Reconciling in vitro and in vivo measurements of aerosol delivery from a metered-dose inhaler during mechanical ventilation and defining efficiency-enhancing factors

被引:92
作者
Fink, JB
Dhand, R
Grychowski, J
Fahey, PJ
Tobin, MJ
机构
[1] Edward Hines Vet Adm Hosp, Div Pulm & Crit Care Med, Hines, IL 60141 USA
[2] Loyola Univ, Stritch Sch Med, Hines, IL USA
关键词
D O I
10.1164/ajrccm.159.1.9803119
中图分类号
R4 [临床医学];
学科分类号
1002 [临床医学]; 100602 [中西医结合临床];
摘要
We attempted to resolve the discrepancies in reported data on aerosol deposition from a chlorofluorocarbon (CFC)-propelled metered-dose inhaler (MDI) during mechanical ventilation, obtained by in vivo and in vitro methodologies. Albuterol delivery to the lower respiratory tract was decreased in a humidified versus a dry circuit (16.2 versus 30.4%, respectively; p < 0.01). In 10 mechanically ventilated patients, 4.8% of the nominal dose was exhaled. When the exhaled aerosol was subtracted from the in vitro delivery of 16.2% achieved in a humidified ventilator circuit, the resulting value (16.2 - 4.8 = 11.4%) was similar to in vivo estimates of aerosol deposition. Having reconciled in vitro with in vivo findings, we then evaluated factors influencing aerosol delivery. A lower inspiratory flow rate (40 versus 80 L/min; p < 0.001), a longer duty cycle (0.50 versus 0.25; p < 0.04), and a shorter interval between successive MDI actuations (15 versus 60 s; p < 0.02) increased aerosol delivery, whereas use of a hydrofluoroalkane (HFA)-propelled MDI decreased aerosol delivery compared with the CFC-propelled MDI. A MDI and actuator combination other than that designed by the manufacturer altered aerosol particle size and decreased drug delivery. In conclusion, aerosol delivery in an in vitro model accurately reflects in vivo delivery, providing a means for investigating methods to improve the efficiency of aerosol therapy during mechanical ventilation.
引用
收藏
页码:63 / 68
页数:6
相关论文
共 24 条
[1]
CHOOSING THE METERED-DOSE INHALER SPACER OR HOLDING CHAMBER THAT MATCHES THE PATIENTS NEED - EVIDENCE THAT THE SPECIFIC DRUG BEING DELIVERED IS AN IMPORTANT CONSIDERATION [J].
AHRENS, R ;
LUX, C ;
BAHL, T ;
HAN, SH .
JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1995, 96 (02) :288-294
[2]
Dose-response to bronchodilator delivered by metered-dose inhaler in ventilator-supported patients [J].
Dhand, R ;
Duarte, AC ;
Jubran, A ;
Jenne, JW ;
Fink, JB ;
Fahey, PJ ;
Tobin, MJ .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1996, 154 (02) :388-393
[3]
BRONCHODILATOR DELIVERY BY METERED-DOSE INHALER IN VENTILATOR-SUPPORTED PATIENTS [J].
DHAND, R ;
JUBRAN, A ;
TOBIN, MJ .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1995, 151 (06) :1827-1833
[4]
Inhaled bronchodilator therapy in mechanically ventilated patients [J].
Dhand, R ;
Tobin, MJ .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1997, 156 (01) :3-10
[5]
ALBUTEROL DELIVERY IN A MODEL OF MECHANICAL VENTILATION - COMPARISON OF METERED-DOSE INHALER AND NEBULIZER EFFICIENCY [J].
DIOT, P ;
MORRA, L ;
SMALDONE, GC .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1995, 152 (04) :1391-1394
[6]
Aerosol delivery from a metered-dose inhaler during mechanical ventilation - An in vitro model [J].
Fink, JB ;
Dhand, R ;
Duarte, AG ;
Jenne, JW ;
Tobin, MJ .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1996, 154 (02) :382-387
[7]
FINK JB, 1995, AM J RESP CRIT CARE, V151, pA436
[8]
EFFICIENCY OF BRONCHODILATOR AEROSOL DELIVERY TO THE LUNGS FROM THE METERED-DOSE INHALER IN MECHANICALLY VENTILATED PATIENTS - A STUDY COMPARING 4 DIFFERENT ACTUATOR DEVICES [J].
FULLER, HD ;
DOLOVICH, MB ;
TURPIE, FH ;
NEWHOUSE, MT .
CHEST, 1994, 105 (01) :214-218
[9]
PRESSURIZED AEROSOL VERSUS JET AEROSOL DELIVERY TO MECHANICALLY VENTILATED PATIENTS - COMPARISON OF DOSE TO THE LUNGS [J].
FULLER, HD ;
DOLOVICH, MB ;
POSMITUCK, G ;
PACK, WW ;
NEWHOUSE, MT .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1990, 141 (02) :440-444
[10]
AEROSOL DELIVERY DURING MECHANICAL VENTILATION - A PREDICTIVE INVITRO LUNG MODEL [J].
FULLER, HD ;
DOLOVICH, MB ;
CHAMBERS, C ;
NEWHOUSE, MT .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 1992, 5 (04) :251-259