Clinical outcomes after heterogeneous overlap stenting with drug-eluting Stents and bare-metal stents for de novo coronary artery narrowings

被引:10
作者
Aoki, Jiro [1 ,2 ]
Kirtane, Ajay J. [1 ,2 ]
Dangas, George D. [1 ,2 ]
Lansky, Alexandra J. [1 ,2 ]
Morales, Andy [1 ,2 ]
Kimura, Masashi [1 ,2 ]
Kim, Young-Hak [1 ,2 ]
Moussa, Issam [1 ,2 ]
Weisz, Giora [1 ,2 ]
Kreps, Edward M. [1 ,2 ]
Collins, Michael [1 ,2 ]
Frankin-Bond, Theresa [1 ,2 ]
Stone, Gregg W. [1 ,2 ]
Moses, Jeffrey W. [1 ,2 ]
Leon, Martin B. [1 ,2 ]
Mehran, Roxana [1 ,2 ]
机构
[1] Columbia Univ, Med Ctr, Ctr Intervent Vasc Therapy, New York, NY 10027 USA
[2] Cardiovasc Res Fdn, New York, NY USA
关键词
D O I
10.1016/j.amjcard.2007.07.049
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
When it is difficult to deliver multiple drug-eluting stents (DES) or when size constraints limit DES implantation, bare-metal stents (BMS) may be implanted contiguous to DES. However, the clinical outcomes after overlapping DES and BMS implantation are not known. From September 2004 to June 2006, 4,872 consecutive patients who underwent percutaneous coronary intervention consented to be enrolled in a prospective registry. Of these patients, 44 (0.9%) with de novo lesions were treated with DES and BMS overlap stenting. All patients were followed to 12 months for the assessment of clinical outcomes. The average implanted stent diameter was 2.68 +/- 0.30 mm for DES and 2.35 +/- 0.38 mm for BMS. Overlapping BMS were implanted distal to DES in all but 1 case. One patient (2.3%) experienced acute stent thrombosis and died 2 days after the Procedure. No other patient died or had a myocardial infarction during 12 months. The target vessel revascularization rate at 12 months, however, was 31.8%, mainly driven by diffuse in-stent restenosis in the BMS segments. In conclusion, the incidence of DES and BMS overlap stenting is rare in daily practice, but this procedure is associated with a high rate of target vessel revascularization. (c) 2008 Elsevier Inc. All rights reserved.
引用
收藏
页码:58 / 62
页数:5
相关论文
共 16 条
[11]   Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery [J].
Moses, JW ;
Leon, MB ;
Popma, JJ ;
Fitzgerald, PJ ;
Holmes, DR ;
O'Shaughnessy, C ;
Caputo, RP ;
Kereiakes, DJ ;
Williams, DO ;
Teirstein, PS ;
Jaeger, JL ;
Kuntz, RE .
NEW ENGLAND JOURNAL OF MEDICINE, 2003, 349 (14) :1315-1323
[12]  
Serruys Patrick W, 2006, EuroIntervention, V2, P286
[13]   Comparison of a polymer-based paclitaxel-eluting stent with a bare metal stent in patients with complex coronary artery disease - A randomized controlled trial [J].
Stone, GW ;
Ellis, SG ;
Cannon, L ;
Mann, JT ;
Greenberg, JD ;
Spriggs, D ;
O'Shaughnessy, CD ;
DeMaio, S ;
Hall, P ;
Popma, JJ ;
Koglin, J ;
Russell, ME .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2005, 294 (10) :1215-1223
[14]   A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease [J].
Stone, GW ;
Ellis, SG ;
Cox, DA ;
Hermiller, J ;
O'Shaughnessy, C ;
Mann, JT ;
Turco, M ;
Caputo, R ;
Bergin, P ;
Greenberg, J ;
Popma, JJ ;
Russell, ME .
NEW ENGLAND JOURNAL OF MEDICINE, 2004, 350 (03) :221-231
[15]   A retrograde approach for the treatment of chronic total occlusion in a patient with acute coronary syndrome [J].
Takano, Masamichi ;
Yamamoto, Masanori ;
Mizuno, Kyoichi .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2007, 119 (01) :E22-E24
[16]   Multiple overlapping drug-eluting stents to treat diffuse disease of the left anterior descending coronary artery [J].
Tsagalou, E ;
Chieffo, A ;
Iakovou, I ;
Ge, L ;
Sangiorgi, GM ;
Corvaja, N ;
Airoldi, F ;
Montorfano, M ;
Michev, L ;
Colombo, A .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 45 (10) :1570-1573