Comparison of computational analysis with clinical measurement of stresses on below-knee residual limb in a prosthetic socket

被引:47
作者
Zhang, M [1 ]
Roberts, C
机构
[1] Hong Kong Polytech Univ, Jockey Club Rehabil Engn Ctr, Kowloon, Hong Kong, Peoples R China
[2] Univ London Kings Coll, Guys Kings & St Thomas Hosp GKT Sch Med, Dept Med Engn & Phys, London SE5 9RS, England
关键词
prosthetic socket; pressure; shear stress; finite element modelling;
D O I
10.1016/S1350-4533(00)00079-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Interface pressures and shear stresses between a below-knee residual limb and prosthetic socket predicted using finite element analyses were compared with experimental measurements. A three-dimensional nonlinear finite element model, based on actual residual geometry and incorporating PTB socket rectification and interfacial friction/slip conditions, was developed to predict the stress distribution. A system for measuring pressures and bi-axial shear stresses was used to measure the stresses in the PTB socket of a trans-tibial amputee. The FE-predicted results indicated that the peak pressure of 226 kPa occurred at the patellar tendon area and the peak shear stress of 50 kPa at the anterolateral tibia area. Quantitatively, FE-predicted pressures were 11%, on average, lower than those measured by triaxial transducers placed at all the measurement sites. Because friction/slip conditions between the residual limb and socket liner were taken into consideration by using interface elements in the FE model, the directions and magnitudes of shear stresses match well between the FE prediction and clinical measurements. The results suggest that the nonlinear mechanical properties of soft tissues and dynamic effects during gait should be addressed in future work. (C) 2001 IPEM. Published by Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:607 / 612
页数:6
相关论文
共 21 条
[1]  
Brennan J.M., 1991, ADV BIOENG, V20, P547
[2]  
Engsberg J. R., 1992, J ASS CHILD PROSTHET, V37, P81
[4]  
LORD M, 1984, PROSTHET ORTHOT INT, V8, P159
[5]  
Pearson J R, 1973, Bull Prosthet Res, V10, P52
[6]  
QUESADA P, 1991, J REHAB RES DEV, V29, P1
[7]   INTERFACE LOAD ANALYSIS FOR COMPUTER-AIDED-DESIGN OF BELOW-KNEE PROSTHETIC SOCKETS [J].
REYNOLDS, DP ;
LORD, M .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1992, 30 (04) :419-426
[8]   CLINICAL MEASUREMENT OF NORMAL AND SHEAR STRESSES ON A TRANS-TIBIAL STUMP - CHARACTERISTICS OF WAVE-FORM SHAPES DURING WALKING [J].
SANDERS, JE ;
DALY, CH ;
BURGESS, EM .
PROSTHETICS AND ORTHOTICS INTERNATIONAL, 1993, 17 (01) :38-48
[9]  
Sanders JE, 1998, J REHABIL RES DEV, V35, P161
[10]  
SANDERS JE, 1993, J REHABIL RES DEV, V30, P191