Development of mechanical circulatory support devices at the University of Tokyo

被引:20
作者
Abe, Yusuke [1 ]
Isoyama, Takashi [1 ]
Saito, Itsuro [2 ]
Mochizuki, Shuichi [1 ]
Ono, Minoru [3 ]
Nakagawa, Hidernoto [2 ]
Taniguchi, Noriyuki [1 ]
Mitsurnune, Noribiko [2 ]
Sugino, Ayaka [1 ]
Mitsui, Mie [2 ]
Takiura, Koki [4 ]
Ono, Toshiya [1 ]
Kouno, Akirnasa [1 ]
Chinzei, Tsuneo [2 ]
Takarnoto, Shinichi [3 ]
Imachi, Kou [4 ]
机构
[1] Univ Tokyo, Dept Biomed Engn, Grad Sch Med, Bunkyo Ku, Tokyo 1130033, Japan
[2] Univ Tokyo, Adv Sci & Technol Res Ctr, Tokyo, Japan
[3] Univ Tokyo, Grad Sch Med, Dept Cardiovasc Surg, Tokyo, Japan
[4] Tohoku Univ, Biomed Engn Org, Sendai, Miyagi 980, Japan
关键词
mechanical circulatory support; undulation pump; total artificial heart; ventricular assist device;
D O I
10.1007/s10047-006-0368-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
The development of mechanical circulatory support devices at the University of Tokyo has focused on developing a small total artificial heart (TAH) since achieving 532 days of survival of an animal with a paracorporial pneumatically driven TAH. The undulation pump was invented to meet this purpose. The undulation pump total artificial heart (UPTAH) is an implantable TAH that uses an undulation pump. To date, the UPTAH has been implanted in 71 goats weighting from 39 to 72 kg. The control methods are very important in animal experiments, and sucking control was developed to prevent atrial sucking. Rapid left-right balance control was performed by monitoring left atria] pressure to prevent acute lung edema caused by the rapid increase in both arterial pressure and venous return associated with the animal becoming agitated. Additionally, 1/R control was applied to stabilize the right atrial pressure. By applying these control methods, seven goats survived more than I month. The maximum survival period was 63 days. We are expecting to carry out longer term animal experiments with a recent model of TAH. In addition to the TAH, an undulation pump ventricular assist device (UPVAD), which is an implantable ventricular assist device (VAD), has been in development since 2002, based on the technology of the UPTAH. The UPVAD was implanted in six goats; three goats survived for more than I month. While further research and development is required to complete the the UPVAD system, the UPVAD has good potential to be realized as an implantable pulsatile-flow VAD.
引用
收藏
页码:60 / 70
页数:11
相关论文
共 42 条
[1]
Development of the undulation pump total artificial heart [J].
Abe, Y ;
Chinzei, T ;
Isoyama, T ;
Ono, T ;
Mochizuki, S ;
Mabuchi, K ;
Baba, K ;
Matsuura, H ;
Sun, YP ;
Kouno, A ;
Atsumi, K ;
Fujimasa, I ;
Imachi, K .
ARTIFICIAL ORGANS, 1997, 21 (07) :665-669
[2]
BASIC STUDY TO DEVELOP THE UNDULATION PUMP FOR PRACTICAL USE - ANTITHROMBOGENICITY, HEMOLYSIS, AND FLOW PATTERNS INSIDE THE PUMP [J].
ABE, Y ;
CHINZEI, T ;
ISOYAMA, T ;
ONE, T ;
MABUCHI, K ;
IMANISHI, K ;
KOUNO, A ;
ATSUMI, K ;
FUJIMASA, I ;
IMACHI, K .
ARTIFICIAL ORGANS, 1995, 19 (07) :691-693
[3]
Physiological control of a total artificial heart: conductance- and arterial pressure-based control [J].
Abe, Y ;
Chinzei, T ;
Mabuchi, K ;
Snyder, AJ ;
Isoyama, T ;
Imanishi, K ;
Yonezawa, T ;
Matsuura, H ;
Kouno, A ;
Ono, T ;
Atsumi, K ;
Fujimasa, I ;
Imachi, K .
JOURNAL OF APPLIED PHYSIOLOGY, 1998, 84 (03) :868-876
[4]
Basic study to develop an electromagnetic drive method for the rotary undulation pump [J].
Abe, Y ;
Chinzei, T ;
Isoyama, T ;
Saito, I ;
Ono, T ;
Mochizuki, S ;
Kouno, A ;
Imachi, K .
ARTIFICIAL ORGANS, 2003, 27 (10) :870-874
[5]
Implantation of the undulation pump total artificial heart in the goat [J].
Abe, Y ;
Chinzei, T ;
Ono, T ;
Mochizuki, S ;
Saito, I ;
Guba, P ;
Karita, T ;
Sun, YP ;
Kouno, A ;
Isoyama, T ;
Baba, K ;
Imachi, K .
ARTIFICIAL ORGANS, 1999, 23 (10) :932-938
[6]
Third model of the undulation pump total artificial heart [J].
Abe, Y ;
Chinzei, T ;
Isoyama, T ;
Ono, T ;
Mochizuki, S ;
Saito, I ;
Iwasaki, K ;
Ishimaru, M ;
Baba, A ;
Kouno, A ;
Ozeki, T ;
Tohyama, T ;
Imachi, K .
ASAIO JOURNAL, 2003, 49 (01) :123-127
[7]
Abe Y, 1994, ASAIO J, V40, pM506, DOI 10.1097/00002480-199407000-00051
[8]
Abe Y, 2001, ARTIF ORGANS, V25, P69
[9]
ABE Y, 1998, HEART REPLACEMENT AR, P34
[10]
ABE Y, 2002, J ARTIF ORGANS, V5, P84