An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery

被引:120
作者
de Wilde, R. B. P.
Schreuder, J. J.
van den Berg, P. C. M.
Jansen, J. R. C.
机构
[1] Leiden Univ, Med Ctr, Dept Intens Care, NL-2300 RC Leiden, Netherlands
[2] Hosp San Raffaele, Dept Cardiac Surg, I-20132 Milan, Italy
关键词
D O I
10.1111/j.1365-2044.2007.05135.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The bias, precision and tracking ability of five different pulse contour methods were evaluated by simultaneous comparison of cardiac output values from the conventional thermodilution technique (COtd). The five different pulse contour methods included in this study were: Wesseling's method (cZ); the Modelflow method; the LiDCO system; the PiCCO system and a recently developed Hemac method. We studied 24 cardiac surgery patients undergoing uncomplicated coronary artery bypass grafting. In each patient, the first series of COtd was used to calibrate the five pulse contour methods. In all, 199 series of measurements were accepted by all methods and included in the study. COtd ranged from 2.14 to 7.55 l.min(-1), with a mean of 4.81 l.min(-1). Bland-Altman analysis showed the following bias and limits of agreement: cZ, 0.23 and - 0.80 to 1.26 l.min(-1); Modelflow, 0.00 and - 0.74 to 0.74 l.min(-1); LiDCO, - 0.17 and - 1.55 to 1.20 l.min(-1); PiCCO, 0.14 and - 1.60 to 1.89 l.min(-1); and Hemac, 0.06 and - 0.81 to 0.93 l.min(-1). Changes in cardiac output larger than 0.5 l.min(-1) (10%) were correctly followed by the Modelflow and the Hemac method in 96% of cases. In this group of subjects, without congestive heart failure, with normal heart rhythm and reasonable peripheral circulation, the best results in absolute values as well as in tracking changes in cardiac output were measured using the Modelflow and Hemac pulse contour methods, based on non-linear three-element Windkessel models.
引用
收藏
页码:760 / 768
页数:9
相关论文
共 30 条
  • [1] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [2] A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques
    Critchley, LAH
    Critchley, JAJH
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) : 85 - 91
  • [3] Monitoring cardiac output using the femoral and radial arterial pressure waveform
    de Wilde, R. B. P.
    Breukers, R. B. G. E.
    van den Berg, P. C. M.
    Jansen, J. R. C.
    [J]. ANAESTHESIA, 2006, 61 (08) : 743 - 746
  • [4] Della Rocca G, 2003, CAN J ANAESTH, V50, P707
  • [5] Continuous and intermittent cardiac output measurement:: pulmonary artery catheter versus aortic transpulmonary technique
    Della Rocca, G
    Costa, MG
    Pompei, L
    Coccia, C
    Pietropaoli, P
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2002, 88 (03) : 350 - 356
  • [6] Lithium dilution cardiac output measurement: A clinical assessment of central venous and peripheral venous indicator injection
    Garcia-Rodriguez, C
    Pittman, J
    Cassell, CH
    Sum-Ping, J
    El-Moalem, H
    Young, C
    Mark, JB
    [J]. CRITICAL CARE MEDICINE, 2002, 30 (10) : 2199 - 2204
  • [7] Continuous, less invasive, hemodynamic monitoring in intensive care after cardiac surgery
    Godje, O
    Hoke, K
    Lamm, P
    Schmitz, C
    Thiel, C
    Weinert, M
    Reichart, B
    [J]. THORACIC AND CARDIOVASCULAR SURGEON, 1998, 46 (04) : 242 - 249
  • [8] Godje O, 2001, Med Sci Monit, V7, P1344
  • [9] Gödje O, 2002, CRIT CARE MED, V30, P52
  • [10] CONTINUOUS NONINVASIVE CARDIAC-OUTPUT AS ESTIMATED FROM THE PULSE CONTOUR CURVE
    GRATZ, I
    KRAIDIN, J
    JACOBI, AG
    DECASTRO, NG
    SPAGNA, P
    LARIJANI, GE
    [J]. JOURNAL OF CLINICAL MONITORING, 1992, 8 (01): : 20 - 27