Direct analysis of inorganic anions in samples with high salt content by capillary zone electrophoresis

被引:20
作者
Carou, MIT [1 ]
Mahía, PL [1 ]
Lorenzo, SM [1 ]
Fernández, EF [1 ]
Rodríguez, DP [1 ]
机构
[1] Univ A Coruna, Dept Analyt Chem, Coruna 15071, Spain
关键词
D O I
10.1093/chromsci/39.9.397
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Trace-level inorganic anions in seawater are separated efficiently by capillary zone electrophoresis using direct UV detection. The carrier electrolyte is 50mM borate at pH 9.3 and contained 1.5M NaCl. This buffer solution is adopted to prevent interference from high concentrations of the chloride ion in seawater. No electro-osmotic flow reverser is used to shorten the analysis time. The experimental conditions such as the concentration of NaCl in the carrier electrolyte, capillary inner diameter, applied current, and temperature are optimized. Linear plots are obtained in the concentration range of 0.1 to 20 μg/mL. The quantitation limits of the anions are in the order of 0.02 to 0.1 μg/mL. The proposed method may be applicable to the determination of inorganic anions in other environmental samples and effluents of a power plant.
引用
收藏
页码:397 / 401
页数:5
相关论文
共 18 条
[1]  
BUCHBERGER W, 1997, HDB CAPILLARY ELECTR, P531
[2]   Capillary zone electrophoresis for the determination of light-absorbing anions in environmental samples [J].
Carou, MIT ;
Mahía, PL ;
Lorenzo, SM ;
Fernández, EF ;
Rodríguez, DP .
JOURNAL OF CHROMATOGRAPHY A, 2001, 918 (02) :411-421
[3]   Capillary isoelectric focusing of physiologically derived proteins with on-line desalting of isotonic salt concentrations [J].
Clarke, NJ ;
Tomlinson, AJ ;
Schomburg, G ;
Naylor, S .
ANALYTICAL CHEMISTRY, 1997, 69 (14) :2786-2792
[4]   Capillary electrophoresis of anions at high salt concentrations [J].
Ding, WL ;
Thornton, MJ ;
Fritz, JS .
ELECTROPHORESIS, 1998, 19 (12) :2133-2139
[5]   Determination of bromide ions in seawater by capillary zone electrophoresis using diluted artificial seawater as the buffer solution [J].
Fukushi, K ;
Watanabe, K ;
Takeda, S ;
Wakida, SI ;
Yamane, M ;
Higashi, K ;
Hiiro, K .
JOURNAL OF CHROMATOGRAPHY A, 1998, 802 (01) :211-217
[6]  
HEIGER D, 1992, HIGH PERFORMANCE CAP
[7]   VARIOUS APPROACHES TO ANALYSIS OF DIFFICULT SAMPLE MATRICES OF ANIONS USING CAPILLARY ION ELECTROPHORESIS [J].
JONES, WR ;
JANDIK, P .
JOURNAL OF CHROMATOGRAPHY, 1992, 608 (1-2) :385-393
[8]   Capillary electrophoresis of inorganic anions [J].
Kaniansky, D ;
Masár, M ;
Marák, J ;
Bodor, R .
JOURNAL OF CHROMATOGRAPHY A, 1999, 834 (1-2) :133-178
[9]   Optimization of conductivity detection of low-molecular-mass anions in capillary zone electrophoresis [J].
Katzmayr, MU ;
Klampfl, CW ;
Buchberger, W .
JOURNAL OF CHROMATOGRAPHY A, 1999, 850 (1-2) :355-362
[10]  
LI SFY, 1993, J CHROMATOGR LIB SER, V52