In vivo validation of cardiac spiral computed tomography using retrospective gating

被引:6
作者
Albers, J [1 ]
Boese, JM
Vahl, CF
Hagl, S
机构
[1] Univ Heidelberg, Dept Cardiac Surg, D-69120 Heidelberg, Germany
[2] German Canc Res Ctr, Dept Biophys & Med Radiat Phys, D-6900 Heidelberg, Germany
关键词
D O I
10.1016/S0003-4975(02)04505-8
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
Background. Cardiac functional assessment represents the basis for diagnostics and cardiac operation planning. Spiral computed tomography (CT) combines the advantages of three-dimensional imaging and high temporal resolution when using gating techniques. However, in vivo validation data of this novel imaging technology are lacking. The purpose of this study was to validate in vivo the new imaging method using retrospective gating and to evaluate the clinical usefulness of the achieved temporal resolution. Methods. In domestic pigs (n = 10, weight 35 to 40 kg) a flowmeter was placed surgically on the ascending aorta. Flow velocity integrated over systole served as the gold standard for left ventricular (LV) stroke volume (LVSV-FM). CT signal, projection data, pacemaker signal, and flow velocity were recorded simultaneously at constant heart rate (pacemaker, 90 beats per minute). End-systolic and end-diastolic frames were calculated by retrospective gating. LV volumes were traced, the difference representing CT stroke volume (LVSV-CT). Image data were three-dimensionally reconstructed using ray-tracing. Results. Temporal resolution was 170 ms. Correlation of stroke volumes was high (r = 0.94, mean difference 1.75 mL). Intraobserver (0.49 mL, for LVEDV, 0.31 for LVESV) and interobserver variability (p = 0.21 and p = 0.06, respectively) were low. Postprocessing resulted in four-dimensional beating-heart models useful for operation planning. Conclusions. Spiral CT using retrospective gating was validated in vivo. Clinically acceptable temporal resolution and accuracy in determining cardiac stroke volumes were found. As a true volumetric imaging modality the method may now play an important role in computer-assisted diagnostics and surgery. (C) 2003 by The Society of Thoracic Surgeons.
引用
收藏
页码:885 / 889
页数:5
相关论文
共 16 条
[1]
Achenbach S, 2001, CIRCULATION, V103, P2535
[2]
Becker CR, 1998, RADIOLOGE, V38, P987, DOI 10.1007/s001170050452
[3]
Becker CR, 1998, RADIOLOGE, V38, P726, DOI 10.1007/s001170050416
[4]
Boese JM, 2000, RADIOLOGE, V40, P123, DOI 10.1007/s001170050020
[5]
Assessment of mitral regurgitant jets by three-dimensional color Doppler [J].
De Simone, R ;
Glombitza, G ;
Vahl, CF ;
Albers, J ;
Meinzer, HP ;
Hagl, S .
ANNALS OF THORACIC SURGERY, 1999, 67 (02) :494-499
[6]
ESTIMATION OF LEFT VENTRICULAR VOLUME BY 1-PLANE CINEANGIOGRAPHY [J].
GREENE, DG ;
CARLISLE, R ;
GRANT, C ;
BUNNELL, IL .
CIRCULATION, 1967, 35 (01) :61-&
[7]
Four multidetector-row helical CT: Image quality and volume coverage speed [J].
Hu, H ;
He, HD ;
Foley, WD ;
Fox, SH .
RADIOLOGY, 2000, 215 (01) :55-62
[8]
GATED MULTIPLANAR CARDIAC COMPUTED-TOMOGRAPHY [J].
JOHNSON, GA ;
GODWIN, JD ;
FRAM, EK .
RADIOLOGY, 1982, 145 (01) :195-197
[9]
Electrocardiogram-correlated image reconstruction from subsecond spiral computed tomography scans of the heart [J].
Kachelriess, M ;
Kalender, WA .
MEDICAL PHYSICS, 1998, 25 (12) :2417-2431
[10]
Cardiac multidetector-row CT:: Retrospectively ECG-gated spiral with optimized temporal and spatial resolution:: First clinical results. [J].
Kopp, AF ;
Ohnesorge, B ;
Flohr, T ;
Georg, C ;
Schröder, S ;
Küttner, A ;
Martensen, J ;
Claussen, CD .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (05) :429-435