On-line enantiomeric analysis using high-performance liquid chromatography in chiral, separation by simulated moving bed

被引:13
作者
Araujo, Joao M. M. [1 ]
Rodrigues, Rui. C. R. [1 ]
Eusebio, Mario F. J. [1 ]
Mota, Jose P. B. [1 ]
机构
[1] Univ Nova Lisboa, Fac Ciencias & Tecnol, Dept Quim, Requim CQFB, P-2829516 Caparica, Portugal
关键词
HPLC; on-line monitoring; chiral technology; simulated moving-bed;
D O I
10.1016/j.chroma.2007.10.102
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An automated on-line enantiomeric analysis system comprising an analytical HPLC set-up with two UV detectors sharing the same light source has been employed to monitor the internal composition profile in chiral simulated moving bed chromatography. This monitoring scheme does not use a polarimeter. Using a sampling interface placed between two SMB columns, effluent samples are directed onto a high-efficiency analytical column at a sampling rate faster than the overall dynamics of the preparative unit to achieve on-line enantiomeric analysis of the composition profile. The other UV detector is placed in the SMB loop before the fraction collector to provide instantaneous measurement of the total enantiomeric concentration. The feasibility and effectiveness of the on-line enantimeric monitoring scheme were assessed experimentally on the separation of Troger's base racemate, using Chiralpak AD as stationary phase and methanol as eluent. It was found that robust monitoring of the concentration profiles of the individual enantiomers is best achieved when the enantiomeric purity obtained from the peak areas of the on-line enantiomer analysis chromatograms is combined with the on-line UV measurement of total enantiomeric concentration. The accuracy and robustness of the proposed on-line enantiomeric monitoring system open up promising perspectives for process control and dynamic optimization of the SMB. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:292 / 301
页数:10
相关论文
共 45 条
[1]  
[Anonymous], 1992, CHIRALITY, V4, P338
[2]  
[Anonymous], PHARM TECH EUR
[3]   Application of gradients in the simulated moving bed process [J].
Antos, D ;
Seidel-Morgenstern, A .
CHEMICAL ENGINEERING SCIENCE, 2001, 56 (23) :6667-6682
[4]  
ARAUJO JMM, UNPUB J CHROMATOGR A
[5]  
ARAUJO JMM, 2006, P SPICA 2006 S INNSB
[6]   Optimal design and operation of a certain class of asynchronous simulated moving bed processes [J].
Araujo, Joao M. M. ;
Rodrigues, Rui. C. R. ;
Mota, Jose P. B. .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1132 (1-2) :76-89
[7]   Use of single-column models for efficient computation of the periodic state of a simulated moving-bed process [J].
Araujo, Joao M. M. ;
Rodrigues, Rui. C. R. ;
Mota, Jose P. B. .
INDUSTRIAL & ENGINEERING CHEMISTRY RESEARCH, 2006, 45 (15) :5314-5325
[8]  
ATTILA R, 2003, J CHROMATOGR A, V986, P207
[9]   Instrumental method for automated on-line fraction analysis and peak deconvolution in multicomponent-overloaded high-performance liquid chromatography [J].
Cavazzini, Alberto ;
Costa, Valentina ;
Nadalini, Giorgio ;
Dondi, Francesco .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1137 (01) :36-41
[10]   Kinetic and equilibrium study of the enantioseparation of fenoprofen in a batch setup [J].
Ching, CB ;
Arlt, W ;
Lisso, M ;
Wozny, G .
INDUSTRIAL & ENGINEERING CHEMISTRY RESEARCH, 2000, 39 (11) :4365-4369