Radiation dose to the eyes and parotids during CT of the sinuses

被引:22
作者
Bassim, MK
Ebert, CS
Sit, RC
Senior, BA
机构
[1] Univ N Carolina, Dept Otolaryngol Head & Neck Surg, Chapel Hill, NC 27599 USA
[2] Univ N Carolina, Dept Radiat Safety, Chapel Hill, NC 27599 USA
关键词
D O I
10.1016/j.otohns.2005.06.018
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
OBJECTIVE: To measure the radiation dose to the lens and parotid during high-resolution computed tomography scan of the sinuses. STUDY DESIGN AND SETTING: Nine cadaver heads were scanned in the axial plane by means of a fine-cut (0.75 mm) protocol. Images were then reconstructed in the coronal and sagittal planes for use with the image guidance software. Thermoluminescent dosimeters were taped over the eyes and parotids and used to measure the radiation dose absorbed by these organs. RESULTS: Doses obtained were 29.5 mGy for the lens and around 30 mGy for the parotid. CONCLUSION: The measured doses are lower than the reported acute thresholds of 500-2000 mGy for lens opacities and well below the threshold of 2500 mGy for damage to the parotid. SIGNIFICANCE: These results demonstrate minimal risk from radiation through the use of high-resolution computed tomography and support the use of such a protocol for diagnosis and preoperative planning. (C) 2005 American Academy of Otolaryngology-Head and Neck Surgery Foundation, Inc. All fights reserved.
引用
收藏
页码:531 / 533
页数:3
相关论文
共 18 条
[1]   CT scanning of the paranasal sinuses:: axial helical CT with reconstruction in the coronal direction versus coronal helical CT [J].
Bernhardt, TM ;
Rapp-Bernhardt, U ;
Fessel, A ;
Ludwig, K ;
Reichel, G ;
Grote, R .
BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (848) :846-851
[2]  
Cohnen M, 1998, Aktuelle Radiol, V8, P34
[3]   Dosage measurement in spiral CT examinations in the head and neck region [J].
Cohnen, M ;
Cohnen, B ;
Ewen, K ;
Teubert, G ;
Modder, U .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1998, 168 (05) :474-479
[4]   Radiation dose to the lens in sequential and spiral CT of the facial bones and sinuses [J].
Czechowski, J ;
Janeczek, J ;
Kelly, G ;
Johansen, J .
EUROPEAN RADIOLOGY, 2001, 11 (04) :711-713
[5]   Radiation exposure with spiral CT of the paranasal sinuses [J].
Dammann, F ;
Momino-Traserra, E ;
Remy, C ;
Pereira, PL ;
Baumann, I ;
Koitschev, A ;
Claussen, CD .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (03) :232-237
[6]  
Hopper KD, 2001, AM J NEURORADIOL, V22, P1194
[7]   CT scanning of the paranasal sinuses - the effect of reducing mAs [J].
Kearney, SE ;
Jones, P ;
Meakin, K ;
Garvey, CJ .
BRITISH JOURNAL OF RADIOLOGY, 1997, 70 (838) :1071-1074
[8]   RADIATION-DOSE TO THE LENS FROM CORONAL CT SCANNING OF THE SINUSES [J].
MACLENNAN, AC .
CLINICAL RADIOLOGY, 1995, 50 (04) :265-267
[9]   PARANASAL SINUSES - LOW-DOSE CT [J].
MARMOLYA, G ;
WIESEN, EJ ;
YAGAN, R ;
HARIA, CD ;
SHAH, AC .
RADIOLOGY, 1991, 181 (03) :689-691
[10]  
Moulin G, 1996, NEURORADIOLOGY, V38, pS127