Fluid absorption study in ultra-high molecular weight polyethylene( UHMWPE) sterilized and unsterilized acetabular cups

被引:25
作者
Affatato, S
Vandelli, C
Bordini, B
Toni, A
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
[2] Ist Ortoped Rizzoli, Div 1, I-40136 Bologna, Italy
关键词
water absorption; serum absorption; polyethylene absorption; fluid absorption; hip simulator;
D O I
10.1243/0954411011533463
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The weight gain due to fluid absorption was measured in gamma -sterilized, ethylene oxide (ETO) gas-sterilized, and unsterilized ultra-high molecular weight (UHMW) polyethylene acetabular cups. After about 2 months the total average fluid absorption gain of the polyethylene cups was: 1.76 +/- 0.45 mg (average +/- standard deviation) for the unsterilized polyethylene cups, 2.81 +/- 0.95 mg for the gamma -sterilized polyethylene cups and 1.51 +/- 0.34 for the ETO gas-sterilized polyethylene cups. There was little difference in fluid absorption between the sterilized and unsterilized specimens. In particular, the gamma -sterilized acetabular cups absorbed more than the ETO gas-sterilized and the unsterilized cups. However, the weight gain was higher in serum compared with that in water.
引用
收藏
页码:107 / 111
页数:5
相关论文
共 18 条
[11]   WEAR CHARACTERISTICS OF UHMW POLYETHYLENE - METHOD FOR ACCURATELY MEASURING EXTREMELY LOW WEAR RATES [J].
MCKELLOP, H ;
CLARKE, IC ;
MARKOLF, KL ;
AMSTUTZ, HC .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1978, 12 (06) :895-927
[12]   Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements [J].
McKellop, H ;
Shen, FW ;
Lu, B ;
Campbell, P ;
Salovey, R .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1999, 17 (02) :157-167
[13]  
MCKELLOP H, 1992, P 38 ANN M ORTH RES, P402
[14]  
MCKELLOP HA, 1981, ENG MED, V10, P123
[15]  
Saikko V, 1992, Proc Inst Mech Eng H, V206, P195, DOI 10.1243/PIME_PROC_1992_206_291_02
[16]  
SCARDOVI I, 1993, METODOLOGIA STAT BIO, P404
[17]   PERIPROSTHETIC BONE LOSS IN TOTAL HIP-ARTHROPLASTY - POLYETHYLENE WEAR DEBRIS AND THE CONCEPT OF THE EFFECTIVE JOINT SPACE [J].
SCHMALZRIED, TP ;
JASTY, M ;
HARRIS, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (06) :849-863
[18]  
Wang A, 1996, Proc Inst Mech Eng H, V210, P141, DOI 10.1243/PIME_PROC_1996_210_407_02