Composite synthetic models of the human femur have recently become commercially available as substitutes for cadaveric specimens. Their quick diffusion was justified by the advantages they offer as a substitute for real femurs. The present investigation concentrated on an extensive experimental validation of the mechanical behaviour of the whole bone composite model, compared to human fresh-frozen and dried-rehydrated specimens for different loading conditions. First, the viscoelastic behaviour of the models was investigated under simulated single leg stance loading, showing that the little time dependent phenomena observed tend to extinguish within a few minutes of the load application. The behaviour under axial loading was then studied by comparing the vertical displacement of the head as well as the axial strains, by application of a parametric descriptive model of the strain distribution. Finally, a four point bending test and a torsional test were performed to characterize the whole bone stiffness of the femur. In all these tests, the composite femurs were shown to fall well within the range for cadaveric specimens, with no significant differences being detected between the synthetic femurs and the two groups of cadaveric femurs. Moreover, the interfemur variability for the composite femurs was 20-200 times lower than that for the cadaveric specimens, thus allowing smaller differences to be characterized as significant using the same simple size, if the composite femurs are employed.
引用
收藏
页码:525 / 535
页数:11
相关论文
共 44 条
[31]   ELASTIC AND ULTIMATE PROPERTIES OF COMPACT BONE TISSUE [J].
REILLY, DT ;
BURSTEIN, AH .
JOURNAL OF BIOMECHANICS, 1975, 8 (06) :393-&
[32]   FINITE-ELEMENT-ANALYSIS AND EXPERIMENTAL INVESTIGATION IN A FEMUR WITH HIP ENDOPROSTHESIS [J].
ROHLMANN, A ;
MOSSNER, U ;
BERGMANN, G ;
KOLBEL, R .
JOURNAL OF BIOMECHANICS, 1983, 16 (09) :727-742
[33]   THE MORPHOLOGY OF THE PROXIMAL FEMUR - A 3-DIMENSIONAL RADIOGRAPHIC ANALYSIS [J].
RUBIN, PJ ;
LEYVRAZ, PF ;
AUBANIAC, JM ;
ARGENSON, JN ;
ESTEVE, P ;
DEROGUIN, B .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01) :28-32
[34]   CROSS-SECTIONAL GEOMETRY OF PECOS PUEBLO FEMORA AND TIBIAE - A BIOMECHANICAL INVESTIGATION .1. METHOD AND GENERAL PATTERNS OF VARIATION [J].
RUFF, CB ;
HAYES, WC .
AMERICAN JOURNAL OF PHYSICAL ANTHROPOLOGY, 1983, 60 (03) :359-381
[35]   STRESS-RELAXATION FUNCTION OF BONE AND BONE-COLLAGEN [J].
SASAKI, N ;
NAKAYAMA, Y ;
YOSHIKAWA, M ;
ENYO, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (12) :1369-1376
[36]  
Shybut G T, 1980, Proc Inst Med Chic, V33, P95
[37]   CHANGES IN TRABECULAR PATTERN OF UPPER END OF FEMUR AS AN INDEX OF OSTEOPOROSIS [J].
SINGH, M ;
NAGRATH, AR ;
MAINI, PS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1970, A 52 (03) :457-&
[38]   COMPARISON OF THE DEFORMATION RESPONSE OF SYNTHETIC AND CADAVERIC FEMORA DURING SIMULATED ONE-LEGGED STANCE [J].
SZIVEK, JA ;
GEALER, RL .
JOURNAL OF APPLIED BIOMATERIALS, 1991, 2 (04) :277-280
[39]   CHARACTERIZATION OF A SYNTHETIC FOAM AS A MODEL FOR HUMAN CANCELLOUS BONE [J].
SZIVEK, JA ;
THOMAS, M ;
BENJAMIN, JB .
JOURNAL OF APPLIED BIOMATERIALS, 1993, 4 (03) :269-272
[40]   VARIABILITY IN THE TORSIONAL AND BENDING RESPONSE OF A COMMERCIALLY AVAILABLE COMPOSITE FEMUR [J].
SZIVEK, JA ;
WENG, M ;
KARPMAN, R .
JOURNAL OF APPLIED BIOMATERIALS, 1990, 1 (02) :183-186