How should airways resistance be measured in young children: mask or mouthpiece?

被引:19
作者
Child, F
Clayton, S
Davies, S
Fryer, AA
Jones, PW
Lenney, W
机构
[1] City Gen Hosp, Acad Dept Paediat, N Staffordshire Royal Infirm, Stoke On Trent ST4 6QG, Staffs, England
[2] Univ Keele, Ctr Cell & Mol Med, N Staffordshire Royal Infirm, Stoke On Trent, Staffs, England
[3] Univ Keele, Dept Math, Keele ST5 5BG, Staffs, England
关键词
airway resistance child; interrupter technique; masks; reproducibility of results; respiratory function tests;
D O I
10.1183/09031936.01.00089501
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
The reproducibility and acceptability of airways resistance measurements using the interrupter technique (MicroRint) obtained using a mouthpiece were compared with those using a face mask. Fifty children aged 4-7 yrs performed four sets of six Rint measurements; tno using a mouthpiece and two using a face mask with integral mouthpiece. Complete data were obtained from 45 (90%) children using the mouthpiece and 43 (86%) children using the mask. The tno methods were equally repeatable with comparable intraclass correlation coefficients (ICC) and coefficients of variation. Mean Rint values obtained using the mouthpiece were significantly lo,rer than those using the face mask ((mean +/- SD) mouthpiece = 0.81 +/-0.18 kPa.L-1.s, mask = 0.88 +/-0.24 kPa.L-1.s p = 0.0002), Although the mean paired differences between the tno methods were small (0.07 kPa.L-1.s), the ICC and limits of agreement confirmed that the two methods could not be used interchangeably. Sixty-seven per cent of children preferred the face mask but this was more time-consuming (p = 0.03), Children did not produce more repeatable results using their preferred method, nor did they improve with practice. Repeatable airway resistance measurements using the interrupter technique ran be obtained from young children using either a mouthpiece or a face mask, but there are significant clinical and statistical differences between the results obtained.
引用
收藏
页码:1244 / 1249
页数:6
相关论文
共 31 条
[11]  
Klug B, 1998, PEDIATR PULM, V25, P322, DOI 10.1002/(SICI)1099-0496(199805)25:5<322::AID-PPUL6>3.0.CO
[12]  
2-K
[13]  
Klug B, 1996, PEDIATR PULM, V21, P290
[14]   Measurement of the specific airway resistance by plethysmography in young children accompanied by an adult [J].
Klug, B ;
Bisgaard, H .
EUROPEAN RESPIRATORY JOURNAL, 1997, 10 (07) :1599-1605
[15]  
KLUG B, 1999, EUR RESPIR J, V14, pA167
[16]  
MERKUS PJ, 2000, AM J RESP CRIT CARE, V161, pA221
[17]   THE ACCURACY OF PORTABLE PEAK FLOW METERS [J].
MILLER, MR ;
DICKINSON, SA ;
HITCHINGS, DJ .
THORAX, 1992, 47 (11) :904-909
[18]   PEAK EXPIRATORY FLOW RATES IN CHILDREN UNDER 5 YEARS OF AGE [J].
MILNER, AD ;
INGRAM, D .
ARCHIVES OF DISEASE IN CHILDHOOD, 1970, 45 (244) :780-+
[19]  
Oswald-Mammosser M, 2000, PEDIATR PULM, V29, P213, DOI 10.1002/(SICI)1099-0496(200003)29:3<213::AID-PPUL10>3.0.CO
[20]  
2-N