Differential diagnosis of gammopathies by capillary electrophoresis and immunosubtraction: Analysis of serum samples problematic by agarose gel electrophoresis

被引:22
作者
Clark, R
Katzmann, JA
Kyle, RA
Fleisher, M
Landers, JP [1 ]
机构
[1] Univ Pittsburgh, Chevron Sci Ctr, Dept Chem, Pittsburgh, PA 15260 USA
[2] Univ Pittsburgh, Inst Canc, Pittsburgh, PA 15260 USA
[3] Mem Sloan Kettering Canc Ctr, Protein Lab, Dept Lab Med & Pathol, New York, NY 10021 USA
[4] Mem Sloan Kettering Canc Ctr, Clin Chem Serv, Dept Clin Labs, New York, NY 10021 USA
关键词
capillary electrophoresis; serum proteins; immunosubtraction; immunofixation electrophoresis; monoclonal protein;
D O I
10.1002/elps.1150191421
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The capabilities of capillary electrophoresis (CE) for serum protein electrophoresis and immunotyping have been demonstrated. CE-based systems specifically designed for serum protein electrophoresis and immunotyping via immunosubtraction (IS) are now available and are being evaluated for efficiency, specificity and sensitivity by several groups. The use of CE for serum protein electrophoresis and immunotyping (IS) in the clinical laboratory compares well with agarose gel electrophoresis (AGE) and immunofixation (IF) for the detection and characterization of monoclonal proteins. In addition to routine use, this technology is useful for a subset of serum samples that are difficult to interpret with conventional technology. In this study, sera abnormalities difficult to detect/interpret by AGE-IF are subdivided into four categories: (i) patients with polyclonal increases in immunoglobulin, (ii) point of application artifacts, (iii) abnormalities in the beta region, and (iv) patients with free light chains. CE is superior to AGE for evaluating samples characterized by the above abnormalities. Sera containing monoclonal proteins within a polyclonal increase are easier to detect by CE as well as being easier to type by IS than by IF. Point-of-application artifacts, periodically observed with AGE, do not exist on CE since the point of detection is remote from the point of application. Enhanced resolution in the beta region allows for increased detection of monoclonal proteins migrating in this region. Some free light chains are undetected by CE as a result of no apparent abnormalities on the CE serum protein profile and, thus, still require IF for detection. CE detects more serum electrophoretic abnormalities than AGE in this clinically important group of patients with Bence Jones proteinemia.
引用
收藏
页码:2479 / 2484
页数:6
相关论文
共 11 条
[1]   RAPID PROTEIN-ANALYSIS BY CAPILLARY ELECTROPHORESIS [J].
CHEN, FTA .
JOURNAL OF CHROMATOGRAPHY, 1991, 559 (1-2) :445-453
[2]  
CHEN FTA, 1991, CLIN CHEM, V37, P14
[3]   Rapid capillary electrophoretic analysis of human serum proteins: Qualitative comparison with high-throughput agarose gel electrophoresis [J].
Clark, R ;
Katzmann, JA ;
Wiegert, E ;
NamystGoldberg, C ;
Sanders, L ;
Oda, RP ;
Kyle, RA ;
Landers, JP .
JOURNAL OF CHROMATOGRAPHY A, 1996, 744 (1-2) :205-213
[4]   EVALUATION OF SERUM-PROTEIN SEPARATION BY CAPILLARY ELECTROPHORESIS - PROSPECTIVE ANALYSIS OF 1000 SPECIMENS [J].
JENKINS, MA ;
KULINSKAYA, E ;
MARTIN, HD ;
GUERIN, MD .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1995, 672 (02) :241-251
[5]   QUANTIFICATION OF SERUM-PROTEINS USING CAPILLARY ELECTROPHORESIS [J].
JENKINS, MA ;
GUERIN, MD .
ANNALS OF CLINICAL BIOCHEMISTRY, 1995, 32 :493-497
[6]   Identification of monoclonal proteins in serum: A quantitative comparison of acetate, agarose gel, and capillary electrophoresis [J].
Katzmann, JA ;
Clark, R ;
Wiegert, E ;
Sanders, E ;
Oda, RP ;
Kyle, RA ;
NamystGoldberg, C ;
Landers, JP .
ELECTROPHORESIS, 1997, 18 (10) :1775-1780
[7]  
Katzmann JA, 1998, AM J CLIN PATHOL, V110, P503
[8]  
KIM JW, 1993, CLIN CHEM, V39, P689
[9]  
KLEIN G, 1993, HDB CAPILLARY ELECTR, P419
[10]  
Kyle R, 1997, MANUAL CLIN LAB IMMU, P156