Effect of collimator choice on quantitative assessment of cardiac iodine 123 MIBG uptake

被引:50
作者
Inoue, Y
Suzuki, A
Shirouzu, I
Machida, T
Yoshizawa, Y
Akita, F
Ohnishi, S
Yoshikawa, KA
Ohtomo, K
机构
[1] Univ Tokyo, Inst Med Sci, Dept Radiol, Minato Ku, Tokyo 1088639, Japan
[2] Univ Tokyo, Dept Radiol, Grad Sch Med, Tokyo 1088639, Japan
[3] Kanto Med Ctr NTT EC, Dept Cardiovasc Med, Tokyo, Japan
[4] Kanto Med Ctr NTT EC, Dept Radiol, Tokyo, Japan
关键词
iodine; 123; metaiodobenzyiguanidine; collimator; quantitation; septal penetration;
D O I
10.1016/S1071-3581(03)00652-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Quantitative accuracy in iodine 123 studies may be impaired by septa, penetration. We evaluated the effect of collimator choice on estimation of the heart-to-mediastinum (H/M) ratio in cardiac 1-123 metaiodobenzylguanidine (MIBG) imaging. Methods and Results. A low-energy high-resolution (LEHR) collimator, special LEHR (SLEHR) collimator, and medium-energy (ME) collimator were used. In experiments in which a phantom of simple geometry was used, the use of the LEHR collimator provided the lowest contrast accuracy, suggesting the effect of septal penetration. Thoracic phantom studies demonstrated contamination of heart and mediastinum counts by lung and liver activities, which was greatest with the LEHR collimator and least with the ME collimator. In 8 patients anterior chest views were acquired successively with the three collimators after 1-123 MIBG injection. H/M ratios were significantly higher with the SLEHR collimator than with the LEHR collimator and were still higher with the ME collimator. The difference in H/M ratios between the LEHR and ME collimators showed a high positive correlation with the lung-to-mediastinum ratio. Conclusions. Collimator choice substantially influences estimation of the H/M ratios in cardiac 1-123 MIBG imaging. The use of an ME collimator provides high quantitative accuracy and may enhance reliability in the evaluation of cardiac sympathetic nerve function.
引用
收藏
页码:623 / 632
页数:10
相关论文
共 17 条
[1]   SCATTER CORRECTION IN SCINTIGRAPHY - THE STATE-OF-THE-ART [J].
BUVAT, I ;
BENALI, H ;
TODDPOKROPEK, A ;
DIPAOLA, R .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1994, 21 (07) :675-694
[2]   Optimal collimator choice for sequential iodine-123 and technetium-99m imaging [J].
DeGeeter, F ;
Franken, PR ;
Defrise, M ;
Andries, H ;
Saelens, E ;
Bossuyt, A .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1996, 23 (07) :768-774
[3]   Influence of high-energy photons on the spectrum of iodine-123 with low- and medium-energy collimators:: consequences for imaging with 123I-labelled compounds in clinical practice [J].
Dobbeleir, AA ;
Hambye, ASE ;
Franken, PR .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1999, 26 (06) :655-658
[4]  
GILLAND DR, 1994, J NUCL MED, V35, P1707
[5]   Effects of scatter and attenuation correction on quantitative analysis of β-CIT brain SPET [J].
Hashimoto, J ;
Sasaki, T ;
Ogawa, K ;
Kubo, A ;
Motomura, N ;
Ichihara, T ;
Amano, T ;
Fukuuchi, Y .
NUCLEAR MEDICINE COMMUNICATIONS, 1999, 20 (02) :159-165
[6]   Metalodobenzylguanidine scintigraphy of the heart: what have we learnt clinically? [J].
Hattori, N ;
Schwaiger, M .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 2000, 27 (01) :1-6
[7]  
ICHIHARA T, 1993, J NUCL MED, V34, P2216
[8]  
Iida H, 1998, J NUCL MED, V39, P181
[9]   Therapeutic effect of co-enzyme Q10 on idiopathic dilated cardiomyopathy: assessment by iodine-123 labelled 15-(p-iodophenyl)-3(R, S)-methylpentadecanoic acid myocardial single-photon mission tomography [J].
Yong-ih Kim ;
Yoshihiro Sawada ;
Go Fujiwara ;
Hiroshi Chiba ;
Tsunehiko Nishimura .
European Journal of Nuclear Medicine, 1997, 24 (6) :629-634
[10]  
MACEY DJ, 1986, J NUCL MED, V27, P1467