Evaluation of the Dade Behring N Latex Cystatin C assay on the Dade Behring Nephelometer II system

被引:233
作者
Erlandsen, EJ
Randers, E
Kristensen, JH
机构
[1] Viborg Cty Hosp, Dept Clin Biochem, DK-8800 Viborg, Denmark
[2] Viborg Cty Hosp, Dept Internal Med, DK-8800 Viborg, Denmark
关键词
creatinine; cysteine proteinase inhibitors; glomerular filtration rate; immunoassay; kidney function tests;
D O I
10.1080/00365519950185940
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
The Dade Behring N Latex Cystatin C assay, a particle-enhanced nephelometric immunoassay for measuring serum cystatin C, was evaluated on the Dade Behring Nephelometer II. The assay time was 6 min and the throughput was 75 samples per hour. The sample volume was 40 mu L and the measuring range was 0.25 - 7.90 mg/L. Imprecision studies revealed within-run CVs <1.8% and between-run CVs <1.8% in the concentration range 0.87 - 4.63 mg/L. Recovery was 92.4 - 101.3%. Linearity studies showed excellent correlation between the theoretical and obtained values. No interferences were detected from haemoglobin <1.0 mmol/L, bilirubin <512 mu mol/L and Intralipid(R) <20 g/L. Stability of cystatin C in serum was 7 days at temperatures from 20 degrees C to - 20 degrees C and 6 months at - 80 degrees C. Measurements of cystatin C in heparin-plasma and EDTA-plasma did not differ significantly from cystatin C measured in serum. Fifty patient samples run on the Dade Behring Nephelometer II (y) were compared to the Dako Cystatin C assay (x). The Passing-Bablok regression analysis revealed y=1.105x - 0.340. In conclusion, the Dade Behring N Latex Cystatin C assay was precise and correlated with the Dako Cystatin C assay.
引用
收藏
页码:1 / 8
页数:8
相关论文
共 23 条
  • [1] THE PLACE OF HUMAN GAMMA-TRACE (CYSTATIN-C) AMONGST THE CYSTEINE PROTEINASE-INHIBITORS
    BARRETT, AJ
    DAVIES, ME
    GRUBB, A
    [J]. BIOCHEMICAL AND BIOPHYSICAL RESEARCH COMMUNICATIONS, 1984, 120 (02) : 631 - 636
  • [2] NOMENCLATURE AND CLASSIFICATION OF THE PROTEINS HOMOLOGOUS WITH THE CYSTEINE-PROTEINASE INHIBITOR CHICKEN CYSTATIN
    BARRETT, AJ
    FRITZ, H
    GRUBB, A
    ISEMURA, S
    JARVINEN, M
    KATUNUMA, N
    MACHLEIDT, W
    MULLERESTERL, W
    SASAKI, M
    TURK, V
    [J]. BIOCHEMICAL JOURNAL, 1986, 236 (01) : 312 - 312
  • [3] COMPARING METHODS OF MEASUREMENT - WHY PLOTTING DIFFERENCE AGAINST STANDARD METHOD IS MISLEADING
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1995, 346 (8982): : 1085 - 1087
  • [4] CATTANEO A, 1986, CYSTEINE PROTEINASES, P507
  • [5] CYSTATIN-C LEVELS IN SERA OF PATIENTS WITH HUMAN-IMMUNODEFICIENCY-VIRUS INFECTION - A NEW AVIDIN-BIOTIN ELISA ASSAY FOR ITS MEASUREMENT
    COLLE, A
    TAVERA, C
    PREVOT, D
    LEUNGTACK, J
    THOMAS, Y
    MANUEL, Y
    BENVENISTE, J
    LEIBOWITCH, J
    [J]. JOURNAL OF IMMUNOASSAY, 1992, 13 (01): : 47 - 60
  • [6] ELRANDSEN EJ, 1998, CLIN CHEM LAB MED, V36, P393
  • [7] Filler G, 1997, CLIN CHEM, V43, P1077
  • [8] Finney H, 1997, CLIN CHEM, V43, P1016
  • [9] Fito M, 1997, EUR J CLIN CHEM CLIN, V35, P787
  • [10] GLICK MR, 1986, CLIN CHEM, V32, P470