In vitro performance of three combinations of spacers and pressurized metered dose inhalers for treatment in children

被引:50
作者
Berg, E
Madsen, J
Bisgaard, H [1 ]
机构
[1] Natl Univ Hosp, Rigshosp, Pulm Serv, Dept Pediat, DK-2100 Copenhagen, Denmark
[2] Astro Draco AB, Analyt Chem, Pharmaceut & Analyt R&D, Lund, Sweden
[3] Univ Copenhagen, Inst Med Physiol, Copenhagen, Denmark
关键词
budesonide; children; electrostatic charge; fluticasone propionate; in vitro testing; spacer devices;
D O I
10.1183/09031936.98.12020472
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
The performance of pressurized metered dose inhalers (pMDIs) and spacers in correct dose recommendations is important, but limited information on dose delivery and fine-particle dose from different combinations of spacers and pMDIs is available. In this study, three combinations of spacers and pMDIs were investigated: NebuChamber(R) and AeroChamber(R) with budesonide pMDI and Babyhaler(R) with fluticasone propionate pMDI. Doses were withdrawn onto a filter either with a breathing simulator (dose to ventilator) or with constant flow (maximal dose). The fine-particle dose was assessed with a cascade impactor (Andersen Sampler). The effect of repeated use and cleaning of the spacers on the passive fallout of aerosol within the spacers was determined by evacuating the dose on a filter 2, 5, 10 and 30 s after actuating the spray. The drugs were quantified by liquid chromatography. The NebuChamber delivered the highest doses, both maximal dose and dose to ventilator. The recovered doses (mean+/-SD) were 55+/-6% and 51+/-2%, respectively, of the delivered dose from the pMDI. The corresponding results for the Babyhaler mere 41+/-7% and 24+/-4% and for the Aerochamber 27+/-3% and 17+/-3%. The passive fallout of aerosol, determined as half-life (t1/2) was around similar to 30 s for the NebuChamber, 9-15 s for the Babyhaler and similar to 10 s for the AeroChamber. The present study confirms that there are significant differences in dose output from different combinations of pressurized metered dose inhalers and spacers, with the NebuChamber giving the highest dose, both as delivered dose and in droplets <4.7 mu m. Interactions with the spacer material, dead space in the inspiratory line and entrainment of air during inhalation due to inefficient valve control could account for these differences.
引用
收藏
页码:472 / 476
页数:5
相关论文
共 10 条
[1]  
[Anonymous], 1966, HEALTH PHYS, V12, P173
[2]   THE EFFECT OF DELAY, MULTIPLE ACTUATIONS AND SPACER STATIC CHARGE ON THE IN-VITRO DELIVERY OF BUDESONIDE FROM THE NEBUHALER [J].
BARRY, PW ;
OCALLAGHAN, C .
BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 1995, 40 (01) :76-78
[3]  
BISGAARD H, 1995, EUR RESPIR J, V8, P856
[4]   Delivery of inhaled medication to children [J].
Bisgaard, H .
JOURNAL OF ASTHMA, 1997, 34 (06) :443-467
[5]   A NONELECTROSTATIC SPACER FOR AEROSOL DELIVERY [J].
BISGAARD, H ;
ANHOJ, J ;
KLUG, B ;
BERG, E .
ARCHIVES OF DISEASE IN CHILDHOOD, 1995, 73 (03) :226-230
[6]  
Bisgaard H, 1996, BMJ-BRIT MED J, V313, P895
[7]   DELIVERY OF SALBUTAMOL BY METERED-DOSE INHALER AND VALVED SPACER TO WHEEZY INFANTS - EFFECT ON BRONCHIAL RESPONSIVENESS [J].
CLARKE, JR ;
ASTON, H ;
SILVERMAN, M .
ARCHIVES OF DISEASE IN CHILDHOOD, 1993, 69 (01) :125-129
[8]   RELIABLE SALBUTAMOL ADMINISTRATION IN 6-MONTH-OLD TO 36-MONTH-OLD CHILDREN BY MEANS OF A METERED DOSE INHALER AND AEROCHAMBER WITH MASK [J].
CONNER, WT ;
DOLOVICH, MB ;
FRAME, RA ;
NEWHOUSE, MT .
PEDIATRIC PULMONOLOGY, 1989, 6 (04) :263-267
[9]   SHORT-TERM EFFECT OF ALBUTEROL, DELIVERED VIA A NEW AUXILIARY DEVICE, IN WHEEZY INFANTS [J].
KRAEMER, R ;
FREY, U ;
SOMMER, CW ;
RUSSI, E .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1991, 144 (02) :347-351
[10]   NEW PERSPECTIVES ON INHALED DRUG-DELIVERY AND SYSTEMIC BIOACTIVITY [J].
LIPWORTH, BJ .
THORAX, 1995, 50 (02) :105-110