Efficacy of the self-inflating bulb in confirming tracheal intubation in the morbidly obese

被引:28
作者
Lang, DJ [1 ]
Wafai, Y [1 ]
Salem, MR [1 ]
Czinn, EA [1 ]
Halim, AA [1 ]
Baraka, A [1 ]
机构
[1] AMER UNIV BEIRUT,DEPT ANESTHESIOL,BEIRUT,LEBANON
关键词
airway; management; self-inflating bulb; difficult; morbid obesity; equipment; esophageal detector device; intubation; esophageal; tracheal;
D O I
10.1097/00000542-199608000-00004
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: This study was designed to determine the incidence of false-negative and false-positive results when the self-inflating bulb (SIB) is used to differentiate tracheal from esophageal intubation in morbidly obese patients using two techniques. In technique 1, the SIB is compressed before it is connected to the tube; in technique 2, the SIB is compressed after connection to the tube. Methods: With institutional review board approval, 54 consenting adult morbidly obese patients (body mass index > 35) undergoing elective surgical procedures were included in the study. After anesthetic induction and muscle relaxation, both the trachea and esophagus were intubated under direct vision with identical cuffed tubes. The efficacy of the sip in verifying the position of both tubes was tested by a second anesthesiologist. The speed of reinflation was graded as rapid (<4 s) or none (>4 s), using both techniques. In the case of tracheal intubation, the absence of reinflation was recorded as a false-negative, whereas in cases of esophageal intubation, rapid reinflation was recorded as a false-positive. Identification of tube location by the second anesthesiologist was based on SIB reinflation results from techniques 1 and 2, as well as the presence of a flatuslike sound elicited by technique 2 in esophageally placed tubes. All patients were retested by the SIB after receiving three breaths of 400-500 ml each. In all patients exhibiting false-negative results, six obese patients exhibiting true-positive results, and four nonobese patients exhibiting true-positive results, tracheal responses to the SIB maneuvers were observed directly by a flexible fiberoptic bronchoscope incorporating an airtight system, 15-20 min after mechanical ventilation was instituted. Results: The incidence of false-negative results was initially 30% with technique 1 and 11% with technique 2, but decreased to 4% when technique 2 was used after the delivery of three breaths. The second anesthesiologist initially identified tube location in 92.5% of patients correctly. After the delivery of three breaths, tube location was correctly identified in 96.3% of patients. Fiberoptic bronchoscopic examination of the patients exhibiting false-negative results revealed exaggerated inward bulging of the posterior tracheal membrane during reinflation of the SIB when technique 1 was used. Conclusions: Contrary to previous investigations in healthy patients, the current study demonstrates a high incidence of false-negative results when the SIB is used to confirm tracheal intubation in morbidly obese patients. If the SIB is used, the technique should include compression of the SIB after connection to the tube and should be used in conjunction with other clinical signs and technical aids. The mechanism of false-negative results in these patients seems to be related to reduction of caliber of airways secondary to a marked decrease in functional residual capacity, and collapse of large airways due to invagination of the posterior tracheal wall when subatmospheric pressure is generated by the SIB.
引用
收藏
页码:246 / 253
页数:8
相关论文
共 29 条
[1]  
Agostoni E., 1964, HDB PHYSL, V1, P387
[2]  
BARAKA A, 1993, ANESTH ANALG, V77, P400
[3]  
BARAKA A, 1991, ANAESTHESIA, V45, P697
[4]   RELATIONSHIP BETWEEN AIRWAY RESISTANCE, AIRWAY CONDUCTANCE AND LUNG VOLUME IN SUBJECTS OF DIFFERENT AGE AND BODY SIZE [J].
BRISCOE, WA ;
DUBOIS, AB .
JOURNAL OF CLINICAL INVESTIGATION, 1958, 37 (09) :1279-1285
[5]  
BUCKLEY FP, 1992, CLIN ANESTH, P1170
[6]   THE ESOPHAGEAL DETECTOR DEVICE [J].
CALDER, I ;
SMITH, M ;
NEWTON, M .
ANAESTHESIA, 1989, 44 (08) :705-705
[7]   ADVERSE RESPIRATORY EVENTS IN ANESTHESIA - A CLOSED CLAIMS ANALYSIS [J].
CAPLAN, RA ;
POSNER, KL ;
WARD, RJ ;
CHENEY, FW .
ANESTHESIOLOGY, 1990, 72 (05) :828-833
[8]   MORTALITY ASSOCIATED WITH ANESTHESIA - A CASE REVIEW STUDY [J].
GANNON, K .
ANAESTHESIA, 1991, 46 (11) :962-966
[9]  
HAYNES SR, 1991, ANAESTHESIA, V45, P1067
[10]   ESOPHAGEAL INTUBATION - AN ANALYSIS OF 2000 INCIDENT REPORTS [J].
HOLLAND, R ;
WEBB, RK ;
RUNCIMAN, WB .
ANAESTHESIA AND INTENSIVE CARE, 1993, 21 (05) :608-610