. Design. Prospective criterion standard study. Setting. University hospital cardiac surgical intensive care unit and cardiac operation room. Patients. Of 33 patients, 24 underwent coronary bypass graft surgery, four had a valve replacement, and five were treated in the intensive care unit. Interventions. Interventions consisted of thermodilution cardiac output measurements. One single duration-controlled injection of cold fluid was used to calculate cardiac output. This controlled injection was performed with a duration equal to one whole ventilation cycle of the ventilator. An algorithm adapted to this duration-controlled injection calculated cardiac output. Moreover, this algorithm has properties to reduce errors caused by artificial ventilation and thermal noise. Measurements and Main Results. In 33 patients, the averaged values of four measurements equally spread over the ventilatory cycle (phase-controlled) were compared with the values of two single duration-controlled measurements. The measurements were performed during periods of stable respiration and circulation. No significant difference was observed between the mean of four phase-controlled measurements and the mean of the two duration-controlled measurements. The cardiac output values in the intensive care patients were significantly higher compared with the two other patient groups (p < .05). The difference between the two methods could not be subdivided for the three patient groups (p > .05). The coefficient of variation of the single duration-controlled thermodilution measurements was significantly lower than the single phase-controlled measurements, 3% vs. 6% (p < .01). Conclusions. One single duration-controlled injection thermodilution measurement is as accurate and repeatable as the mean of four phase-controlled measurements and is clinically feasible.
引用
收藏
页码:1868 / 1873
页数:6
相关论文
共 24 条
[11]   CONTINUOUS CARDIAC-OUTPUT MONITORING BY PULSE CONTOUR DURING CARDIAC-SURGERY [J].
JANSEN, JRC ;
WESSELING, KH ;
SETTELS, JJ ;
SCHREUDER, JJ .
EUROPEAN HEART JOURNAL, 1990, 11 :26-32
[12]  
JANSEN JRC, 1986, INTENS CARE MED, V12, P71
[13]   AN ADEQUATE STRATEGY FOR THE THERMODILUTION TECHNIQUE IN PATIENTS DURING MECHANICAL VENTILATION [J].
JANSEN, JRC ;
SCHREUDER, JJ ;
SETTELS, JJ ;
KLOEK, JJ ;
VERSPRILLE, A .
INTENSIVE CARE MEDICINE, 1990, 16 (07) :422-425
[14]   THERMODILUTION TECHNIQUE FOR MEASUREMENT OF CARDIAC-OUTPUT DURING ARTIFICIAL-VENTILATION [J].
JANSEN, JRC ;
SCHREUDER, JJ ;
BOGAARD, JM ;
VANROOYEN, W ;
VERSPRILLE, A .
JOURNAL OF APPLIED PHYSIOLOGY, 1981, 51 (03) :584-591
[15]   Single injection thermodilution - A flow-corrected method [J].
Jansen, JRC ;
Schreuder, JJ ;
Settels, JJ ;
Kornet, L ;
Penn, OCKM ;
Mulder, PGH ;
Versprille, A ;
Wesseling, KH .
ANESTHESIOLOGY, 1996, 85 (03) :481-490
[16]  
LATSON TW, 1993, ANESTHESIOLOGY, V79, P1233
[17]   A multicenter evaluation of a new continuous cardiac output pulmonary artery catheter system [J].
Mihm, FG ;
Gettinger, A ;
Hanson, CW ;
Gilbert, HC ;
Stover, EP ;
Vender, JS ;
Beerle, B ;
Haddow, G .
CRITICAL CARE MEDICINE, 1998, 26 (08) :1346-1350
[18]   EFFECTS OF INTERMITTENT POSITIVE-PRESSURE VENTILATION ON CARDIAC-OUTPUT MEASUREMENTS BY THERMODILUTION [J].
OKAMOTO, K ;
KOMATSU, T ;
KUMAR, V ;
SANCHALA, V ;
KUBAL, K ;
BHALODIA, R ;
SHIBUTANI, K .
CRITICAL CARE MEDICINE, 1986, 14 (11) :977-980
[19]   Delayed time response of the continuous cardiac output pulmonary artery catheter [J].
Siegel, LC ;
Hennessy, MM ;
Pearl, RG .
ANESTHESIA AND ANALGESIA, 1996, 83 (06) :1173-1177
[20]   EFFECTS OF MECHANICAL VENTILATION ON THE MEASUREMENT OF CARDIAC-OUTPUT BY THERMODILUTION [J].
SNYDER, JV ;
POWNER, DJ .
CRITICAL CARE MEDICINE, 1982, 10 (10) :677-682