Tune compounds for electrospray ionisation/in-source collision-induced dissociation with mass spectral library searching

被引:69
作者
Weinmann, W
Stoertzel, M
Vogt, S
Wendt, J
机构
[1] Univ Freiburg Klinikum, Inst Legal Med, D-79104 Freiburg, Germany
[2] Agilent Technol Deutschland GmbH, Chem Anal Tech, D-76337 Waldbronn, Germany
关键词
libraries; MS; mass spectrometry; haloperidol; paracetamol; metronidazole; metamizole; benzoylecgonine;
D O I
10.1016/S0021-9673(01)01066-4
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Haloperidol, paracetamol, metronidazole and metamizole have been tested as tune compounds for electrospray ionisation in-source collision-induced dissociation MS (ESI-CID-MS) with two different mass spectrometers (Sciex API 365 and Agilent 1100 MSD SL). The different electrospray sources of API 365 and MSD 1100 SL consist of an orifice with nitrogen curtain gas and a capillary interface, respectively. In-source CID occurs in both interfaces in front of the skimmers, which separate a region with a vacuum of approximately 300 Pa and the high vacuum (< 10(-3) Pa). Comparison of the breakdown curves of selected tune compounds, depending on collision energy (orifice or fragment or voltage), showed, that very similar fragmentation can be obtained with both instruments, when adjusting the fragment or voltage of the MSD 1100 SL to higher values than the orifice voltage of the API 365. For three energy levels - low, medium and high - the corresponding voltages were 20, 50 and 80 V for the API 365 and 110, 190, 230 V for the MSD 1100 SL. These voltages resulted in the most similar spectra for haloperidol and paracetamol with both instruments. The comparison of ESI-CID-MS of all tune compounds at three energy levels showed, that - despite variations in relative ion abundances - all significant ions were present in one of the three CID spectra. Therefore, mass spectral library searching of an ESI-CID-MS library set-up with one of the two instruments should be possible with the other instrument after adjusting the CID energies by means of at least two tune compounds such as haloperidol and paracetamol, metronidazole or metamizole. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:199 / 209
页数:11
相关论文
共 24 条
[1]   Poor reproducibility of in-source collisional atmospheric pressure ionization mass spectra of toxicologically relevant drugs [J].
Bogusz, MJ ;
Maier, RD ;
Krüger, KD ;
Webb, KS ;
Romeril, J ;
Miller, ML .
JOURNAL OF CHROMATOGRAPHY A, 1999, 844 (1-2) :409-418
[2]   MASS-SPECTROMETRY WITH ION SOURCES OPERATING AT ATMOSPHERIC-PRESSURE [J].
BRUINS, AP .
MASS SPECTROMETRY REVIEWS, 1991, 10 (01) :53-77
[3]  
GRINDSTAFF DJ, 1992, 40 C MASS SPECTR ALL, P37
[4]  
HEEREMANS CEF, 1990, 38TH P ASMS C MASS S, P1216
[5]  
HENION JD, 1992, 40 C MASS SPECTR ALL, P499
[6]   Evaluation of electrospray transport CID for the generation of searchable libraries [J].
Hough, JM ;
Haney, CA ;
Voyksner, RD ;
Bereman, RD .
ANALYTICAL CHEMISTRY, 2000, 72 (10) :2265-2270
[7]   CHARACTERIZATION OF OVER-THE-COUNTER COUGH COLD MEDICATIONS BY LIQUID-CHROMATOGRAPHY ELECTROSPRAY MASS-SPECTROMETRY [J].
JOSEPHS, JL .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1995, 9 (13) :1270-1274
[8]   USE OF A SINGLE-QUADRUPOLE MASS-SPECTROMETER FOR COLLISION-INDUCED DISSOCIATION STUDIES OF MULTIPLY CHARGED PEPTIDE IONS PRODUCED BY ELECTROSPRAY IONIZATION [J].
KATTA, V ;
CHOWDHURY, SK ;
CHAIT, BT .
ANALYTICAL CHEMISTRY, 1991, 63 (02) :174-178
[9]  
KATTA V, 1990, 38TH P ASMS C MASS S, P214
[10]  
KEEVER J, 1998, 46 ASMS C MASS SPECT, P86