Variation in Tube Voltage for Adult Neck MDCT: Effect on Radiation Dose and Image Quality

被引:20
作者
Hoang, Jenny K. [1 ,2 ]
Yoshizumi, Terry T. [1 ]
Giao Nguyen [1 ]
Toncheva, Greta [1 ]
Choudhury, Kingshuk Roy [1 ]
Gafton, Andreia R. [1 ]
Eastwood, James D. [1 ]
Lowry, Carolyn [1 ]
Hurwitz, Lynne M. [1 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
[2] Duke Univ, Med Ctr, Dept Radiat Oncol, Durham, NC 27710 USA
关键词
CT; kilovoltage; neck; radiation dose; tube current; CHEST CT; REDUCTION; ENHANCEMENT; ANGIOGRAPHY;
D O I
10.2214/AJR.11.6831
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
OBJECTIVE. The purpose of this study was to assess the effect of peak kilovoltage on radiation dose and image quality in adult neck MDCT. MATERIALS AND METHODS. An anthropomorphic phantom with metal oxide semiconductor field effect transistor detectors was imaged with a 64-MDCT scanner. The reference CT protocol called for 120 kVp, and images obtained with that protocol were compared with CT images obtained with protocols entailing 80, 100, and 140 kVp. All imaging was performed with automatic tube current modulation. Organ dose and effective dose were determined for each protocol and compared with those obtained with the 120-kVp protocol. Image noise was evaluated objectively and subjectively for each protocol. RESULTS. The highest organ doses for all protocols were to the thyroid, ocular lens, skin, and mandible. The greatest reductions in organ dose were for the bone marrow of the cervical spine and mandible: 43% and 35% with the 100-kVp protocol and 63% and 53% with the 80-kVp protocol. Effective dose decreased as much as 9% with the 100-kVp protocol and 12% with the 80-kVp protocol. Use of the 140-kVp protocol was associated with an increase in organ dose as high as 64% for bone marrow in the cervical spine and a 19% increase in effective dose. Image noise increased with lower peak kilovoltage. The measured noise difference was greatest at 80 kVp, absolute increases were less than 2.5 HU. There was no difference in subjective image quality among protocols. CONCLUSION. Reducing the voltage from 120 to 80 kVp for neck CT can result in greater than 50% reduction in the absorbed organ dose to the bone marrow of the cervical spine and mandible without impairment in subjective image quality.
引用
收藏
页码:621 / 627
页数:7
相关论文
共 12 条
[1]
[Anonymous], 2006, HLTH RISKS EXP LOW L
[2]
Multi-detector row CT: Radiation dose characteristics [J].
Hamberg, LM ;
Rhea, JT ;
Hunter, GJ ;
Thrall, JH .
RADIOLOGY, 2003, 226 (03) :762-772
[3]
Image quality and radiation exposure at pulmonary CT angiography with 100-or 120-kVp protocol: prospective randomized study [J].
Heyer, Christoph M. ;
Mohr, Patrick S. ;
Lemburg, Stefan P. ;
Peters, Soeren A. ;
Nicolas, Volkmar .
RADIOLOGY, 2007, 245 (02) :577-583
[4]
Radiation dose reduction to breast and thyroid during MDCT:: Effectiveness of an in-plane bismuth shield [J].
Hohl, C ;
Wildberger, JE ;
Süss, C ;
Thomas, C ;
Mühlenbruch, G ;
Schmidt, T ;
Honnef, D ;
Günther, RW ;
Mahnken, AH .
ACTA RADIOLOGICA, 2006, 47 (06) :562-567
[5]
International Commission on Radiological Protection, 2007, INT COMM RAD PROT PU
[6]
Evaluation of a Radiation Dose Reduction Strategy for Pediatric Chest CT [J].
Kim, Jee-Eun ;
Newman, Beverley .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (05) :1188-1193
[7]
Thyroid Shields versus z-Axis Automatic Tube Current Modulation for Dose Reduction at Neck CT [J].
Leswick, David A. ;
Hunt, Megan M. ;
Webster, Steven T. ;
Fladeland, Derek A. .
RADIOLOGY, 2008, 249 (02) :572-580
[8]
Vascular Enhancement and Image Quality of MDCT Pulmonary Angiography in 400 Cases: Comparison of Standard and Low Kilovoltage Settings [J].
Matsuoka, Shin ;
Hunsaker, Andetta R. ;
Gill, Ritu R. ;
Oliva, Isabel B. ;
Trotman-Dickenson, Beatrice ;
Jacobson, Francine L. ;
Hatabu, Hiroto .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 192 (06) :1651-1656
[9]
Abdominal CT with low tube voltage: Preliminary observations about radiation dose, contrast enhancement, image quality, and noise [J].
Nakayama, Y ;
Awai, K ;
Funama, Y ;
Hatemura, M ;
Imuta, M ;
Nakaura, T ;
Ryu, D ;
Morishita, S ;
Sultana, S ;
Sato, N ;
Yamashita, Y .
RADIOLOGY, 2005, 237 (03) :945-951
[10]
Effect of Automatic Tube Current Modulation on Radiation Dose and Image Quality for Low Tube Voltage Multidetector Row CT Angiography: Phantom Study [J].
Schindera, Sebastian T. ;
Nelson, Rendon C. ;
Yoshizumi, Terry ;
Toncheva, Greta ;
Nguyen, Giao ;
DeLong, David M. ;
Szucs-Farkas, Zsolt .
ACADEMIC RADIOLOGY, 2009, 16 (08) :997-1002