Voltage-addressable on/off microvalves for high-pressure microchip separations

被引:50
作者
Kirby, BJ [1 ]
Shepodd, TJ [1 ]
Hasselbrink, EF [1 ]
机构
[1] Sandia Natl Labs, Microfluid Dept, Livermore, CA 94550 USA
关键词
chip technology; instrumentation;
D O I
10.1016/S0021-9673(02)01453-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We present a microchip-based, voltage-addressable on/off valve architecture that is fundamentally consistent with the pressures and solvents employed for high-pressure liquid chromatography. Laser photopatterning of polymer monoliths inside glass microchannels is used to fabricate mobile fluid control elements, which are opened and closed by electrokinetic pressures. The glass substrates and crosslinked polymer monoliths operate in water-acetonitrile mixtures and have been shown to hold off pressures as high as 350 bar (5000 p.s.i.). Open/closed flow ratios of 104 to 106 have been demonstrated over the pressure range 1.5-70 bar (20-1000 p.s.i.), and the pressure-leak relationship shows the potential for valving control of flow through packed or monolithic chromatography columns. We expect that this valve platform will enable multiplexing of multiple chromatographic separations on single microchips. Published by Elsevier Science B.V.
引用
收藏
页码:147 / 154
页数:8
相关论文
共 10 条
[1]   Functional hydrogel structures for autonomous flow control inside microfluidic channels [J].
Beebe, DJ ;
Moore, JS ;
Bauer, JM ;
Yu, Q ;
Liu, RH ;
Devadoss, C ;
Jo, BH .
NATURE, 2000, 404 (6778) :588-+
[2]   High-pressure microfluidic control in lab-on-a-chip devices using mobile polymer monoliths [J].
Hasselbrink, EF ;
Shepodd, TJ ;
Rehm, JE .
ANALYTICAL CHEMISTRY, 2002, 74 (19) :4913-4918
[3]  
Madou M., 1997, Fundamentals of Microfabrication
[4]   Conduct-as-cast polymer monoliths as separation media for capillary electrochromatography [J].
Ngola, SM ;
Fintschenko, Y ;
Choi, WY ;
Shepodd, TJ .
ANALYTICAL CHEMISTRY, 2001, 73 (05) :849-856
[5]  
PAUL PH, 2000, MICRO TOTAL ANAL SYS
[6]   MICROFLOW DEVICES AND SYSTEMS [J].
SHOJI, S ;
ESASHI, M .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 1994, 4 (04) :157-171
[7]   Electrochromatography in microchips: Reversed-phase separation of peptides and amino acids using photopatterned rigid polymer monoliths [J].
Throckmorton, DJ ;
Shepodd, TJ ;
Singh, AK .
ANALYTICAL CHEMISTRY, 2002, 74 (04) :784-789
[8]   Monolithic microfabricated valves and pumps by multilayer soft lithography [J].
Unger, MA ;
Chou, HP ;
Thorsen, T ;
Scherer, A ;
Quake, SR .
SCIENCE, 2000, 288 (5463) :113-116
[9]   Soft lithography in biology and biochemistry [J].
Whitesides, GM ;
Ostuni, E ;
Takayama, S ;
Jiang, XY ;
Ingber, DE .
ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 2001, 3 :335-373
[10]   Fabrication and characterization of electroosmotic micropumps [J].
Zeng, SL ;
Chen, CH ;
Mikkelsen, JC ;
Santiago, JG .
SENSORS AND ACTUATORS B-CHEMICAL, 2001, 79 (2-3) :107-114