Accurate measurement of total attenuation coefficient of thin tissue with optical coherence tomography

被引:43
作者
Kholodnykh, AI [1 ]
Petrova, IY
Motamedi, M
Esenaliev, RO
机构
[1] Univ Texas, Med Branch, Ctr Biomed Engn, Galveston, TX 77555 USA
[2] Univ Texas, Med Branch, Dept Ophthalmol & Visual Sci, Galveston, TX 77555 USA
[3] Univ Texas, Med Branch, Dept Physiol & Biophys, Galveston, TX 77555 USA
[4] Univ Texas, Med Branch, Dept Anesthesiol, Galveston, TX 77555 USA
基金
美国国家卫生研究院;
关键词
medical diagnostics; noninvasive monitoring; optical coherence tomography (OCT); tissue optical properties;
D O I
10.1109/JSTQE.2003.814194
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Noninvasive accurate measurements of tissue optical properties are needed for many diagnostic and therapeutic applications. Optical coherence tomography (OCT) recently proposed for high-resolution imaging in tissue can potentially be applied for accurate, noninvasive, and high-resolution measurement of tissue total attenuation coefficient. However, confocal function (dependence of OCT sensitivity on the distance of probed site from the focal plane of the objective lens) and multiple scattering substantially limit the accuracy of the measurement with the OCT technique. We studied the influence of the confocal function and multiple scattering on the accuracy of the measurement and proposed methods that provide measurement of the total attenuation coefficient with a significantly reduced systematic error. Experiments were performed in tissue phantoms and porcine and human skin in vitro and in vivo. Our data indicate that the tissue total attenuation coefficient can noninvasively be measured in vivo with the accuracy of 5%-10% in the range from 0.5 to 17 mm(-1) and about 20% in the range up to 40 mm(-1). These results suggest that the proper correction of the OCT-based measurement for the confocal function and multiple scattering provides absolute values of tissue total attenuation coefficient with high accuracy and resolution that may not be achievable by other optical techniques in vivo.
引用
收藏
页码:210 / 221
页数:12
相关论文
共 26 条
[1]  
[Anonymous], OPT THERM RESPONSE L
[2]   A REVIEW OF THE OPTICAL-PROPERTIES OF BIOLOGICAL TISSUES [J].
CHEONG, WF ;
PRAHL, SA ;
WELCH, AJ .
IEEE JOURNAL OF QUANTUM ELECTRONICS, 1990, 26 (12) :2166-2185
[3]   TOTAL ATTENUATION COEFFICIENTS AND SCATTERING PHASE FUNCTIONS OF TISSUES AND PHANTOM MATERIALS AT 633-NM [J].
FLOCK, ST ;
WILSON, BC ;
PATTERSON, MS .
MEDICAL PHYSICS, 1987, 14 (05) :835-841
[4]   Optical coherence tomography: An emerging technology for biomedical imaging and optical biopsy [J].
Fujimoto, JG ;
Pitris, C ;
Boppart, SA ;
Brezinski, ME .
NEOPLASIA, 2000, 2 (1-2) :9-25
[5]   SOME FUNDAMENTAL PROPERTIES OF SPECKLE [J].
GOODMAN, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1976, 66 (11) :1145-1150
[6]   OPTICAL-PROPERTIES OF HUMAN DERMIS INVITRO AND INVIVO [J].
GRAAFF, R ;
DASSEL, ACM ;
KOELINK, MH ;
DEMUL, FFM ;
AARNOUDSE, JG ;
ZIJLSTRA, WG .
APPLIED OPTICS, 1993, 32 (04) :435-447
[7]   OPTICAL-CONSTANTS OF WATER IN 200-NM TO 200-MUM WAVELENGTH REGION [J].
HALE, GM ;
QUERRY, MR .
APPLIED OPTICS, 1973, 12 (03) :555-563
[8]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[9]   OPTICAL COHERENCE MICROSCOPY IN SCATTERING MEDIA [J].
IZATT, JA ;
HEE, MR ;
OWEN, GM ;
SWANSON, EA ;
FUJIMOTO, JG .
OPTICS LETTERS, 1994, 19 (08) :590-592
[10]  
Jacques S. L., 1987, Lasers in the Life Sciences, V1, P309