Quantitative IVUS blood flow:: Validation in vitro, in animals and in patients

被引:18
作者
Lupotti, FA
Mastik, F
Carlier, SG
de Korte, CL
van der Giessen, WJ
Serruys, PW
van der Steen, AFW
机构
[1] ICIN, Utrecht, Netherlands
[2] Erasmus MC, Thoraxctr, Rotterdam, Netherlands
[3] Thorax Ctr Rotterdam, Dept Intervent Cardiol, Rotterdam, Netherlands
关键词
intravascular ultrasound; correlation coefficients; dynamic noise suppression; catheter eccentricity; quantitative transverse blood flow; coronary flow reserve;
D O I
10.1016/S0301-5629(02)00774-3
中图分类号
O42 [声学];
学科分类号
070206 [声学]; 082403 [水声工程];
摘要
In recent years, a new method to measure transverse blood flow based on the decorrelation of the radiofrequency (RF) signals of intravascular ultrasound (IVUS) rotating single-element scanners was introduced. We report here in vitro, animal and patient testing to evaluate the correlation-based method using an IVUS array catheter. A new correlation-based method to dynamically correct the correlation coefficients for noise is implemented. The decorrelation due to noise was estimated from the correlation coefficients from flowing blood obtained at increasing time lags. First, blood flow experiments were carried out with different catheters in a tissue-mimicking flow phantom with an inner diameter ranging from 3.0 to 5.0 mm. A calibrated electromagnetic flow meter (EMF, range: 0 to 250 cc/min) was used as a reference. Good linear relationships were found between the IVUS-derived flow and the calibrated EMF (all R-2 > 0.96). The catheter position within the flow phantom and the size of the ring-down were theoretically analyzed. These elements, and noise in the RF signals, have an important influence on the IVUS blood flow measurements reflected by the offset and the slope of the linear relationships. By placing the IVUS catheter outside the flow phantom, parabolic blood flow profiles were also measured. Second, IVUS blood flow measurements were performed in the carotid artery of two Yorkshire pigs, which showed linear relationships (all R-2 > 0.85) between the IVUS-derived flow and the calibrated EMF. Experimentally, the offset was lower than 3 mL/min and the slope was close to 1. Third, IVUS blood flow measurements were performed in coronary arteries in patients. Preliminary results for the coronary flow reserve (CFR = high flow/baseline flow) in patients using the decorrelation method of RF signals of an array IVUS scanner were comparable with CFR based on Doppler measurements. (E-mail: a.vandersteen@erasmusmc.nl) (C) 2003 World Federation for Ultrasound in Medicine Biology.
引用
收藏
页码:507 / 515
页数:9
相关论文
共 33 条
[1]
Simulation of circular array ultrasound transducers for intravascular applications [J].
Borsboom, JMG ;
Céspedes, EI ;
van der Steen, AFW ;
Lancèe, CT ;
Deprettere, EF .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2000, 108 (02) :827-835
[2]
Simultaneous morphological and functional assessment of a renal artery stent intervention with intravascular ultrasound [J].
Carlier, SG ;
Li, WG ;
Céspedes, EI ;
van der Steen, AFW ;
Hamburger, JN ;
Bom, N ;
Serruys, PW .
CIRCULATION, 1998, 97 (25) :2575-2576
[3]
Coronary Flow reserve versus geometric measurements of coronary dimensions: Advantages and limitations of the functional stenosis assessment [J].
Carlier, SG ;
Van Langenhove, G ;
Lupotti, F ;
Albertal, M ;
Mastik, F ;
Bom, K ;
Serruys, PW .
JOURNAL OF INTERVENTIONAL CARDIOLOGY, 1999, 12 (06) :411-424
[4]
CARLIER SG, 1998, TXB INTERVENTIONAL C, P748
[5]
CESPEDES EI, 1998, J VASC INVEST, V4, P39
[6]
The combined effect of signal decorrelation and random noise on the variance of time delay estimation [J].
Cespedes, I ;
Ophir, J ;
Alam, SK .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1997, 44 (01) :220-225
[7]
MEASUREMENT OF VOLUMETRIC CORONARY BLOOD-FLOW BY SIMULTANEOUS INTRAVASCULAR 2-DIMENSIONAL AND DOPPLER ULTRASOUND - VALIDATION IN AN ANIMAL-MODEL [J].
CHOU, TM ;
SUDHIR, K ;
IWANAGA, S ;
CHATTERJEE, K ;
YOCK, PG .
AMERICAN HEART JOURNAL, 1994, 128 (02) :237-243
[8]
Characterization of plaque components with intravascular ultrasound elastography in human femoral and coronary arteries in vitro [J].
de Korte, CL ;
Pasterkamp, G ;
van der Steen, AFW ;
Woutman, HA ;
Bom, N .
CIRCULATION, 2000, 102 (06) :617-623
[9]
Di Mario C, 2000, CIRCULATION, V102, P2938
[10]
Accuracy of coronary flow measurements performed by means of Doppler wires [J].
Doriot, PA ;
Dorsaz, PA ;
Dorsaz, L ;
Chatelain, P .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 (02) :221-228