A non-invasive immobilization system and related quality assurance for dynamic intensity modulated radiation therapy of intracranial and head and neck disease

被引:49
作者
Tsai, JS
Engler, MJ
Ling, MN
Wu, JK
Kramer, B
Dipetrillo, T
Wazer, DE
机构
[1] Tufts Univ, New England Med Ctr Hosp, Dept Radiat Oncol, Boston, MA 02111 USA
[2] Tufts Univ, New England Med Ctr Hosp, Dept Neurosurg, Boston, MA 02111 USA
[3] Tufts Univ, Sch Med, Boston, MA 02111 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1999年 / 43卷 / 02期
关键词
non-invasive immobilization; polyurethane cradle; alpha cradle; thermoplastic mask; aquaplast; conformal radiotherapy; intensity-modulated radiotherapy; IMRT; topograms; quality assurance;
D O I
10.1016/S0360-3016(98)00398-8
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To develop and implement a non-invasive immobilization system guided by a dedicated quality assurance (QA) program for dynamic intensity-modulated radiotherapy (IMRT) of intracranial and head and neck disease, with IMRT delivered using the NOMOS Corporation's Peacock System and MIMiC collimator. Methods and Materials: Thermoplastic face masks are combined with cradle-shaped polyurethane foaming agents and a dedicated quality assurance program to create a customized headholder system (CHS). Plastic shrinkage was studied to understand its effect on immobilization. Fiducial points for computerized tomography (CT) are obtained by placing multiple dabs of barium paste on mask surfaces at intersections of laser projections used for patient positioning. Fiducial lines are drawn on the cradle along laser projections aligned with nasal surfaces. Lateral CT topograms are annotated with a crosshair indicating the origin of the treatment planning and delivery coordinate system, and with lines delineating the projections of superior-inferior field borders of the linear accelerator's secondary collimators, or with those of the fully open MIMiC. Port films exposed with and without the MIMIC are compared to annotated topograms to measure positional variance (PV) in superior-inferior (SI), right-left (RL), and anterior posterior (AP) directions. MIMiC vane patterns superposed on port films are applied to verify planned patterns. A 12-patient study of PV was performed by analyzing positions of 10 anatomic points on repeat CT topograms, plotting histograms of PV, and determining average PV. Results and Discussion: A 1.5 +/- 0.3 mm SD shrinkage per 70 cm of thermoplastic was observed over 24 h. Average PV of 1.0 +/- 0.8, 1.2 +/- 1.1, and 1.3 +/- 0.8 mm were measured in SI, AP, and RL directions, respectively. Lateral port films exposed with and without the MIMiC showed PV of 0.2 +/- 1.3 and 0.8 +/- 2.2 mm in AP and SI directions. Vane patterns superimposed on port films consistently verified the planned patterns. Conclusion: The CHS provided adequately reproducible immobilization for dynamic IMRT, and may be applicable to decrease PV for other cranial and head and neck external beam radiation therapy. (C) 1999 Elsevier Science Inc.
引用
收藏
页码:455 / 467
页数:13
相关论文
共 32 条
[1]   Comparison of two dead and neck immobilization systems [J].
Bentel, GC ;
Marks, LB ;
Hendren, K ;
Brizel, DM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04) :867-873
[2]   A CUSTOMIZED HEAD AND NECK SUPPORT SYSTEM [J].
BENTEL, GC ;
MARKS, LB ;
SHEROUSE, GW ;
SPENCER, DP .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 32 (01) :245-248
[3]   The field-matching problem as it applies to the peacock three dimensional conformal system for intensity modulation [J].
Carol, M ;
Grant, WH ;
Bleier, AR ;
Kania, AA ;
Targovnik, HS ;
Butler, EB ;
Woo, SW .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 34 (01) :183-187
[4]  
Carol M. P., 1994, P 11 INT C US COMP R, P172
[5]   PEACOCK(TM) - A SYSTEM FOR PLANNING AND ROTATIONAL DELIVERY OF INTENSITY-MODULATED FIELDS [J].
CAROL, MP .
INTERNATIONAL JOURNAL OF IMAGING SYSTEMS AND TECHNOLOGY, 1995, 6 (01) :56-61
[6]  
CAROL MP, 1993, 3D RAD TREATMENT PLA, P249
[7]   EXTERNAL STEREOTACTIC IRRADIATION BY LINEAR-ACCELERATOR [J].
COLOMBO, F ;
BENEDETTI, A ;
POZZA, F ;
AVANZO, RC ;
MARCHETTI, C ;
CHIEREGO, G ;
ZANARDO, A .
NEUROSURGERY, 1985, 16 (02) :154-160
[8]   FRACTIONATED RADIOTHERAPY OF SMALL INOPERABLE LESIONS OF THE BRAIN USING A NONINVASIVE STEREOTAXIC FRAME [J].
DELANNES, M ;
DALY, NJ ;
BONNET, J ;
SABATIER, J ;
TREMOULET, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (03) :749-755
[9]   ASSURANCE OF HIGH-QUALITY LINAC-BASED STEREOTAXIC RADIOSURGERY [J].
DRZYMALA, RE ;
KLEIN, EE ;
SIMPSON, JR ;
RICH, KM ;
WASSERMAN, TH ;
PURDY, JA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 30 (02) :459-472
[10]   Application of dynamic intensity-modulated radiation therapy to the treatment of prostate cancer [J].
Engler, MJ ;
Tsai, JS ;
Vora, S ;
Wazer, DE ;
DiPetrillo, T .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 39 (02) :193-193