Monte Carlo simulation for the ECAT EXACT HR plus system using GATE

被引:41
作者
Jan, S [1 ]
Comtat, C
Strul, D
Santin, G
Trébossen, R
机构
[1] Serv Hosp Frederic Joliot, F-91406 Orsay, France
[2] Ecole Polytech Fed Lausanne, LPHE, High Energy Phys Lab, CH-1015 Lausanne, Switzerland
关键词
counting rate; GATE; Geant4; magnetic field; Monte Carlo simulations; positron emission tomography (PET); positron range;
D O I
10.1109/TNS.2005.851461
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
GATE (Geant4 Application for Tomographic Emission) is a Monte Carlo simulation platform for nuclear medicine imaging, based on the Geant4 toolkit package. In order to improve data analysis and image quantification, a new simulator for the ECAT EXACT HR+ scanner was originally developed with GATE. In this study, we validate GATE simulations for positron emission topography (PET) imaging systems: a good agreement is obtained between simulated and actual experimental results. Additional GATE applications are also presented in the paper: an implementation of the voxelized Hoffman phantom and an example of a MRI/PET coupling system for improving image resolution.
引用
收藏
页码:627 / 633
页数:7
相关论文
共 11 条
[1]  
[Anonymous], NEMA STAND PUBL
[2]  
Brix G, 1997, J NUCL MED, V38, P1614
[3]  
Buvat I, 2002, Q J NUCL MED, V46, P48
[4]   MEASUREMENT OF POSITRON RANGE IN MATTER IN STRONG MAGNETIC-FIELDS [J].
HAMMER, BE ;
CHRISTENSEN, NL .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :1371-1376
[5]   gamma-Ray spectra from positron annihilation on atoms and molecules [J].
Iwata, K ;
Greaves, RG ;
Surko, CM .
PHYSICAL REVIEW A, 1997, 55 (05) :3586-3604
[6]   GePEToS: A Geant4 Monte Carlo simulation package for positron emission tomography [J].
Jan, S ;
Collot, J ;
Gallin-Martel, ML ;
Martin, P ;
Mayet, F ;
Tournefier, E .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (01) :102-106
[7]   GATE:: a simulation toolkit for PET and SPECT [J].
Jan, S ;
Santin, G ;
Strul, D ;
Staelens, S ;
Assié, K ;
Autret, D ;
Avner, S ;
Barbier, R ;
Bardiès, M ;
Bloomfield, PM ;
Brasse, D ;
Breton, V ;
Bruyndonckx, P ;
Buvat, I ;
Chatziioannou, AF ;
Choi, Y ;
Chung, YH ;
Comtat, C ;
Donnarieix, D ;
Ferrer, L ;
Glick, SJ ;
Groiselle, CJ ;
Guez, D ;
Honore, PF ;
Kerhoas-Cavata, S ;
Kirov, AS ;
Kohli, V ;
Koole, M ;
Krieguer, M ;
van der Laan, DJ ;
Lamare, F ;
Largeron, G ;
Lartizien, C ;
Lazaro, D ;
Maas, MC ;
Maigne, L ;
Mayet, F ;
Melot, F ;
Merheb, C ;
Pennacchio, E ;
Perez, J ;
Pietrzyk, U ;
Rannou, FR ;
Rey, M ;
Schaart, DR ;
Schmidtlein, CR ;
Simon, L ;
Song, TY ;
Vieira, JM ;
Visvikis, D .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (19) :4543-4561
[8]  
JAN S, 2002, THESIS U GRENOBLE GR
[9]   GATE: A Geant4-based simulation platform for PET and SPECT integrating movement and time management [J].
Santin, G ;
Strul, D ;
Lazaro, D ;
Simon, L ;
Krieguer, M ;
Martins, MV ;
Breton, V ;
Morel, C .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (05) :1516-1521
[10]  
SIMON L, IN PRESS P 2 ITBS C