Microwave ablation with triaxial antennas tuned for lung: Results in an in vivo porcine model

被引:49
作者
Durick, Nathan A. [1 ]
Laeseke, Paul F. [1 ]
Broderick, Lynn S. [1 ]
Lee, Fred T., Jr. [1 ]
Sampson, Lisa A. [1 ]
Frey, Tina M. [1 ]
Warner, Thomas F. [2 ]
Fine, Jason P. [3 ]
Van der Weide, Daniel W. [1 ,4 ]
Brace, Christopher L. [1 ,4 ]
机构
[1] Univ Wisconsin, Dept Radiol, Madison, WI 53792 USA
[2] Univ Wisconsin, Dept Pathol & Lab Sci, Madison, WI 53792 USA
[3] Univ Wisconsin, Dept Biostat, Madison, WI 53792 USA
[4] Univ Wisconsin, Dept Elect & Comp Engn, Madison, WI 53792 USA
关键词
D O I
10.1148/radiol.2471062123
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Purpose: To prospectively determine in swine the size and shape of coagulation zones created in normal lung tissue by using small-diameter triaxial microwave antennas and to prospectively quantify the effects of bronchial occlusion and multiple antennas on the coagulation zone. Materials and Methods: The study was approved by the research animal care and use committee, and all husbandry and experimental studies were compliant with the National Research Council's Guide for the Care and Use of Laboratory Animals. Twentyfour coagulation zones (three per animal) were created at thoracotomy in eight female domestic swine (mean weight, 55 kg) by using a microwave ablation system with 17-gauge lung-tuned triaxial antennas. Ablations were performed for 10 minutes each by using (a) a single antenna, (b) a single antenna with bronchial occlusion, and (c) an array of three antennas powered simultaneously. The animals were sacrificed immediately after ablation. The coagulation zones were excised en bloc and sectioned into approximately 4-mm slices for measurement of size, shape, and circularity. Analysis of variance and two-sample t tests were used to identify differences between the three ablation groups. Results: The overall mean diameters of coagulation achieved with a single antenna and bronchial occlusion (4.11 cm +/- 1.09 [standard deviation]) and with multiple-antenna arrays (4.05 cm +/- 0.69) were significantly greater than the overall mean diameter achieved with a single antenna alone (3.09 cm +/- 0.83) (P = .016 for comparison with multiple antennas, P = .032 for comparison with bronchial occlusion). No significant differences in size were seen between the coagulation zones created with bronchial occlusion and those created with multiple antennas (P = .68). The coagulation zones in all groups were very circular (isoperimetric ratio > 0.80) at cross-sectional analysis. Conclusion: A 17-gauge triaxial microwave ablation system tuned for lung tissue yielded large circular zones of coagulation in vivo in porcine lungs. The coagulation zones created with bronchial occlusion and multiple antennas were significantly larger than those created with one antenna. (c) RSNA, 2008.
引用
收藏
页码:80 / 87
页数:8
相关论文
共 40 条
[1]
Radiofrequency ablation: Effect of surrounding tissue composition on coagulation necrosis in a canine tumor model [J].
Ahmed, M ;
Liu, ZJ ;
Afzal, KS ;
Weeks, D ;
Lobo, SM ;
Kruskal, JB ;
Lenkinski, RE ;
Goldberg, SN .
RADIOLOGY, 2004, 230 (03) :761-767
[2]
Percutaneous radiofrequency ablation of lung neoplasms: Initial therapeutic response [J].
Akeboshi, M ;
Yamakado, K ;
Nakatsuka, A ;
Hataji, O ;
Taguchi, O ;
Takao, M ;
Takeda, K .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2004, 15 (05) :463-470
[3]
BENUMOF JL, 1987, ANESTHESIA THORACIC, P224
[4]
Microwave ablation with multiple simultaneously powered small-gauge triaxial antennas: Results from an in vivo swine liver model [J].
Brace, Christopher L. ;
Laeseke, Paul F. ;
Sampson, Lisa A. ;
Frey, Tina M. ;
van der Weide, Daniel W. ;
Lee, Fred T., Jr. .
RADIOLOGY, 2007, 244 (01) :151-156
[5]
Microwave ablation with a single small-gauge triaxial antenna: In vivo porcine liver model [J].
Brace, Christopher L. ;
Laeseke, Paul F. ;
Sampson, Lisa A. ;
Frey, Tina M. ;
van der Weide, Daniel W. ;
Lee, Fred T., Jr. .
RADIOLOGY, 2007, 242 (02) :435-440
[6]
Microwave ablation with a triaxial antenna:: Results in ex vivo bovine liver [J].
Brace, CL ;
Laeseke, PF ;
van der Weide, DW ;
Lee, FT .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2005, 53 (01) :215-220
[7]
Effect of vascular occlusion on radiofrequency ablation of the liver: Results in a porcine model [J].
Chinn, SB ;
Lee, FT ;
Kennedy, GD ;
Chinn, C ;
Johnson, CD ;
Winter, TC ;
Warner, TF ;
Mahvi, DM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (03) :789-795
[8]
Midterm local efficacy and survival after radiofrequency ablation of lung tumors with minimum follow-up of 1 year:: Prospective evaluation [J].
de Baere, Thierry ;
Palussiere, Jean ;
Auperin, Anne ;
Hakime, Antoine ;
Abdel-Rehim, Mohamed ;
Kind, Michele ;
Dromain, Clarisse ;
Ravaud, Alain ;
Tebboune, Nathalie ;
Boige, Valerie ;
Malka, David ;
Lafont, Clarisse ;
Ducreux, Michel .
RADIOLOGY, 2006, 240 (02) :587-596
[9]
Control of interstitial thermal coagulation: Comparative evaluation of microwave and ultrasound applicators [J].
Deardorff, DL ;
Diederich, CJ ;
Nau, WH .
MEDICAL PHYSICS, 2001, 28 (01) :104-117
[10]
DOCARMO M, 1976, DIFFERENTIAL GEOMETR, P393