Evaluation of the accuracy of a wireless pressure sensor in a canine model of retrograde-collateral (type II) endoleak and correlation with histologic analysis

被引:24
作者
Chaer, Rabih A.
Trocciola, Susan
DeRubertis, Brian
Hynecek, Robert
Xu, Qiang
Lam, Russell
Kent, K. Craig
Faries, Peter L.
机构
[1] Cornell Univ, New York Presbyterian Hosp, Div Vasc Surg, Weill Med Sch, New York, NY 10021 USA
[2] Columbia Univ, Coll Phys & Surg, New York, NY 10027 USA
关键词
D O I
10.1016/j.jvs.2006.08.031
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background. The utility of intra-aneurysmal pressure determination is dependent on the ability to measure pressure in the presence of endoleak and thrombosis. In this study, the accuracy of a CardioMEMS wireless pressure sensor (CardioMEMS, Atlanta, Ga) transducer in the presence of thrombus associated with type 11 endoleak was measured. Methods. Type II endoleaks were created in four mongrel dogs by implanting four collateral arterial side branches (lumbar and caudal mesenteric arteries) as a Carrel patch onto a 3-cm prosthetic polytetrafluoroethylene abdominal aortic aneurysm (AAA). The aneurysm was excluded 2 weeks later from antegrade perfusion by a stent graft. The wireless pressure sensor was positioned in the AAA external to the stent graft. A Konigsberg intraluminal solid-state strain-gauge pressure transducer (Konigsberg Instruments, Pasadena, Calif) that is accurate in the presence of thrombus served as the control to determine AAA pressure. Both of the transducers were implanted on the luminal surface of the aneurysm, 180 degrees opposite from the Carrel patch and endoleak channel. Intra-aneurysmal pressure resulting from the type II endoleak was measured twice daily for 4 weeks using both transducers. A total of 56 pre-exclusion and 224 post-exclusion distinct pressure determinations were made. Intra-aneurysmal pressure was indexed to the systemic pressure that was simultaneously measured by a strain-gauge pressure transducer implanted in the native aorta. Histologic analysis of the aneurysm contents was performed with hematoxylin and eosin. Results. The intra-aneurysmal systolic, mean, and pulse pressures produced by the type II endoleak were significantly lower than systemic pressure in all animals and were < 60% of systemic pressure (P <.001). Close correlation between the wireless transducer and the control strain-gauge transducer was observed (R = 0.83, P < .001). Arteriography and Doppler ultrasound documented retrograde flow through the aneurysm side branches and persistent endoleak patency up to the time of euthanasia. Pathologic analysis demonstrated the endoleak channel to be patent and separated from the transducers by thrombus, which surrounded both transducers. Conclusions: Intra-aneurysmal pressure generated by type II endoleaks may be accurately measured through thrombus using a wireless pressure sensor in the canine model. The wireless sensor has the potential for clinical applicability in diagnosing and characterizing type II endoleaks.
引用
收藏
页码:1306 / 1313
页数:8
相关论文
共 26 条
[1]   Aneurysm sac pressure measurements after endovascular repair of abdominal aortic aneurysms [J].
Baum, RA ;
Carpenter, RP ;
Cope, C ;
Golden, MA ;
Velazquez, OC ;
Neschis, DG ;
Mitchell, ME ;
Barker, CF ;
Fairman, RM .
JOURNAL OF VASCULAR SURGERY, 2001, 33 (01) :32-40
[2]  
Chong CK, 2003, J ENDOVASC THER, V10, P424, DOI 10.1583/1545-1550(2003)010<0424:MEACRF>2.0.CO
[3]  
2
[4]   Characterization of retrograde collateral (type II) endoleak using a new canine model [J].
Dayal, R ;
Mousa, A ;
Bernheim, J ;
Hollenbeck, S ;
Henderson, P ;
Prince, M ;
Gordon, R ;
Badimon, J ;
Fuster, V ;
Marin, ML ;
Kent, KC ;
Faries, PL .
JOURNAL OF VASCULAR SURGERY, 2004, 40 (05) :985-994
[5]   Intra-aneurysm sac pressure measurements after endovascular aneurysm repair: Differences between shrinking, unchanged, and expanding aneurysms with and without endoleaks [J].
Dias, NV ;
Ivancev, K ;
Malina, M ;
Resch, T ;
Lindblad, B ;
Sonesson, B .
JOURNAL OF VASCULAR SURGERY, 2004, 39 (06) :1229-1235
[6]  
DORMER KJ, 1980, MED ELECT, V11, P69
[7]   Abdominal aortic aneurysm sac shrinkage after endovascular aneurysm repair: Correlation with chronic sac pressure measurement [J].
Ellozy, SH ;
Carroccio, A ;
Lookstein, RA ;
Jacobs, TS ;
Addis, MD ;
Teodorescu, VJ ;
Marin, ML .
JOURNAL OF VASCULAR SURGERY, 2006, 43 (01) :2-6
[8]   First experience in human beings with a permanently implantable intrasac pressure transducer for monitoring endovascular repair of abdominal aortic aneurysms [J].
Ellozy, SH ;
Carroccio, A ;
Lookstein, RA ;
Minor, ME ;
Sheahan, CM ;
Juta, J ;
Cha, A ;
Valenzuela, R ;
Addis, MD ;
Jacobs, MS ;
Teodorescu, VJ ;
Marin, ML .
JOURNAL OF VASCULAR SURGERY, 2004, 40 (03) :405-411
[9]  
Faries PL, 1997, J ENDOVASC SURG, V4, P290, DOI 10.1583/1074-6218(1997)004<0290:AEMFTA>2.0.CO
[10]  
2