Multi-detector row computed tomography of the heart: does a multi-segment reconstruction algorithm improve left ventricular volume measurements?

被引:27
作者
Juergens, KU
Maintz, D
Grude, M
Boese, JM
Heimes, B
Fallenberg, EM
Heindel, W
Fischbach, R
机构
[1] Univ Munster, Dept Clin Radiol, D-48149 Munster, Germany
[2] Univ Munster, Dept Cardiol & Angiol, D-48149 Munster, Germany
[3] Siemens Med Solut, D-91301 Forchheim, Germany
关键词
multi-detector row computed tomography; multi-segment reconstruction algorithm; magnetic resonance imaging; steady-state free precession cine sequences; global left ventricular function;
D O I
10.1007/s00330-004-2530-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A multi-segment cardiac image reconstruction algorithm in multi-detector row computed tomography (MDCT) was evaluated regarding temporal resolution and determination of left ventricular (LV) volumes and global LV function. MDCT and cine magnetic resonance (CMR) imaging were performed in 12 patients with known or suspected coronary artery disease. Patients gave informed written consent for the MDCT and the CMR exam. MDCT data were reconstructed using the standard adaptive cardiac volume (ACV) algorithm as well as a multi-segment algorithm utilizing data from three, five and seven rotations. LV end-diastolic (LV-EDV) and end-systolic volumes and ejection fraction (LV-EF) were determined from short-axis image reformations and compared to CMR data. Mean temporal resolution achieved was 192 24 ms using the ACV algonthm and improved significantly utilizing the three, five and seven data segments to 139 +/- 12, 113 +/- 13 and 96 +/- 11 ms (P<0.001 for each). Mean LV-EDV was without significant differences using the ACV algorithm, the multi-segment approach and CMR imaging. Despite improved temporal resolution with multi-segment image reconstruction, end-systolic volumes were less accurately measured (mean differences 3.9 +/- 11.8 ml to 8.1 +/- 13.9 ml), resulting in a consistent underestimation of LV-EF by 2.3-5.4% in comparison to CMR imaging (Bland-Altman analysis). Multi-segment image reconstruction improves temporal resolution compared to the standard ACV algorithm, but this does not result in a benefit for determination of LV volume and function.
引用
收藏
页码:111 / 117
页数:7
相关论文
共 26 条
[1]   In vivo validation of cardiac spiral computed tomography using retrospective gating [J].
Albers, J ;
Boese, JM ;
Vahl, CF ;
Hagl, S .
ANNALS OF THORACIC SURGERY, 2003, 75 (03) :885-889
[2]   Normal human left and right ventricular dimensions for MRI as assessed by turbo gradient echo and steady-state free precession imaging sequences [J].
Alfakih, K ;
Plein, S ;
Thiele, H ;
Jones, T ;
Ridgway, JP ;
Sivananthan, MU .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 17 (03) :323-329
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]  
Boese JM, 2000, RADIOLOGE, V40, P123, DOI 10.1007/s001170050020
[5]  
COLTART J, 1975, BRIT HEART J, V37, P357
[6]  
DEHMER GJ, 1981, BRIT HEART J, V45, P656
[7]   Usefulness of dynamic multislice computed tomography of left ventricular function in unstable angina pectoris and comparison with echocardiography [J].
Dirksen, MS ;
Bax, JJ ;
de Roos, A ;
Jukema, JW ;
van der Geest, RJ ;
Geleijns, K ;
Boersma, E ;
van der Wall, EE ;
Lamb, HJ .
AMERICAN JOURNAL OF CARDIOLOGY, 2002, 90 (10) :1157-+
[8]   Heart rate adaptive optimization of spatial and temporal resolution for electrocardiogram-gated multislice spiral CT of the heart [J].
Flohr, T ;
Ohnesorge, B .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2001, 25 (06) :907-923
[9]   Evaluation of global left ventricular myocardial function with electrocardiogram-gated multidetector computed tomography - Comparison with magnetic resonance imaging [J].
Grude, M ;
Juergens, KU ;
Wichter, T ;
Paul, M ;
Fallenberg, EM ;
Muller, JG ;
Heindel, W ;
Breithardt, G ;
Fischbach, R .
INVESTIGATIVE RADIOLOGY, 2003, 38 (10) :653-661
[10]   Evaluation of left ventricular dysfunction using multiphasic reconstructions of coronary multi-slice computed tomography data in patients with chronic ischemic heart disease: validation against cine magnetic resonance imaging [J].
Halliburton, SS ;
Petersilka, M ;
Schvartzman, PR ;
Obuchowski, N ;
White, RD .
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2003, 19 (01) :73-83