Fluid volumes determination by impedance spectroscopy and hematocrit monitoring: Application to pediatric hemodialysis

被引:25
作者
Fenech, M
Maasrani, M
Jaffrin, MY
机构
[1] Technol Univ Compiegne, CNRS, UMR 6600, Compiegne, France
[2] Islam Univ Lebanon, Beirut, Lebanon
[3] Inst Univ France, Paris, France
关键词
pediatric hemodialysis; bioimpedance; extracellular; interstitial; intracellular fluid;
D O I
10.1046/j.1525-1594.2001.025002089.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
A method for extracting fluid volumes from multifrequency bioimpedance, which takes into account the body geometry and the presence of nonconducting elements, was tested on 12 young dialyzed patients against correlations for total body water volumes (TBW) from Watson et al. and Humes et al. Our calculations of TBW from impedance were found to overestimate Humes' values by 0.25 L (0.8%) postdialysis and by 2.08 L (6.5%) predialysis. Extracellular water (ECW) was found to contribute an average of 93% of ultrafiltered volume. Intracellular water volume (ICW) determination from impedance was found to be too imprecise to predict its variation during dialysis; therefore, ICW variations were calculated as the difference between ultrafiltration and ECW changes. The continuous recording of hematocrit by an optical device monitored changes in plasma and interstitial volumes. In most cases, ultrafiltration was compensated mainly by a contribution from interstitial fluid, and the drop in plasma volume was generally moderate.
引用
收藏
页码:89 / 98
页数:10
相关论文
共 22 条
[1]
de Vries P M, 1988, ASAIO Trans, V34, P623
[2]
Predicting body cell mass with bioimpedance by using theoretical methods: A technological review [J].
DeLorenzo, A ;
Andreoli, A ;
Matthie, J ;
Withers, P .
JOURNAL OF APPLIED PHYSIOLOGY, 1997, 82 (05) :1542-1558
[3]
MEASUREMENT OF TRANS-CELLULAR FLUID SHIFT DURING HEMODIALYSIS .2. INVITRO AND CLINICAL-EVALUATION [J].
DEVRIES, PMJM ;
MEIJER, JH ;
VLAANDEREN, K ;
VISSER, V ;
OE, PL ;
DONKER, AJM ;
SCHNEIDER, H .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1989, 27 (02) :152-158
[4]
FINBERG L, 1993, WATER ELECTROLYTES P
[5]
Hanai T., 1968, EMULSION SCI
[6]
BIOIMPEDANCE ANALYSIS OF TOTAL-BODY WATER IN HEMODIALYSIS-PATIENTS [J].
HO, LT ;
KUSHNER, RF ;
SCHOELLER, DA ;
GUDIVAKA, R ;
SPIEGEL, DM .
KIDNEY INTERNATIONAL, 1994, 46 (05) :1438-1442
[8]
JABARA AE, 1995, ASAIO J, V41, pM682, DOI 10.1097/00002480-199507000-00098
[9]
Extra- and intracellular volume monitoring by impedance during haemodialysis using Cole-Cole extrapolation [J].
Jaffrin, MY ;
Maasrani, M ;
LeGourrier, A ;
Boudailliez, B .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1997, 35 (03) :266-270
[10]
Comparison of optical, electrical, and centrifugation techniques for haematocrit monitoring of dialysed patients [J].
Jaffrin, MY ;
Fournier, C .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1999, 37 (04) :433-439