Unreliable post event report from an automated external defibrillator

被引:4
作者
Calle, PA
Monsieurs, KG
Buylaert, WA
机构
[1] State Univ Ghent Hosp, Dept Emergency Med, B-9000 Ghent, Belgium
[2] State Univ Ghent Hosp, Intens Care Unit, B-9000 Ghent, Belgium
关键词
cardiopulmonary resuscitations; defibrillation; automated external defibrillator; medicolegal problem;
D O I
10.1016/S0300-9572(01)00358-6
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Medical supervision of the use of automated external defibrillators (AEDs) is possible by the incorporation of a solid state memory system recording electrocardiography (ECG) tracings and information about the operation of the device. Since a post event report suggested inappropriate AED use erroneously, the information storage and printing processes of the Laerdal (R) AED system were investigated. This analysis strongly suggests (yet unpredictable) incompatibilities between the software built in the solid state memory modules and the different components of the printing system. Although no problems were encountered during the resuscitation attempts, these findings may be clinically relevant because an unreliable post event report from a solid state memory module may lead to erroneous criticism of an AED user, (C) 2001 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:357 / 361
页数:5
相关论文
共 10 条
[1]   When an AED meets an ICD .... [J].
Calle, PA ;
Buylaert, W .
RESUSCITATION, 1998, 38 (03) :177-183
[2]   Anomalous ECG downloads from semi-automatic external defibrillators [J].
Calle, PA ;
Vanhaute, O ;
Ranhoff, JF ;
Buylaert, WA .
RESUSCITATION, 1998, 38 (02) :119-125
[3]  
Kerber RE, 1997, CIRCULATION, V95, P1677
[4]  
*LAERD MED CORP, HEARTST 3000 OP STRU
[5]   SEMIAUTOMATIC EXTERNAL DEFIBRILLATION AND IMPLANTED CARDIAC-PACEMAKERS - UNDERSTANDING THE INTERACTIONS DURING RESUSCITATION [J].
MONSIEURS, KG ;
CONRAADS, VMA ;
GOETHALS, MPN ;
SNOECK, JP ;
BOSSAERT, LL .
RESUSCITATION, 1995, 30 (02) :127-131
[6]   Sudden death from cardiac arrest - Improving the odds. [J].
Robertson, RM .
NEW ENGLAND JOURNAL OF MEDICINE, 2000, 343 (17) :1259-1260
[7]   EFFICACY OF OUT OF HOSPITAL DEFIBRILLATION BY AMBULANCE TECHNICIANS USING AUTOMATED EXTERNAL DEFIBRILLATORS - THE HEARTSTART-SCOTLAND PROJECT [J].
SEDGWICK, ML ;
WATSON, J ;
DALZIEL, K ;
CARRINGTON, DJ ;
COBBE, SM .
RESUSCITATION, 1992, 24 (01) :73-87
[8]   States set to pass laws limiting liability for lay users of automated external defibrillators [J].
SoRelle, R .
CIRCULATION, 1999, 99 (20) :2606-2607
[9]   Quality assessment of defibrillation and advanced life support using data from the medical control module of the defibrillator [J].
Sunde, K ;
Eftestol, T ;
Askenberg, C ;
Steen, PA .
RESUSCITATION, 1999, 41 (03) :237-247
[10]   AMERICAN-HEART-ASSOCIATION REPORT ON THE PUBLIC-ACCESS DEFIBRILLATION CONFERENCE - DECEMBER 8-10, 1994 [J].
WEISFELDT, ML ;
KERBER, RE ;
MCGOLDRICK, RP ;
MOSS, AJ ;
NICHOL, G ;
ORNATO, JP ;
PALMER, DG ;
RIEGEL, B ;
SMITH, SC .
CIRCULATION, 1995, 92 (09) :2740-2747