All sealed endoleaks are not the same:: A treatment strategy based on an ex-vivo analysis

被引:27
作者
Mehta, M
Ohki, T
Veith, FJ
Lipsitz, EC
机构
[1] Montefiore Med Ctr, Div Vasc Surg, Bronx, NY 10467 USA
[2] Albert Einstein Coll Med, Bronx, NY 10467 USA
关键词
sealed endoleaks;
D O I
10.1053/ejvs.2001.1349
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: factors contributing to pressure transmission through thrombosed or sealed endoleaks have not been elucidated. The purpose of this investigation was to create an ex-vivo model that mimics patent and sealed endoleaks and that can quantitatively analyse the effects of length, diameter and thrombus on pressure transmission to the interior of the aneurysm sac. Methods: in the ex-vivo model, endoleak channels (ELCs) of various lengths (2 cm, 6 cm, 10 cm) and diameters (0.6 cm, 1.0 cm, 1.4 cm) were constructed using polytetrafluoroethylene (PTFE) grafts and attached to an artificial aneurysm sac. These ELCs were incorporated within a mock circulation made of rubber tubing connected to a pulsatile pump. Peak systolic pressure (PSP) was recorded in the aneurysm sac, distal to each ELC. Subsequently the ELCs were filled with human thrombus, and the pressure measurements repeated (n=5). Data was evaluated by regression analysis. Results: peak systolic pressure in the artificial circulation was maintained at 150 mmHg. In the absence of thrombus pressure did not change across the ELC, regardless of its length or diameter. In the presence of organised thrombus, the pressure curves distal to the ELC were dampened, and the pressure reduction was directly proportional to the length and inversely proportional to the diameter of the ELC. Regression analysis indicated statistical significance. Conclusions: in the absence of thrombosis, pressure transmitted via an ELC to the aneurysm sac is unchanged regardless of its length or diameter. All sealed endoleaks also transmit pressure. However, when an endoleak has thrombosed, pressure reduction is directly proportional to the length and inversely proportional to the diameter of its channel. This ex-vivo model suggests that Type 2 endoleaks with longer channels and smaller diameters would derive a greater benefit from adjunctive manoeuvres (coil embolisation) that hasten thrombosis. On the other hand, thrombosis of endoleaks with short and wide channels (e.g. Type 1) may not result in substantial pressure reduction within the aneurysm sac and a successful outcome.
引用
收藏
页码:541 / 544
页数:4
相关论文
共 17 条
[1]   Diagnosis and treatment of inferior mesenteric arterial endoleaks after endovascular repair of abdominal aortic aneurysms [J].
Baum, RA ;
Carpenter, JP ;
Tuite, CM ;
Velazquez, OC ;
Soulen, MC ;
Barker, CF ;
Golden, MA ;
Pyeron, AM ;
Fairman, RM .
RADIOLOGY, 2000, 215 (02) :409-413
[2]   Aneurysm sac pressure measurements after endovascular repair of abdominal aortic aneurysms [J].
Baum, RA ;
Carpenter, RP ;
Cope, C ;
Golden, MA ;
Velazquez, OC ;
Neschis, DG ;
Mitchell, ME ;
Barker, CF ;
Fairman, RM .
JOURNAL OF VASCULAR SURGERY, 2001, 33 (01) :32-40
[3]   Endoluminal stent-grafts for infrarenal abdominal aortic aneurysms [J].
Blum, U ;
Voshage, G ;
Lammer, J ;
Beyersdorf, F ;
Tollner, D ;
Kretschmer, G ;
Spillner, G ;
Polterauer, P ;
Nagel, G ;
Holzenbein, T ;
Thurnher, S ;
Langer, M .
NEW ENGLAND JOURNAL OF MEDICINE, 1997, 336 (01) :13-20
[4]   Freedom from endoleak after endovascular aneurysm repair does not equal treatment success [J].
Gilling-Smith, GL ;
Martin, J ;
Sudhindran, S ;
Gould, DA ;
McWilliams, RG ;
Bakran, A ;
Brennen, JA ;
Harris, PL .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2000, 19 (04) :421-425
[5]  
Görich J, 2000, RADIOLOGY, V215, P414
[6]   Embolization of perigraft leaks after endovascular stent-graft treatment of aortic aneurysms [J].
Kato, N ;
Semba, CP ;
Dake, MD .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1996, 7 (06) :805-811
[7]   Fate of endoleaks after endoluminal repair of abdominal aortic aneurysms with the EVT® device [J].
Makaroun, M ;
Zajko, A ;
Sugimoto, H ;
Eskandari, M ;
Webster, M .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 1999, 18 (03) :185-190
[8]   Endoleak after endovascular graft repair of experimental aortic aneurysms: Does coil embolization with angiographic "seal" lower intraaneurysmal pressure? [J].
Marty, B ;
Sanchez, LA ;
Ohki, T ;
Wain, RA ;
Faries, PL ;
Cynamon, J ;
Marin, ML ;
Veith, FJ .
JOURNAL OF VASCULAR SURGERY, 1998, 27 (03) :454-461
[9]   Reduction in aortic aneurysm size: Early results after endovascular graft placement [J].
Matsumura, JS ;
Pearce, WH ;
McCarthy, WJ ;
Yao, JST .
JOURNAL OF VASCULAR SURGERY, 1997, 25 (01) :113-123
[10]   Transfemoral endovascular repair of abdominal aortic aneurysm: Results of the North American EVT phase 1 trial [J].
Moore, WS ;
Rutherford, RB .
JOURNAL OF VASCULAR SURGERY, 1996, 23 (04) :543-553