Determination of trace iodide in seawater by capillary electrophoresis following transient isotachophoretic preconcentration - Comparison with ion chromatography

被引:97
作者
Ito, K
Ichihara, T
Zhuo, HA
Kumamoto, K
Timerbaev, AR
Hirokawa, T
机构
[1] Kinki Univ, Sch Engn, Dept Biotechnol & Chem, Higashihiroshima 7392116, Japan
[2] Hiroshima Univ, Grad Sch Engn, Dept Appl Chem, Higashihiroshima 7398527, Japan
[3] VI Vernadskii Inst Geochem & Analyt Chem, Moscow 117975, Russia
关键词
seawater analysis; capillary electrophoresis; sample preconcentration; ion chromatography; iodide; ZONE-ELECTROPHORESIS; BACKGROUND ELECTROLYTE; ARTIFICIAL SEAWATER; INORGANIC ANIONS; SAMPLE STACKING; NATURAL-WATERS; NITRATE; NITRITE; SENSITIVITY;
D O I
10.1016/j.aca.2003.08.052
中图分类号
O65 [分析化学];
学科分类号
070302 [分析化学];
摘要
Usefulness of capillary electrophoresis (CE) in combination with transient isotachophoresis (tITP) as on-line preconcentration technique was examined for routine analysis of trace iodide in seawater. The combined method was based on an electrolyte system where 0.5 mol l(-1) sodium chloride and 25 mmol l(-1) cetyltrimethylammonium chloride at an acidic pH and 0.5 mol l(-1) 2-(N-morpholino)ethanesulfonate (pH 6.0) served as carrier and terminating electrolyte, respectively. Under optimized tITP-CE conditions (a carrier electrolyte pH of 2.4), the peak of 1 mug l(-1) iodide in artificial seawater can be clearly detected at 226 nm, with the detection limit as low as 0.2 mug l(-1). Calibration curve was linear in the range of 0-40 mug l(-1) (n = 10; regression coefficient = 0.99992). Relative standard deviation values of the migration time, peak height and area of surface seawater iodide (0.6, 3.1 and 1.5%, respectively) proved an excellent run-to-run reproducibility performance of the proposed method. The method was applied to seawater samples containing sub- and low-mug l(-1) levels of iodide, and the results obtained agreed well with ion chromatographic data. Detection of nitrate and nitrite was also possible in the same tITP-CE run (limits of detection 25 and 20 mug l(-1), respectively, at 210 nm). The analytical characteristics of the method were discussed in comparison with an ion chromatographic procedure previously developed in authors' laboratory. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:67 / 74
页数:8
相关论文
共 41 条
[1]
Beckers JL, 2000, ELECTROPHORESIS, V21, P2788, DOI 10.1002/1522-2683(20000801)21:14<2788::AID-ELPS2788>3.0.CO
[2]
2-L
[3]
Beckers JL, 2000, ELECTROPHORESIS, V21, P2747, DOI 10.1002/1522-2683(20000801)21:14<2747::AID-ELPS2747>3.0.CO
[4]
2-Z
[5]
MATRIX-ELIMINATION ION CHROMATOGRAPHY WITH POSTCOLUMN REACTION DETECTION FOR THE DETERMINATION OF IODIDE IN SALINE WATERS [J].
BRANDAO, ACM ;
BUCHBERGER, WW ;
BUTLER, ECV ;
FAGAN, PA ;
HADDAD, PR .
JOURNAL OF CHROMATOGRAPHY A, 1995, 706 (1-2) :271-275
[6]
Breadmore MC, 2001, ELECTROPHORESIS, V22, P2464
[7]
BREWER PG, 1975, CHEM OCEANOGR, P456
[8]
Direct analysis of inorganic anions in samples with high salt content by capillary zone electrophoresis [J].
Carou, MIT ;
Mahía, PL ;
Lorenzo, SM ;
Fernández, EF ;
Rodríguez, DP .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2001, 39 (09) :397-401
[9]
Fukushi K, 2000, ELECTROPHORESIS, V21, P2866, DOI 10.1002/1522-2683(20000801)21:14<2866::AID-ELPS2866>3.0.CO
[10]
2-V