An in vitro assessment of gas trapping during high frequency oscillation

被引:4
作者
Leipälä, JA [1 ]
Sharma, A [1 ]
Lee, S [1 ]
Milner, AD [1 ]
Greenough, A [1 ]
机构
[1] Univ London Kings Coll, Guys Kings & St Thomas Sch Med, Dept Child Hlth, Div Asthma Allergy & Lung Biol, London WC2R 2LS, England
关键词
high frequency oscillatory ventilation; gas-trapping; inspiratory-to-expiratory ratio; neonate; oscillatory frequency;
D O I
10.1088/0967-3334/26/3/016
中图分类号
Q6 [生物物理学];
学科分类号
071011 [生物物理学];
摘要
Our aim was to assess the influence of oscillatory frequency, inspiratory-to-expiratory (I:E) ratio and airway pressure on gas trapping during high frequency oscillatory ventilation (HFOV). A lung model was used, which had a high compliance (4.0 ml (cmH(2)O)(-1)) and resistance (160 cmH(2)O l(-1) s(-1)) resulting in a long time constant (0.64 s). To assess whether gas trapping occurred, the mean pressure within the lung model (equivalent to alveolar pressure) was compared to the mean airway pressure (MAP) measured at the manifold (manifold MAP) of the two oscillators used, the SLE 5000 (I:E ratio 1:1) and the Sensor Medics 3100A (tested at l:E ratios of 1:2 and 1:1). The effects were assessed of raising the MAP from 15 to 35 cmH(2)O, the oscillatory amplitude from 30 to 70 cmH(2)O and the frequency from 5 to 15 Hz (5 to 20 Hz SLE only). There were no significant trends for differences between the pressure within the lung model and the 'manifold' MAP to increase (i.e. no evidence of gas trapping), as MAP, amplitude or frequency was increased, regardless of which oscillator or LE ratio was used. Increasing the pressure amplitude led to a progressive fall in the pressure within the lung model when an I:E ratio of 1:2 was used (p < 0.05). Our results suggest that significant gas trapping does not occur during HFOV even if there is high compliance and resistance.
引用
收藏
页码:329 / 336
页数:8
相关论文
共 14 条
[1]
Determination of gas-trapping during high frequency oscillatory ventilation [J].
Alexander, J ;
Milner, AD .
ACTA PAEDIATRICA, 1997, 86 (03) :268-273
[2]
BRYAN AC, 1986, AM REV RESPIR DIS, V133, P928
[3]
HIGH-FREQUENCY OSCILLATION FOR PRETERM INFANTS WITH SEVERE RESPIRATORY-FAILURE [J].
CHAN, V ;
GREENOUGH, A ;
GAMSU, HR .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 1994, 70 (01) :F44-F46
[4]
Comparison of two inspiratory: expiratory ratios during high frequency oscillation [J].
Dimitriou, G ;
Greenough, A ;
Kavvadia, V ;
Milner, AD .
EUROPEAN JOURNAL OF PEDIATRICS, 1999, 158 (10) :796-799
[5]
FROESE AB, 1987, AM REV RESPIR DIS, V135, P1363, DOI 10.1007/BF03009896
[6]
PROXIMAL, TRACHEAL, AND ALVEOLAR PRESSURES DURING HIGH-FREQUENCY OSCILLATORY VENTILATION IN A NORMAL RABBIT MODEL [J].
GERSTMANN, DR ;
FOUKE, JM ;
WINTER, DC ;
TAYLOR, AF ;
DELEMOS, RA .
PEDIATRIC RESEARCH, 1990, 28 (04) :367-374
[7]
MEASUREMENT OF TIDAL LUNG-VOLUMES IN NEONATES DURING HIGH-FREQUENCY OSCILLATION [J].
HOSKYNS, EW ;
MILNER, AD ;
HOPKIN, IE .
JOURNAL OF BIOMEDICAL ENGINEERING, 1992, 14 (01) :16-20
[8]
High-frequency oscillatory ventilation for the prevention of chronic lung disease of prematurity [J].
Johnson, AH ;
Peacock, JL ;
Greenough, A ;
Marlow, N ;
Limb, ES ;
Marston, L ;
Calvert, SA .
NEW ENGLAND JOURNAL OF MEDICINE, 2002, 347 (09) :633-642
[9]
PHYSICAL PROPERTIES OF HUMAN LUNGS MEASURED DURING SPONTANEOUS RESPIRATION [J].
MEAD, J ;
WHITTENBERGER, JL .
JOURNAL OF APPLIED PHYSIOLOGY, 1953, 5 (12) :779-796
[10]
Effects of smoking in pregnancy on neonatal lung function [J].
Milner, AD ;
Marsh, MJ ;
Ingram, DM ;
Fox, GF ;
Susiva, C .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 1999, 80 (01) :F8-F14