Hepatic bipolar radio-frequency ablation between separated multiprong electrodes

被引:93
作者
Haemmerich, D
Staelin, ST
Tungjitkusolmun, S
Lee, FT
Mahvi, DM
Webster, JG
机构
[1] Univ Wisconsin, Dept Biomed Engn, Madison, WI 53706 USA
[2] Univ Wisconsin, Dept Surg, Madison, WI 53792 USA
[3] King Mongkuts Inst Technol Ladkrabang, Dept Elect Engn, Bangkok 10520, Thailand
[4] Univ Wisconsin, Dept Radiol, Madison, WI 53792 USA
关键词
bipolar ablation; electrode; finite element model; liver ablation; radio-frequency; RF ablation;
D O I
10.1109/10.951517
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Radio-frequency (RF) ablation has become an important means of treatment of nonresectable primary and metastatic liver tumors. Major limitations are small lesion size, which make multiple applications necessary, and incomplete killing of tumor cells, resulting in high recurrence rates. We examined a new bipolar RF ablation method incorporating two probes with hooked electrodes (RITA model 30). We performed monopolar and bipolar in vivo experiments on three pigs. The electrodes were 2.5 cm apart and rotated 45 degrees relative to each other. We used temperature-controlled mode at 95 degreesC. Lesion volumes were 3.9 +/- 1.8 cm(3) (n = 7) for the monopolar case and 12.2 +/- 13 cm(3) (n = 10) for the bipolar case. We generated finite-element models (FEMs) of monopolar and bipolar configurations. We analyzed the distribution of temperature and electric field of the finite element model. The lesion volumes for the FEM are 7.95 cm(3) for the monopolar and 18.79 cm(3) for the bipolar case. The new bipolar method creates larger lesions and is less dependent on local inhomogenities in liver tissue-such as blood perfusion-compared with monopolar RF ablation. A limitation of the new method is that the power dissipation of the two probes cannot be controlled independently in response to different conditions in the vicinity of each probe. This may result in nonuniform lesions and decreased lesion size.
引用
收藏
页码:1145 / 1152
页数:8
相关论文
共 26 条
[1]   RECENT DEVELOPMENTS IN MODELING HEAT-TRANSFER IN BLOOD-PERFUSED TISSUES [J].
ARKIN, H ;
XU, LX ;
HOLMES, KR .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (02) :97-107
[2]   Hepatic lesion ablation with bipolar saline-enhanced radiofrequency in the audible spectrum [J].
Burdío, F ;
Güemes, A ;
Burdío, JM ;
Castiella, T ;
De Gregorio, MA ;
Lozano, R ;
Livraghi, T .
ACADEMIC RADIOLOGY, 1999, 6 (11) :680-686
[3]   CT versus sonography for monitoring radiofrequency ablation in a porcine liver [J].
Cha, CH ;
Lee, FT ;
Gurney, JM ;
Markhardt, BK ;
Warner, TF ;
Kelcz, F ;
Mahvi, DM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 175 (03) :705-711
[4]  
Curley MG, 1997, P ANN INT IEEE EMBS, V19, P2516, DOI 10.1109/IEMBS.1997.756840
[5]   Laparoscopically guided bipolar radiofrequency ablation of areas of porcine liver [J].
Curley, SA ;
Davidson, BS ;
Fleming, RY ;
Izzo, F ;
Stephens, LC ;
Tinkey, P ;
Cromeens, D .
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1997, 11 (07) :729-733
[6]   Radiofrequency ablation of unresectable primary and metastatic hepatic malignancies - Results in 123 patients [J].
Curley, SA ;
Izzo, F ;
Delrio, P ;
Ellis, LM ;
Granchi, J ;
Vallone, P ;
Fiore, F ;
Pignata, S ;
Daniele, B ;
Cremona, F .
ANNALS OF SURGERY, 1999, 230 (01) :1-8
[7]   VASCULAR OCCLUSIONS FOR LIVER RESECTIONS - OPERATIVE MANAGEMENT AND TOLERANCE TO HEPATIC ISCHEMIA - 142 CASES [J].
DELVA, E ;
CAMUS, Y ;
NORDLINGER, B ;
HANNOUN, L ;
PARC, R ;
DERIAZ, H ;
LIENHART, A ;
HUGUET, C .
ANNALS OF SURGERY, 1989, 209 (02) :211-218
[8]   Minimally invasive treatment of malignant hepatic tumors: At the threshold of a major breakthrough [J].
Dodd, GD ;
Soulen, MC ;
Kane, RA ;
Livraghi, T ;
Lees, WR ;
Yamashita, Y ;
Gillams, AR ;
Karahan, OI ;
Rhim, H .
RADIOGRAPHICS, 2000, 20 (01) :9-27
[9]   A CYLINDRICAL-SECTION ULTRASOUND PHASED-ARRAY APPLICATOR FOR HYPERTHERMIA CANCER-THERAPY [J].
EBBINI, ES ;
UMEMURA, S ;
IBBINI, M ;
CAIN, CA .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1988, 35 (05) :561-572
[10]   Percutaneous radiofrequency tissue ablation: Optimization of pulsed radiofrequency technique to increase coagulation necrosis [J].
Goldberg, SN ;
Stein, MC ;
Gazelle, GS ;
Sheiman, RG ;
Kruskal, JB ;
Clouse, ME .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1999, 10 (07) :907-916