Performance characteristics of the Pantak DXT-300 kilovoltage X-ray treatment machine

被引:21
作者
Aukett, RJ [1 ]
Thomas, DW [1 ]
Seaby, AW [1 ]
Gittins, JT [1 ]
机构
[1] SINGLETON HOSP,DEPT MED PHYS & CLIN ENGN,SWANSEA NHS TRUST,SWANSEA SA2 8QA,W GLAM,WALES
关键词
D O I
10.1259/0007-1285-69-824-726
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The beam and performance characteristics of a new orthovoltage X-ray unit, the Pantak DXT-300 have been evaluated. Data were collated for four qualities: 3.27 mmAl, 7.15 mmAl, 1.65 mmCu and 3.51 mmCu half value layer (HVL) (SE = 0.04 mm). Parameters which were investigated included beam quality, central axis depth dose, relative output, backscatter factors, field uniformity, peripheral dose and head leakage. The calibration procedure and the performance of the dosimetry system have also been described.
引用
收藏
页码:726 / 734
页数:9
相关论文
共 8 条
[1]  
*BRIT I RAD HOSP P, 1983, CENTR AX DEPTH DOS S, V17
[2]   BEAM CHARACTERISTICS OF THE THERAPAX DXT300 ORTHOVOLTAGE THERAPY UNIT [J].
GERIG, L ;
SOUBRA, M ;
SALHANI, D .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (09) :1377-1392
[3]   NEW SUPERFICIAL X-RAY THERAPY UNIT [J].
LILLICRAP, SC ;
WILSON, BC ;
EDSER, E ;
STANLEY, D .
BRITISH JOURNAL OF RADIOLOGY, 1974, 47 (555) :193-195
[4]  
*NAT RAD PROT BOAR, 1988, GUID NOT PROT PERS I
[5]   PERFORMANCE-CHARACTERISTICS OF AN ORTHOVOLTAGE X-RAY THERAPY MACHINE [J].
NIROOMANDRAD, A ;
GILLIN, MT ;
LOPEZ, F ;
GRIMM, DF .
MEDICAL PHYSICS, 1987, 14 (05) :874-878
[6]  
ROSSER KE, 1992, MEASUREMETN ABSORBED
[7]   PERFORMANCE-CHARACTERISTICS OF A WIDELY USED ORTHOVOLTAGE X-RAY THERAPY MACHINE [J].
SCRIMGER, JW ;
CONNORS, SG .
MEDICAL PHYSICS, 1986, 13 (02) :267-269
[8]  
1991, PHYS MED BIOL, V36, P1027