Determination of uric acid and p-aminohippuric acid in human saliva and urine using capillary electrophoresis with electrochemical detection -: Potential application in fast diagnosis of renal disease

被引:41
作者
Guan, YQ [1 ]
Wu, T [1 ]
Ye, JN [1 ]
机构
[1] E China Normal Univ, Dept Chem, Shanghai 200062, Peoples R China
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2005年 / 821卷 / 02期
关键词
capillary electrophoresis; electrochemical detection; renal disease; saliva; urine; uric acid; p-aminohippuric acid;
D O I
10.1016/j.jchromb.2005.03.049
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The monitoring of uric acid (UA) and p-aminohippuric acid (PAH) levels in biological samples is routinely carried out in clinical laboratories as an indication of renal disease. With the aim of investigation of the correlation between the trace amounts of UA and PAH in human saliva or urine and renal diseases, we carried out the determination of UA and PAH in human saliva and urine by using capillary electrophoresis with electrochemical detection (CE-ED) in this work. Under the optimum conditions, UA, PAH and three coexisting analytes could be well separated within 21 min at the separation voltage of 14 kV in 80 mmol/L borax running buffer (pH 9.2). Good linear relationship was established between peak current and concentration of analytes over two orders of magnitude with detection limits (S/N = 3) ranged from 5.01 x 10(-7) to 2.00 x 10(-6) mol/L for all analytes. The result shows that this proposed method could be successfully applied for the study on the correlation between the levels of UA and PAH in human saliva and urine and renal diseases, and provide an alternative and convenient method for the fast diagnosis of renal disease. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:229 / 234
页数:6
相关论文
共 30 条
[11]   Simultaneous measurement of allantoin, uric acid, xanthine and hypoxanthine in blood by high-performance liquid chromatography [J].
Czauderna, M ;
Kowalczyk, J .
JOURNAL OF CHROMATOGRAPHY B, 1997, 704 (1-2) :89-98
[12]   A COLORIMETRIC METHOD USING URICASE AND PEROXIDASE FOR DETERMINATION OF URIC ACID [J].
DOMAGK, GF ;
SCHLICKE, HH .
ANALYTICAL BIOCHEMISTRY, 1968, 22 (02) :219-&
[13]   Simultaneous determination of p-aminohippuric acid, acetyl-p-aminohippuric acid and iothalamate in human plasma and urine by high-performance liquid chromatography [J].
Dowling, TC ;
Frye, RF ;
Zemaitis, MA .
JOURNAL OF CHROMATOGRAPHY B, 1998, 716 (1-2) :305-313
[14]  
Dryhurst G., 1977, ELECTROCHEMISTRY BIO
[15]  
Fanguy JC, 2002, ELECTROPHORESIS, V23, P767, DOI 10.1002/1522-2683(200203)23:5<767::AID-ELPS767>3.0.CO
[16]  
2-8
[17]   Simultaneous determination of the urinary metabolites of toluene, xylene and styrene using high-performance capillary electrophoresis - Comparison with high-performance liquid chromatography [J].
Fujii, T ;
Kawabe, S ;
Horike, T ;
Taguchi, T ;
Ogata, M .
JOURNAL OF CHROMATOGRAPHY B, 1999, 730 (01) :41-47
[18]   Direct detection of renal function markers using microchip CE with pulsed electrochemical detection [J].
Garcia, CD ;
Henry, CS .
ANALYST, 2004, 129 (07) :579-584
[19]  
KIMMEL PL, 1993, J AM SOC NEPHROL, V4, P12
[20]   Psychosocial factors in dialysis patients [J].
Kimmel, PL ;
Levy, N ;
Vassalotti, J ;
Christensen, A ;
Friend, R ;
Veis, J ;
Peterson, R ;
Devins, G ;
Lew, SQ ;
Star, R ;
Reiss, D .
KIDNEY INTERNATIONAL, 2001, 59 (04) :1599-1613