Detection performance of SIGaAs detectors for nuclear medicine

被引:6
作者
Bertolucci, E
Bisogni, MG
Bottigli, U
Conti, M
Cola, A
Fantacci, ME
Maestro, P
Mettivier, G
Quaranta, F
Rosso, V
Russo, P
Stefanini, A
Vasanelli, L
机构
[1] Univ Naples Federico II, Ist Nazl Fis Nucl, Sez Napoli, Dipartimento Sci Fisiche, I-80126 Naples, Italy
[2] Univ Pisa, Dipartimento Fis, I-56100 Pisa, Italy
[3] Ist Nazl Fis Nucl, Sez Pisa, I-56100 Pisa, Italy
[4] CNR, Ist IME, Lecce, Italy
[5] Ist Nazl Fis Nucl, Sez Lecce, Lecce, Italy
关键词
D O I
10.1016/S0168-9002(00)01106-2
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 [仪器科学与技术]; 080401 [精密仪器及机械]; 081102 [检测技术与自动化装置];
摘要
This collaboration has already obtained encouraging results about the performance of 200 and 600 mum thick semiinsulating GaAs detectors, in terms of charge collection efficiency (c.c.e.), detection efficiency and imaging capabilities, to be used as single photon counting detectors with photons in the radiographic energy range. The possible use of such detectors in nuclear medicine implies detection of more energetic photons, up to the 140 keV of the Tc-99m source; thus efficient detectors must be thicker than those previously studied. In this paper, we present the results of the investigation of electrical and detection properties (spectra, c.c.e., energy resolution) of a series of detectors, whose thickness ranges from 200 to 1000 mum, when irradiated with radioactive sources over the energy range used in nuclear medicine. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:123 / 126
页数:4
相关论文
共 8 条
[1]
GaAs detector optimization for different medical imaging applications [J].
Amendolia, SR ;
Bertolucci, E ;
Bisogni, MG ;
Bottigli, U ;
Ciocci, MA ;
Conti, M ;
Delogu, P ;
Fantacci, ME ;
Maestro, P ;
Marzulli, V ;
Pernigotti, E ;
Romeo, N ;
Rosso, V ;
Russo, P ;
Stefanini, A ;
Stumbo, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 434 (01) :14-17
[2]
Amendolia SR, 1999, NUOVO CIMENTO A, V112, P167
[3]
Experimental study of LEC GaAs detectors for X-ray digital radiography [J].
Amendolia, SR ;
Bisogni, MG ;
Campbell, M ;
Cola, A ;
DAuria, S ;
DaVia, C ;
Heijne, EHM ;
Fantacci, ME ;
OShea, V ;
Rosso, V ;
Smith, K ;
Vasanelli, L .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 380 (1-2) :410-413
[4]
Amendolia SR, 1998, PHYS MEDICA, V14, P17
[5]
BENCIVELLI W, 1993, MATER RES SOC SYMP P, V302, P381, DOI 10.1557/PROC-302-381
[6]
BISOGNI MG, 1998, 43 SPIE P, P3445
[7]
A readout chip for a 64 x 64 pixel matrix with 15-bit single Photon Counting [J].
Campbell, M ;
Heijne, EHM ;
Meddeler, G ;
Pernigotti, E ;
Snoeys, W .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) :751-753
[8]
OTT RJ, 1995, PHYSICS MED IMAGING, P42