Cragg covered stents in hemodialysis access: Initial and midterm results

被引:76
作者
Sapoval, MR
TurmelRodrigues, LA
Raynaud, AC
Bourquelot, P
Rodrigue, H
Gaux, JC
机构
[1] CLIN ST GATIEN, DEPT CARDIOVASC RADIOL, TOURS, FRANCE
[2] CLIN LABROUSTE, DEPT CARDIOVASC RADIOL, PARIS, FRANCE
[3] HOP ST JOSEPH, DEPT UROL, F-75674 PARIS, FRANCE
[4] CLIN ST COME, DEPT NEPHROL, BLOIS, FRANCE
关键词
D O I
10.1016/S1051-0443(96)72863-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To report midterm follow-up after implantation of covered stents for hemodialysis access. PATIENTS AND METHODS: Over a 2-year period, a Cragg Endopro stent was placed in 14 patients (mean age, 66.6 years +/- 15) to treat angioplasty-induced ruptures (n = 3), pseudoaneurysm (n = 1), postangioplasty residual stenosis (n = 2), and early restenosis (n = 8, four of them in a Wallstent). RESULTS: Initial placement was successful in all cases. A clinical inflammatory reaction was observed in all three cases of placement in the forearm. When the covered stent was placed in a stenotic vessel, restenosis always occurred within 6 months. Primary and secondary patencies were 28.5% +/- 13.9 and 67.8% +/- 14.5, respectively, at 6 months, Covered stents were of undoubtable benefit in one case of rupture after Wallstent failure and in one case of restenosis in a Wallstent. CONCLUSION: Covered Cragg stents are effective in controlling angioplasty-induced rupture and sometimes for maintaining patency after restenosis in a Wallstent. They do not prevent restenosis and are responsible for an inflammatory reaction of unknown origin and long-term effect.
引用
收藏
页码:335 / 342
页数:8
相关论文
共 33 条
  • [11] INSUFFICIENT HEMODIALYSIS ACCESS FISTULAS - LATE RESULTS OF TREATMENT WITH PERCUTANEOUS BALLOON ANGIOPLASTY
    GMELIN, E
    WINTERHOFF, R
    RINAST, E
    [J]. RADIOLOGY, 1989, 171 (03) : 657 - 660
  • [12] HEALING OF POLYTETRAFLUOROETHYLENE ARTERIAL GRAFTS IS INFLUENCED BY GRAFT POROSITY
    GOLDEN, MA
    HANSON, SR
    KIRKMAN, TR
    SCHNEIDER, PA
    CLOWES, AW
    [J]. JOURNAL OF VASCULAR SURGERY, 1990, 11 (06) : 838 - 845
  • [13] USE OF WALLSTENTS FOR HEMODIALYSIS ACCESS-RELATED VENOUS STENOSES AND OCCLUSIONS UNTREATABLE WITH BALLOON ANGIOPLASTY
    GRAY, RJ
    HORTON, KM
    DOLMATCH, BL
    RUNDBACK, JH
    ANAISE, D
    AQUINO, AO
    CURRIER, CB
    LIGHT, JA
    SASAKI, TM
    [J]. RADIOLOGY, 1995, 195 (02) : 479 - 484
  • [14] MACROPHAGE BIOMATERIAL INTERACTIONS - THE STIMULATION OF ENDOTHELIALIZATION
    GREISLER, HP
    DENNIS, JW
    ENDEAN, ED
    ELLINGER, J
    FRIESEL, R
    BURGESS, W
    [J]. JOURNAL OF VASCULAR SURGERY, 1989, 9 (04) : 588 - 593
  • [15] GUNTHER RW, 1989, RADIOLOGY, V170, P401
  • [16] Henry M, 1994, J Endovasc Surg, V1, P31
  • [17] CORRECTION OF CENTRAL VENOUS STENOSES - USE OF ANGIOPLASTY AND VASCULAR WALLSTENTS
    KOVALIK, EC
    NEWMAN, GE
    SUHOCKI, P
    KNELSON, M
    SCHWAB, SJ
    [J]. KIDNEY INTERNATIONAL, 1994, 45 (04) : 1177 - 1181
  • [18] EARLY INFLAMMATORY RESPONSE TO GELATIN-SEALED AND COLLAGEN-SEALED DACRON PROSTHESES
    LACROIX, H
    BOEL, K
    NEVELSTEEN, A
    SUY, R
    [J]. ANNALS OF VASCULAR SURGERY, 1995, 9 (02) : 152 - 154
  • [19] LINDENAUER SM, 1989, VASCULAR SURG, P450
  • [20] PERCUTANEOUS ILIAC ARTERY STENT - ANGIOGRAPHIC LONG-TERM FOLLOW-UP
    LONG, AL
    PAGE, PE
    RAYNAUD, AC
    BEYSSEN, BM
    FIESSINGER, JN
    DUCIMETIERE, P
    RELLAND, JY
    GAUX, JC
    [J]. RADIOLOGY, 1991, 180 (03) : 771 - 778