Thermal response of contrast agent microbubbles: Preliminary results from physico-chemical and US-imaging characterization

被引:18
作者
Guiot, C.
Pastore, G.
Napoleone, M.
Gabriele, P.
Trotta, M.
Cavalli, R.
机构
[1] Univ Turin, Dept Neurosci, I-10125 Turin, Italy
[2] Univ Turin, Dept Drug Sci & Technol, Turin, Italy
[3] IRCC & Mauriziano Hosp, Radiotherapy Unit, Candiolo, Italy
关键词
ultrasound contrast agents; hyperthermia; microbubbles; non-invasive temperature monitoring;
D O I
10.1016/j.ultras.2006.06.031
中图分类号
O42 [声学];
学科分类号
070206 [声学]; 082403 [水声工程];
摘要
Hyperthermia ( HT) is a therapeutic strategy based on the selective damaging of tumoral cells when heated at temperatures in the range 41 - 45 degrees C. We are currently investigating the feasibility of Ultrasound ( US) imaging to perform a non- invasive, efficient and cost effective temperature monitoring of heated tissues. Commercial US contrast agents ( Sonovue (R), Bracco), consisting in microbubbles of SF6 coated with a phospholipidic shell, greatly improve the US echo signal from tissues. Further investigations have been performed, consisting in physico- chemical and US- imaging characterization. In conclusion, we demonstrate that Sonovue (R) microbubbles reach their maximal diameter at 40 degrees C, and then a sharp decrease is observed, possible due to the occurrence of gel - sol transition of the phospholipidic shell. At the same temperature the maximal backscattering intensity is predicted and actually experimentally observed. Sonovue (R), as well as other contrast agents based only on phospholipids, are, therefore, not suitable for use as non- invasive temperature monitoring medium since it is sensitive to temperatures below the hyperthermic range. Although microbubbles are in principle thermally effective, other coating materials should be investigated in order to increase their operative thermal range. (c) 2006 Elsevier B. V. All rights reserved.
引用
收藏
页码:E127 / E130
页数:4
相关论文
共 10 条
[1]
PHANTOM CHARACTERIZATION OF APPLICATORS BY LIQUID-CRYSTAL-PLATE DOSIMETRY [J].
ANDREUCCETTI, D ;
BINI, M ;
IGNESTI, A ;
OLMI, R ;
VANNI, R .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1991, 7 (01) :175-183
[2]
BAMBER J C, 1979, Ultrasound in Medicine and Biology, V5, P149, DOI 10.1016/0301-5629(79)90083-8
[3]
ABSORPTION AND SCATTER OF ENCAPSULATED GAS FILLED MICROSPHERES - THEORETICAL CONSIDERATIONS AND SOME MEASUREMENTS [J].
DEJONG, N ;
HOFF, L ;
SKOTLAND, T ;
BOM, N .
ULTRASONICS, 1992, 30 (02) :95-103
[4]
Influence of bubble size distribution on the echogenicity of ultrasound contrast agents -: A study of SonoVue™ [J].
Gorce, JM ;
Arditi, M ;
Schneider, M .
INVESTIGATIVE RADIOLOGY, 2000, 35 (11) :661-671
[5]
Temperature monitoring using ultrasound contrast agents: in vitro investigation on thermal stability [J].
Guiot, C ;
Cavalli, R ;
Gaglioti, P ;
Danelon, D ;
Musacchio, C ;
Trotta, M ;
Todros, T .
ULTRASONICS, 2004, 42 (1-9) :927-930
[6]
Perfusion and thermal field during hyperthermia. Experimental measurements and modelling in recurrent breast cancer [J].
Guiot, C ;
Madon, E ;
Allegro, D ;
Pianta, PG ;
Baiotto, B ;
Gabriele, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (10) :2831-2843
[7]
THERMAL-TREATMENT PARAMETERS ARE MOST PREDICTIVE OF OUTCOME IN PATIENTS WITH SINGLE TUMOR NODULES PER TREATMENT FIELD IN RECURRENT ADENOCARCINOMA OF THE BREAST [J].
KAPP, DS ;
COX, RS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (04) :887-899
[8]
PC-aided assessment of the thermal performances of a MW applicator for oncological hyperthermia [J].
Marini, P ;
Guiot, C ;
Baiotto, B ;
Gabriele, P .
COMPUTERS IN BIOLOGY AND MEDICINE, 2004, 34 (01) :3-13
[9]
COUNTING BUBBLES ACOUSTICALLY - REVIEW [J].
MEDWIN, H .
ULTRASONICS, 1977, 15 (01) :7-13
[10]
THE CURRENT AND POTENTIAL ROLE OF HYPERTHERMIA IN RADIOTHERAPY [J].
OVERGAARD, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 16 (03) :535-549