Development of a new predictive modelling technique to find with confidence equivalence zone and design space of chromatographic analytical methods

被引:63
作者
Lebrun, Pierre [1 ]
Govaerts, Bernadette [2 ]
Debrus, Benjamin [1 ]
Ceccato, Attilio
Caliaro, Gabriel
Hubert, Philippe [1 ]
Boulanger, Bruno
机构
[1] Univ Liege, Inst Pharm, Analyt Chem Lab, B-4000 Liege, Belgium
[2] Univ Catholique Louvain, Inst Stat, B-1348 Louvain, Belgium
关键词
HPLC; chromatography; design of experiments; multi-criteria optimization; statistical models; error propagation; design space;
D O I
10.1016/j.chemolab.2007.05.010
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A new method for modelling chromatographic responses is presented as a critical piece for the achievement of automated development of analytical methods. This methodology is based on four parts. First, we propose to use a very little set of statistical equations to create predictive models for retention time-based responses as the apex, the width and the asymmetry of peaks. Second, an experimental design is set up to realize experiments. Third, using grid search over the domain, multi-criteria decision is taken with respect to different local or global optimization criteria, used as desirability functions. This allows finding an optimal chromatogram. Fourth, we advice to investigate how the predictive error of the models propagates around optimal solution. This allows to give confidence in the optimal solution, in finding a set of zones that presumably will give an acceptable solution. Design spaces can be derived with a similar technique. The approach is exemplified with a real case and predictions of models at optimal analytical conditions are validated through new experiments. Flexibility is left over all the presented methodology. (C) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:4 / 16
页数:13
相关论文
共 15 条
  • [1] [Anonymous], EXPT DESIGNS
  • [2] BOULANGER B, 2006, C CHIM 2006 PAR DEC
  • [3] DERRINGER G, 1980, J QUAL TECHNOL, V12, P214, DOI 10.1080/00224065.1980.11980968
  • [4] DETILLEGHEM CL, 2005, UNCERTAINTY PROPAGAT, V532
  • [5] DETILLEGHEM CL, 2005, DISTRIBUTION DESIRAB, V531
  • [6] Development of response models for optimising HPLC methods
    Dewé, W
    Marini, RD
    Chiap, P
    Hubert, P
    Crommen, J
    Boulanger, B
    [J]. CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2004, 74 (02) : 263 - 268
  • [7] Harrington E. C., 1965, IND QUALITY CONTROL, V21, P494, DOI DOI 10.1128/AM.13.3.494-495.1965
  • [8] *ICH, Q8 DRAFT GUID PHARM
  • [9] Khui A., 1987, RESPONSE SURFACES DE
  • [10] Schoenmakers P., 1988, ANAL CHIM ACTA, V208, P357