Bone Remodeling Around Porous Metal Cementless Acetabular Components

被引:125
作者
Meneghini, R. Michael [1 ]
Ford, Kerry S. [3 ]
McCollough, Cynthia H. [2 ]
Hanssen, Arlen D. [3 ]
Lewallen, David G. [3 ]
机构
[1] Univ Connecticut, Ctr Hlth, Dept Orthopaed Surg, New England Musculoskeletal Inst, Farmington, CT 06034 USA
[2] Mayo Clin, Dept Radiol, Rochester, MN USA
[3] Mayo Clin, Dept Orthopaed Surg, Rochester, MN USA
关键词
bone mineral density; total hip arthroplasty; cementless; quantitative computed tomography; porous tantalum; bone-remodeling; TOTAL HIP-ARTHROPLASTY; X-RAY ABSORPTIOMETRY; MINERAL DENSITY; PROXIMAL FEMUR; PRESS-FIT; REPLACEMENT; CUPS; DEXA;
D O I
10.1016/j.arth.2009.04.025
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
Bone remodeling around cementless acetabular components after total hip arthroplasty has not been well characterized. A randomized, prospective study of total hip arthroplasty was performed comparing 2 cementless acetabular implants: a solid titanium and a more elastic porous tantalum design. Seventeen hips (9 porous tantalum, 8 titanium) underwent quantitative computed tomography at mean of 7.7 years, and adjacent bone mineral density (BMD) was calculated. The absolute and relative decrease in BMD from preoperative level was less in zones 9 to 15 mm adjacent to the porous tantalum compared to the titanium component (P <= .02) and predominated posterosuperiorly. The relative BMD increased in all regions adjacent to the porous tantalum component from 5% to 40% over the control. This data demonstrates stress-shielding likely occurs less around a highly porous metal implant of material with an elastic modulus similar to bone.
引用
收藏
页码:741 / 747
页数:7
相关论文
共 23 条
[1]
AN APPROACH FOR TIME-DEPENDENT BONE MODELING AND REMODELING - THEORETICAL DEVELOPMENT [J].
BEAUPRE, GS ;
ORR, TE ;
CARTER, DR .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1990, 8 (05) :651-661
[2]
Characteristics of bone ingrowth and interface mechanics of a new porous tantalum biomaterial [J].
Bobyn, JD ;
Stackpool, GJ ;
Hacking, SA ;
Tanzer, M ;
Krygier, JJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (05) :907-914
[3]
Brown T, 1999, T ORTHOP RES SOC, V24, P747
[4]
TRABECULAR BONE-DENSITY AND LOADING HISTORY - REGULATION OF CONNECTIVE-TISSUE BIOLOGY BY MECHANICAL ENERGY [J].
CARTER, DR ;
FYHRIE, DP ;
WHALEN, RT .
JOURNAL OF BIOMECHANICS, 1987, 20 (08) :785-+
[5]
Cohen Robert, 2002, Am J Orthop (Belle Mead NJ), V31, P216
[6]
Different loss of BMD using uncemented press-fit and whole polyethylene cups fixed with cement -: Repeated DXA studies in 96 hips randomized to 3 types of fixation [J].
Digas, G ;
Kärrholm, J ;
Thanner, J .
ACTA ORTHOPAEDICA, 2006, 77 (02) :218-226
[7]
ENGH CA, 1992, CLIN ORTHOP RELAT R, P13
[8]
A QUANTITATIVE-EVALUATION OF PERIPROSTHETIC BONE-REMODELING AFTER CEMENTLESS TOTAL HIP-ARTHROPLASTY [J].
ENGH, CA ;
MCGOVERN, TF ;
BOBYN, JD ;
HARRIS, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (07) :1009-1020
[9]
Clinical consequences of stress shielding after porous-coated total hip arthroplasty [J].
Engh, CA ;
Young, AM ;
Engh, CA ;
Hopper, RH .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (417) :157-163
[10]
ENGH CA, 1994, CLIN ORTHOP RELAT R, P89