Technical variability of the RT3 accelerometer

被引:92
作者
Powell, SM
Jones, DI
Rowlands, AV [1 ]
机构
[1] Univ Wales, Sch Sport Hlth & Exercise Sci, Bangor LL57 2PX, Gwynedd, Wales
[2] Univ Wales, Sch Informat, Bangor LL57 2PX, Gwynedd, Wales
关键词
physical activity; activity assessment; triaxial accelerometry; activity monitor;
D O I
10.1249/01.MSS.0000089341.68754.BA
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
Purpose: To evaluate the technical performance of the RT3 triaxial accelerometer. Methods: Twenty-three RT3 accelerometers were subjected to a specific vibration along each sensitive axis in isolation, using a motorized vibration table that produced frequencies of 2.1, 5.1, and 10.2 Hz, respectively. Data were analyzed for frequency and axis effects and inter- and intra-instrument variability. Results: ANOVA showed a frequency by axis interaction (F-2,F-1,F-36,F-8 = 19.9 P < 0.001). Post hoc tests revealed the Y axis count to be significantly higher than the X and Z axes counts at 5.1 and 10.2 Hz. There was no difference in counts between axes at 2.1 Hz. Interinstrument coefficients of variation (CV) decreased as frequency increased (21.9 to 26.7% at 2.1 Hz, 6.3 to 9.0% at 5.1 Hz, and 4.2 to 7.2% at 10.2 Hz). The intraclass correlation (ICC) between RT3s was 0.99. regardless of the axis. Intra-instrument CV also decreased as frequency increased (2.1 to 56.2%, 0.3 to 2.5%, and 0.2 to 2.9% at 2.1, 5.1, and 10.2 Hz, respectively. Conclusion: There were no differences in counts recorded on the X, Y, and Z axes at 2.1 Hz: however, the counts recorded along the Y axis were significantly higher than the counts at the X and Z axes at 5.1 and 10.2 Hz. Due to large coefficients of variation for both inter- and intra-instrument variability at 2.1 Hz, testing the inter- and intra-instrument variability of the accelerometers before use is recommended.
引用
收藏
页码:1773 / 1778
页数:6
相关论文
共 11 条
[1]   Validity and reliability issues in objective monitoring of physical activity [J].
Bassett, DR .
RESEARCH QUARTERLY FOR EXERCISE AND SPORT, 2000, 71 (02) :S30-S36
[2]   PHYSICAL-FITNESS AND ALL-CAUSE MORTALITY - A PROSPECTIVE-STUDY OF HEALTHY-MEN AND WOMEN [J].
BLAIR, SN ;
KOHL, HW ;
PAFFENBARGER, RS ;
CLARK, DG ;
COOPER, KH ;
GIBBONS, LW .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1989, 262 (17) :2395-2401
[3]   The physical activity, fitness and health of children [J].
Boreham, C ;
Riddoch, C .
JOURNAL OF SPORTS SCIENCES, 2001, 19 (12) :915-929
[4]   Validity of heart rate, pedometry, and accelerometry for predicting the energy cost of children's activities [J].
Eston, RG ;
Rowlands, AV ;
Ingledew, DK .
JOURNAL OF APPLIED PHYSIOLOGY, 1998, 84 (01) :362-371
[5]  
FRANKLIN GF, 2002, FEEDBACK CONTROL DYN, P910
[6]   Validity of accelerometry for the assessment of moderate intensity physical activity in the field [J].
Hendelman, D ;
Miller, K ;
Bagget, C ;
Debold, E ;
Freedson, P .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2000, 32 (09) :S442-S449
[7]   Technical reliability of the CSA activity monitor: The EarlyBird Study [J].
Metcalf, BS ;
Curnow, JSH ;
Evans, C ;
Voss, LD ;
Wilkin, TJ .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2002, 34 (09) :1533-1537
[8]  
Rowlands A.V., 2001, KINANTHROPOMETRY EXE, P151
[9]  
Rowlands AV, 2000, ANN HUM BIOL, V27, P479, DOI 10.1080/030144600419314
[10]  
STEVENS J, 1996, APPL MULTIVARIATE ST, P479