Drain entrapment and titanium to ceramic head deposition - Two unique complications following closed reduction of a dislocated total hip arthroplasty

被引:27
作者
Luchetti, WT [1 ]
Copley, LA [1 ]
Vresilovic, EJ [1 ]
Black, J [1 ]
Steinberg, ME [1 ]
机构
[1] Hosp Univ Penn, Dept Orthopaed Surg, Philadelphia, PA 19104 USA
关键词
total hip arthroplasty; femoral head; closed reduction; titanium; drainage tube;
D O I
10.1016/S0883-5403(98)80018-1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Postoperative dislocation remains one of the most frequent complications following total hip replacement. In this report, a case is presented that illustrates two potential concerns with postoperative dislocation and subsequent closed reduction. The first complication presented in this report is entrapment of a closed drainage system tube in the joint space following closed reduction. The second complication, transfer of metallic debris to a ceramic femoral head from contact with an acetabular shell during closed reduction, was documented by analysis of a femoral head using scanning electron microscopy and energy dispersive x-ray spectrometry. This report emphasizes the need for the surgeon to express caution when relocating a dislocated hip, particularly when a closed drainage system is used postoperatively.
引用
收藏
页码:713 / 717
页数:5
相关论文
共 13 条
[1]
METALLIC WEAR IN FAILED TITANIUM-ALLOY TOTAL HIP REPLACEMENTS - A HISTOLOGICAL AND QUANTITATIVE-ANALYSIS [J].
AGINS, HJ ;
ALCOCK, NW ;
BANSAL, M ;
SALVATI, EA ;
WILSON, PD ;
PELLICCI, PM ;
BULLOUGH, PG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) :347-356
[2]
BLACK J, 1991, BIOL PERFORMANCE MAT, P110
[3]
GLANT TT, 1991, ORTHOP T, V15, P540
[4]
KELMAN GJ, 1989, COMPLICATIONS ORTHOP, V4, P68
[5]
IN-VIVO AND IN-VITRO ANALYSIS OF MEMBRANES FROM HIP PROSTHESES INSERTED WITHOUT CEMENT [J].
KIM, KJ ;
CHIBA, J ;
RUBASH, HE .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1994, 76A (02) :172-180
[6]
POSTOPERATIVE DISLOCATION OF A TOTAL HIP-PROSTHESIS DUE TO A SURGICAL DRAIN [J].
LARSON, BJ ;
ZINDRICK, M ;
SCHWARTZ, C ;
DEMOS, TC .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1987, 149 (05) :971-972
[7]
LINDER L, 1983, CLIN ORTHOP RELAT R, P93
[8]
ASEPTIC LOOSENING IN TOTAL HIP-ARTHROPLASTY SECONDARY TO OSTEOLYSIS INDUCED BY WEAR DEBRIS FROM TITANIUM-ALLOY MODULAR FEMORAL HEADS [J].
LOMBARDI, AV ;
MALLORY, TH ;
VAUGHN, BK ;
DROUILLARD, P .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1989, 71A (09) :1337-1342
[9]
MORREY BF, 1992, ORTHOP CLIN N AM, V23, P237
[10]
COMPLICATIONS OF TOTAL HIP ARTHROPLASTY TREATED BY REOPERATION [J].
NOLAN, DR ;
FITZGERALD, RH ;
BECKENBAUGH, RD ;
COVENTRY, MB .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1975, 57 (07) :977-981