Validation of a fully automated high throughput liquid chromatographic/tandem mass spectrometric method for roxithromycin quantification in human plasma. Application to a bioequivalence study

被引:13
作者
Kousoulos, Constantinos [1 ]
Tsatsou, Georgia [1 ]
Dotsikas, Yannis [1 ]
Apostolou, Constantinos [1 ]
Loukas, Yannis L. [1 ]
机构
[1] Univ Athens, Sch Pharm, Dept Pharmaceut Chem, Lab Pharmaceut Anal & Biequivalence Serv GLP Comp, GR-15771 Athens, Greece
关键词
roxithromycin; bioequivalence; 96-well; electrospray; validation;
D O I
10.1002/bmc.958
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A fully automated high-throughput liquid chromatography/tandem mass spectrometry (LC-MS/MS) method was developed for the determination of roxithromycin in human plasma. The plasma samples were treated by liquid-liquid extraction (LLE) in 2.2mL 96-deep-well plates. Roxithromycin and the internal standard clarithromycin were extracted from 100 mu L of human plasma by LLE, using methyl t-butyl ether as the organic solvent. All liquid transfer steps were performed automatically using robotic liquid handling workstations. After vortexing, centrifugation and freezing, the supernatant organic solvent was evaporated and reconstituted. Sample analysis was performed by reversed-phase LC-MS/MS, with positive ion electrospray ionization, using multiple-re action monitoring. The method had a very short chromatographic run time of 1.6 min. The calibration curve was linear for the range of concentrations 50.0-20.0 x 10(3) ng mL(-1). The proposed method was fully validated and it was proven to be selective, accurate, precise, reproducible and suitable for the determination of roxithromycin in human plasma. Therefore, it was applied to the rapid and reliable determination of roxithromycin in a bioequivalence study after per os administration of 300 mg tablet formulations of roxithromycin. Copyright (C) 2007 John Wiley & Sons, Ltd.
引用
收藏
页码:494 / 501
页数:8
相关论文
共 22 条
[1]   Linear regression for calibration lines revisited: weighting schemes for bioanalytical methods [J].
Almeida, AM ;
Castel-Branco, MM ;
Falcao, AC .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2002, 774 (02) :215-222
[2]   Model validation through the linear regression fit to actual versus predicted values [J].
Analla, M .
AGRICULTURAL SYSTEMS, 1998, 57 (01) :115-119
[3]  
Centre for Drug Evaluation and Research (CDER), 2001, GUID IND BIOAN METH
[4]   Comparative study of coulometric and amperometric detection for the determination of macrolides in human urine using high-performance liquid chromatography [J].
de la Huebra, MJG ;
Bordin, G ;
Rodríguez, AR .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2003, 375 (08) :1031-1037
[5]   An improved high-throughput liquid chromatographic/tandem mass spectrometric method for terbinafine quantification in human plasma, using automated liquid-liquid extraction based on 96-well format plates [J].
Dotsikas, Yannis ;
Apostolou, Constantinos ;
Kousoulos, Constantinos ;
Tsatsou, Georgia ;
Loukas, Yannis L. .
BIOMEDICAL CHROMATOGRAPHY, 2007, 21 (02) :201-208
[6]   ANALYSIS OF ERYTHROMYCIN AND ROXITHROMYCIN IN PLASMA OR SERUM BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY USING ELECTROCHEMICAL DETECTION [J].
GRGURINOVICH, N ;
MATTHEWS, A .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1988, 433 :298-304
[7]   LIQUID-CHROMATOGRAPHIC DETERMINATION OF THE MACROLIDE ANTIBIOTICS ROXITHROMYCIN AND CLARITHROMYCIN IN PLASMA BY AUTOMATED SOLID-PHASE EXTRACTION AND ELECTROCHEMICAL DETECTION [J].
HEDENMO, M ;
ERIKSSON, BM .
JOURNAL OF CHROMATOGRAPHY A, 1995, 692 (1-2) :161-166
[8]   Determination of macrolides in biological matrices by high-performance liquid chromatography with electrochemical detection [J].
Kees, F ;
Spangler, S ;
Wellenhofer, M .
JOURNAL OF CHROMATOGRAPHY A, 1998, 812 (1-2) :287-293
[9]   Determination of N-methyl-4-isoleucine-cyclosporin (NIM811) in human whole blood by high performance liquid chromatography-tandem mass spectrometry [J].
Li, Wenkui ;
Luo, Suyi ;
Hayes, Michael ;
He, Hanclan ;
Tse, Francis L. S. .
BIOMEDICAL CHROMATOGRAPHY, 2007, 21 (03) :249-256
[10]   Determination of roxithromycin residues in the flounder muscle with electrospray liquid chromatography-mass spectrometry [J].
Lim, JH ;
Jang, BS ;
Lee, RK ;
Park, SC ;
Yun, HI .
JOURNAL OF CHROMATOGRAPHY B, 2000, 746 (02) :219-225