Whole Blood Metal Ion Concentrations in Correlation With Activity Level in Three Different Metal-On-Metal Bearings

被引:36
作者
Pattyn, Christophe A. [1 ]
Lauwagie, Sofie N. [1 ]
Verdonk, Rene C. [1 ]
机构
[1] Ghent Univ Hosp, Dept Traumat & Orthopaed Surg, B-9000 Ghent, Belgium
关键词
hip arthroplasty; hip resurfacing; metal-on-metal; metal ion release; activity level; TOTAL HIP-REPLACEMENT; HIGH-CARBON CONTENT; IN-VIVO; COBALT; ARTHROPLASTY; CHROMIUM; INCLINATION; WEAR;
D O I
10.1016/j.arth.2009.11.007
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
Seventy consecutive osteoarthritis patients younger than 65 years undergoing a hip arthroplasty were included. Three different metal bearings were used (2 different resurfacing systems and one 28-mm metal bearing) and compared with a ceramic control group. Whole blood cobalt and chromium levels were analyzed preoperatively, at 3 months, 6 months, 1 year, and 2 years. Patients with steep cups (abduction >55 degrees) were excluded. No significant differences in functional scores and activity level were found between either groups. Ion levels were significantly lower with one type of resurfacing. No correlation was found between ion and activity level. In conclusion, although the current second-generation metal bearings may show differences in ion release, more attention should be paid to the correlation between ion release and implant positioning.
引用
收藏
页码:58 / 64
页数:7
相关论文
共 28 条
[1]
Predictive role of the Λ ratio in the evaluation of metal-on-metal total hip replacement [J].
Affatato, S. ;
Spinelli, M. ;
Zavalloni, M. ;
Leardini, W. ;
Viceconti, M. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2008, 222 (H5) :617-628
[2]
THE MCKEE-FARRAR HIP-ARTHROPLASTY - A LONG-TERM STUDY [J].
AUGUST, AC ;
ALDAM, CH ;
PYNSENT, PB .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1986, 68 (04) :520-527
[3]
BEAULE PE, 2004, W ORTH ASS 68 ANN M
[4]
Cup inclination and serum concentration of cobalt and chromium after metal-on-metal total hip arthroplasty [J].
Brodner, W ;
Grübl, A ;
Jankovsky, R ;
Meisinger, V ;
Lehr, S ;
Gottsauner-Wolf, F .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :66-70
[5]
The effect of the diameter of metal-on-metal bearings on systemic exposure to cobalt and chromium [J].
Daniel, J ;
Ziaee, H ;
Salama, A ;
Pradhan, C ;
McMinn, DJW .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2006, 88B (04) :443-448
[6]
DAUBIGNE RM, 1953, ACTA ORTHOP BELG, V19, P81
[7]
Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement [J].
De Haan, R. ;
Pattyn, C. ;
Gill, H. S. ;
Murray, D. W. ;
Campbell, P. A. ;
De Smet, K. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (10) :1291-1297
[8]
THE INCIDENCE OF CANCER FOLLOWING TOTAL HIP-REPLACEMENT [J].
GILLESPIE, WJ ;
FRAMPTON, CMA ;
HENDERSON, RJ ;
RYAN, PM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1988, 70 (04) :539-542
[9]
Head replacement, head rotation, and surface damage effects on metal-on-metal total hip replacements: a hip simulator study [J].
Hardaker, C ;
Dowson, D ;
Isaac, GH .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) :209-217
[10]
HARRIS H., 1969, J BONE JOINT SURG AM, V51-A, P737