Electrodeposited iridium oxide pH electrode for measurement of extracellular myocardial acidosis during acute ischemia

被引:194
作者
Marzouk, SAM
Ufer, S
Buck, RP [1 ]
Johnson, TA
Dunlap, LA
Cascio, WE
机构
[1] Univ N Carolina, Dept Chem, Chapel Hill, NC 27599 USA
[2] Univ N Carolina, Dept Med, Chapel Hill, NC 27599 USA
关键词
D O I
10.1021/ac980608e
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In the present paper, fabrication, characterization, and physiological applications of a solid-state pH electrode are described. The pH sensing layer was based on an anodic electrodeposited iridium oxide film (AEIROF), Sputtered platinum electrodes (1 mm diameter) fabricated on flexible Kapton films or platinum wires were used as planar or cylindrical supports. Each electrode site was coated with Nafion to attenuate the interference of anionic redox species and to protect the electrode surface during in vivo measurements. Performance of the AEIROF was evaluated, for the first time, as a pH electrode and proved to have a slightly super-Nernstian response with slope of -63.5 +/- 2.2 mV/pH unit for both wire and planar sputtered platinum electrodes. Linear pH responses were obtained in the pH range 2-10, The electrodes have a working lifetime of at least 1 month with accuracy of about 0.02 pH unit and fast response time. The electrodes showed very low sensitivities for different species, such as Na+, K+, Li+, NH4+, Ca2+, Mg2+, dissolved oxygen, lactate, ascorbate, and mate, which are important for physiological applications, The electrodes were applied in extracellular pH measurements during brief regional ischemia in a swine heart and no-flow ischemia in an isolated rabbit papillary muscle. A first report on extracellular pH, K+, and lactate simultaneous measurements during no-flow ischemia using the AEIROF pH electrode and the previously described K+ and lactate electrodes is presented as well.
引用
收藏
页码:5054 / 5061
页数:8
相关论文
共 36 条
[1]   PROPERTIES OF THERMALLY PREPARED IRIDIUM DIOXIDE ELECTRODES [J].
ARDIZZONE, S ;
CARUGATI, A ;
TRASATTI, S .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1981, 126 (1-3) :287-292
[2]   PREPARATION OF AN OXIDIZED IRIDIUM ELECTRODE AND THE VARIATION OF ITS POTENTIAL WITH PH [J].
BURKE, LD ;
MULCAHY, JK ;
WHELAN, DP .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1984, 163 (1-2) :117-128
[3]   PASSIVE ELECTRICAL-PROPERTIES, MECHANICAL-ACTIVITY, AND EXTRACELLULAR POTASSIUM IN ARTERIALLY PERFUSED AND ISCHEMIC RABBIT VENTRICULAR MUSCLE - EFFECTS OF CALCIUM ENTRY BLOCKADE OR HYPOCALCEMIA [J].
CASCIO, WE ;
YAN, GX ;
KLEBER, AG .
CIRCULATION RESEARCH, 1990, 66 (06) :1461-1473
[4]   EARLY CHANGES IN EXTRACELLULAR POTASSIUM IN ISCHEMIC RABBIT MYOCARDIUM - THE ROLE OF EXTRACELLULAR CARBON-DIOXIDE ACCUMULATION AND DIFFUSION [J].
CASCIO, WE ;
YAN, GX ;
KLEBER, AG .
CIRCULATION RESEARCH, 1992, 70 (02) :409-422
[5]   SURFACE AND BULK PROCESSES AT OXIDIZED IRIDIUM ELECTRODES .2. CONDUCTIVITY-SWITCHED BEHAVIOR OF THICK OXIDE-FILMS [J].
CONWAY, BE ;
MOZOTA, J .
ELECTROCHIMICA ACTA, 1983, 28 (01) :9-16
[6]   MICROFABRICATED SENSOR ARRAYS SENSITIVE TO PH AND K+ FOR IONIC DISTRIBUTION MEASUREMENTS IN THE BEATING HEART [J].
COSOFRET, VV ;
ERDOSY, M ;
JOHNSON, TA ;
BUCK, RP ;
ASH, RB ;
NEUMAN, MR .
ANALYTICAL CHEMISTRY, 1995, 67 (10) :1647-1653
[7]   ENHANCED OXYGEN EVOLUTION ON AN ELECTRODEPOSITED RUTHENIUM+IRIDIUM COATING ON TITANIUM [J].
CUKMAN, D ;
VUKOVIC, M ;
MILUN, M .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1995, 389 (1-2) :209-213
[8]   EMF-MEASUREMENTS OF CELLS EMPLOYING METAL-METAL OXIDE ELECTRODES IN AQUEOUS CHLORIDE AND SULFATE ELECTROLYTES AT TEMPERATURES BETWEEN 25-250DEGREESC [J].
DOBSON, JV ;
SNODIN, PR ;
THIRSK, HR .
ELECTROCHIMICA ACTA, 1976, 21 (07) :527-533
[9]   EFFECT OF SERIAL BRIEF ISCHEMIC EPISODES ON EXTRACELLULAR K+, PH, AND ACTIVATION IN THE PIG [J].
FLEET, WF ;
JOHNSON, TA ;
GRAEBNER, CA ;
GETTES, LS .
CIRCULATION, 1985, 72 (04) :922-932
[10]   ELECTRONIC SEMICONDUCTING OXIDES AS PH SENSORS [J].
FOG, A ;
BUCK, RP .
SENSORS AND ACTUATORS, 1984, 5 (02) :137-146