Statistics for critical clinical decision making based on readings of pairs of implanted sensors

被引:19
作者
Schmidtke, DW [1 ]
Pishko, MV [1 ]
Quinn, CP [1 ]
Heller, A [1 ]
机构
[1] UNIV TEXAS,DEPT CHEM ENGN,AUSTIN,TX 78712
关键词
D O I
10.1021/ac9602027
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Low error rates are essential if lives of patients are to depend on readings of implanted sensors, such as glucose sensors in insulin-dependent diabetic patients, To verify the operation and to calibrate on demand an implanted sensor, it is necessary that calibration through a single, independent measurement involving withdrawal of only one sample of blood and its independent analysis be feasible, Such a one-point calibration must be accurate, Borrowing from nuclear reactor safety assurance, where a Likelihood ratio test is applied to readings of pairs of pressure sensors for shutdown/no shutdown decisions, we apply a similar test to sensor pairs implanted in rats, We show, for five sets of glucose sensor pairs, calibrated in vivo by withdrawal of a single sample of blood, that application of the likelihood ratio test increases the fraction of the clinically correct readings from 92.4% for their averaged readings to 98.8%.
引用
收藏
页码:2845 / 2849
页数:5
相关论文
共 9 条
[1]  
[Anonymous], 1987, DIABETES CARE, V10, P95
[2]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[3]  
CSOREGI E, 1994, ANAL CHEM, V66, P3131
[4]   SEQUENTIAL PROBABILITY RATIO TEST FOR NUCLEAR-PLANT COMPONENT SURVEILLANCE [J].
GROSS, KC ;
HUMENIK, KE .
NUCLEAR TECHNOLOGY, 1991, 93 (02) :131-137
[5]   ELIMINATION OF ELECTROOXIDIZABLE INTERFERANTS IN GLUCOSE ELECTRODES [J].
MAIDAN, R ;
HELLER, A .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1991, 113 (23) :9003-9004
[6]   ELIMINATION OF ELECTROOXIDIZABLE INTERFERANT-PRODUCED CURRENTS IN AMPEROMETRIC BIOSENSORS [J].
MAIDAN, R ;
HELLER, A .
ANALYTICAL CHEMISTRY, 1992, 64 (23) :2889-2896
[7]   GLUCOSE ELECTRODES BASED ON CROSS-LINKED [OS(BPY)(2)](+/2+) COMPLEXED POLY(L-VINYLIMIDAZOLE) FILMS [J].
OHARA, TJ ;
RAJAGOPALAN, R ;
HELLER, A .
ANALYTICAL CHEMISTRY, 1993, 65 (23) :3512-3517
[8]   PHOTO-CROSS-LINKED COPOLYMERS OF 2-HYDROXYETHYL METHACRYLATE, POLY(ETHYLENE GLYCOL) TETRA-ACRYLATE AND ETHYLENE DIMETHACRYLATE FOR IMPROVING BIOCOMPATIBILITY OF BIOSENSORS [J].
QUINN, CP ;
PATHAK, CP ;
HELLER, A ;
HUBBELL, JA .
BIOMATERIALS, 1995, 16 (05) :389-396
[9]   OPTIMUM CHARACTER OF THE SEQUENTIAL PROBABILITY RATIO TEST [J].
WALD, A ;
WOLFOWITZ, J .
ANNALS OF MATHEMATICAL STATISTICS, 1948, 19 (03) :326-339